BeGraft与BeGraft+支架移植物作为开窗EVAR桥接装置的体外性能研究
本研究在模拟支架移植物开窗的体外模型中,测试了20枚BeGraft和20枚BeGraft+球囊扩张覆膜支架(6mm和8mm直径)在喇叭口扩张后的形态及抗移位能力。结果显示喇叭口扩张改变支架几何形态但无ePTFE覆盖破损或支架断裂,BeGraft+在拉脱力和剪切力方面均优于BeGraft,表现出更强的顺应性。
要点
- 喇叭口扩张改变支架几何形态但未导致支架断裂或ePTFE覆盖破损。
- 6mm和8mm直径BeGraft+的拉脱力均显著高于BeGraft。
- BeGraft+在150%支架直径时的剪切力显著高于BeGraft。
核心结论
- 喇叭口扩张改变支架几何形态但不引起支架断裂,BeGraft+在拉脱力和剪切力方面优于BeGraft,表现出更强的顺应性。
研究设计与样本
- 体外实验研究,共测试20枚BeGraft和20枚BeGraft+钴铬合金ePTFE覆膜支架,直径6mm和8mm,在模拟支架移植物开窗的薄片中测试。
主要发现
- 喇叭口扩张后未检测到ePTFE覆盖改变,数字X线和CT显示支架几何形态明显改变,BeGraft组更显著,但未检测到断裂。
- 6mm支架拉脱力中位数:BeGraft为17.1 N(15.8-19.6),BeGraft+为30.4 N(20.2-31.9),p=0.006。
- 8mm支架拉脱力中位数:BeGraft为11.3 N(11-12.1),BeGraft+为21.8 N(18.2-25.5),p<0.001。
- 剪切力测试在150%支架直径时:6mm BeGraft与BeGraft+中位力分别为10.5 vs 15.28 N,8mm模型分别为15.23 vs 20.72 N(p=0.016和0.017)。
临床怎么用
- 在开窗EVAR中选择桥接支架时,BeGraft+可能提供更高的抗拉脱和抗剪切能力,但需结合临床实际评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Purpose: To investigate 2 generations of balloon-expandable covered stents as potential bridging devices using an in vitro model of stent-graft fenestrations. Materials and Methods: Twenty BeGraft and 20 BeGraft+ cobalt-chromium stents covered in expanded polytetrafluoroethylene (ePTFE) in 6- and 8-mm diameters were tested in sheets mimicking stent-graft fenestrations. Microscopy and radiography were employed to evaluate stent morphology after flaring. In vitro bench tests measured maximum pullout (perpendicular displacement) and the shear stress (axial displacement) forces needed to dislocate the stents. Results: No alteration of ePTFE coverage was detected in the flared stents. Digital radiography and computed tomography showed marked alteration of the stent geometry, which was more pronounced in the BeGraft group. No fractures were detected. Median (minimum-maximum) pullout forces for the 6-mm stent-grafts were 17.1 N (15.8-19.6) for the BeGraft device and 30.4 N (20.2-31.9) for the BeGraft+ device (p=0.006). Median (minimum-maximum) pullout forces for the 8-mm stent-grafts were 11.3 N (11-12.1) for the BeGraft device and 21.8 N (18.2-25.5) for the BeGraft+ device (p<0.001). The shear stress test showed median forces of 10.5 vs 15.28 N at 150% of the stent diameter for the 6-mm BeGraft and BeGraft+ stent-grafts, respectively, and 15.23 vs 20.72 N at 150% stent diameter for the 8-mm models (p=0.016 and 0.017, respectively). Conclusion: Flaring changed the stent geometry but did not provoke stent fractures. The BeGraft+ is superior to the BeGraft in terms of pullout and shear stress forces, demonstrating greater resilience.