球囊扩张式烟囱移植物联合Endurant内移植物在肾动脉的体外评估

In Vitro Evaluation of Balloon-Expandable Chimney Grafts in the Renal Arteries Combined With the Endurant Endograft.
PubMed 文献 · J Endovasc Ther · 2019-08-28

本研究在基于患者CT血管造影制作的硅胶模型中,比较了三种球囊扩张式覆膜支架(Advanta V12、VBX、BeGraft+)作为肾动脉烟囱移植物与Endurant内移植物联合使用时的表现。结果显示三种支架在沟槽面积和烟囱移植物通畅管腔直径上无显著差异,读者间一致性良好,但需进一步临床评估确认。

要点

  • 三种球囊扩张式覆膜支架作为肾动脉烟囱移植物时,沟槽面积无显著差异。
  • 烟囱移植物管腔直径在三个测量水平上组间均无显著差异。
  • 体外结果需临床研究验证。

核心结论

  • 体外实验表明,Advanta V12与VBX、BeGraft+作为肾动脉烟囱移植物联合Endurant内移植物时结果相当。

研究设计与样本

  • 基于患者CT血管造影制作硅胶模型,置于37°C模拟血液 pulsed 泵系统(140/60 mm Hg)中。
  • 使用36 mm Endurant II内移植物和6×59 mm球囊扩张式覆膜支架烟囱移植物,共9次测试(每种组合3次)。
  • 两名盲法放射科医生独立测量沟槽面积和烟囱移植物管腔直径。

主要发现

  • 平均沟槽面积:Advanta 19.3±7 mm²,VBX 20.2±8 mm²,BeGraft+ 22.3±8 mm²(均p>0.05)。
  • 肾动脉上支架杆水平平均直径:Advanta 4.46±0.3 mm,VBX 4.12±0.4 mm,BeGraft+ 4.12±0.3 mm(均p>0.05)。
  • 烟囱移植物最大成角水平平均直径:Advanta 3.77±0.3 mm,VBX 4.16±0.1 mm,BeGraft+ 3.74±0.3 mm(均p>0.05)。
  • 右肾动脉内平均直径:Advanta 3.91±0.2 mm,VBX 4.05±0.9 mm,BeGraft+ 4.3±0.4 mm(均p>0.05)。
  • 读者间组内相关系数(ICC)为0.7至0.9,一致性良好。

临床怎么用

  • 在体外模型中,三种球囊扩张式覆膜支架作为肾动脉烟囱移植物联合Endurant内移植物时表现相当。

局限性

  • 需进一步临床评估确认体外结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Purpose: To compare balloon-expandable covered stents (BECS) available for use as renal chimney grafts to the Advanta V12 approved for use with the Endurant endograft. Materials and Methods: A silicone model was manufactured based on preoperative computed tomography angiography (CTA) scans of a patient with a 7-cm juxtarenal aneurysm treated with an Endurant bifurcated endograft and an Advanta V12 covered stent for the 6-mm-diameter right renal artery. The model was placed in a flow box filled with a 37°C blood-mimicking solution equipped with a pulsatile pumping system (140/60 mm Hg). The tested BECS were the Advanta V12, the VBX, and the BeGraft+. A 36-mm-diameter Endurant II endograft with a 16-mm-diameter contralateral limb and a 6×59-mm BECS chimney graft were used in 9 consecutive tests (3 for each combination). After each implantation, the model was placed in the CT scanner, and 2 radiologists blinded to the test device independently measured the gutter areas and the patent chimney graft lumen at the level of the Endurant's suprarenal stent, at the level of the chimney's maximum curvature, and 10 mm inside the renal artery. The intraclass correlation coefficients (ICC) were calculated to assess interreader reliability. Results: The mean gutter areas were 19.3±7, 20.2±8, and 22.3±8 mm2 for the Advanta, VBX, and BeGraft+, respectively (all p>0.05). At the level of the aortic endograft's suprarenal stent struts, the mean diameter of the Advanta V12, VBX, and BeGraft+ were 4.46±0.3, 4.12±0.4, and 4.12±0.3 mm, respectively (all p>0.05). At the level of the maximum chimney graft angulation, the mean diameters were 3.77±0.3, 4.16±0.1, and 3.74±0.3 mm, respectively (all p>0.05). In the right renal artery, the mean diameters were 3.91±0.2, 4.05±0.9, and 4.3±0.4 mm, respectively (all p>0.05). The ICCs varied between 0.7 and 0.9, indicating good agreement between readers. Conclusion: These in vitro findings showed comparable results between the Advanta V12 and the available BECS used in conjunction with the Endurant endograft according to the instructions for use. Further clinical evaluation is needed to confirm these results.