文献速递静脉血栓栓塞

下腔静脉滤器支柱穿透的长期CT随访结果

Long-term computed tomography follow-up results of strut penetration of inferior vena cava filters.
PubMed 文献 · J Vasc Surg Venous Lymphat Disord · 2019-06-21

本研究纳入66例未取出下腔静脉滤器的深静脉血栓患者,随访CT中位14个月,所有患者均检出支柱穿透,其中35%为累及器官的3级穿透。

要点

  • 所有未取出滤器患者均存在支柱穿透,3级穿透占35%。
  • 滤器留置≥30个月与3级穿透风险增加独立相关(OR 4.395)。
  • 可回收滤器无需时应尽早取出并密切随访。

核心结论

  • 无论滤器类型,下腔静脉滤器支柱穿透发生率均高,且邻近器官受累风险随时间增加。

研究设计与样本

  • 回顾性研究,纳入2003年6月至2015年8月138例置入下腔静脉滤器的深静脉血栓患者,其中104例未取出滤器者的66例有CT随访被纳入分析。

主要发现

  • 中位随访14个月,所有患者均检出支柱穿透,1级、2级、3级分别为29%、36%、35%。

临床怎么用

  • 对留置可回收下腔静脉滤器的患者应密切CT随访,无需时尽早取出以降低器官受累风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: This study aimed to investigate the incidence of inferior vena cava (IVC) filter strut penetration and risk factors of organ involvement. METHODS: From June 2003 to August 2015, there were 138 patients with deep venous thrombosis who received an IVC filter. Among 104 patients who did not have the IVC filter retrieved, 66 had follow-up computed tomography and were included in this study. The IVC filters used were 21 Günther Tulip (Cook Medical, Bloomington, Ind), 26 Celect (Cook Medical), and 19 OptEase (Cordis Corp, Bridgewater, NJ) filters. Filter strut penetration was categorized by a previously published scale of grade 0 to grade 3, and organ involvement was specifically assessed. Multivariate analysis was used to identify risk factors for organ-involving strut penetration (grade 3). RESULTS: The median age of the patients was 66 years (27-84 years), and 46% were male. Median computed tomography follow-up duration was 14 months (1-137 months). IVC strut penetration was detected in all patients. Grade 1, grade 2, and grade 3 were 29%, 36%, and 35%, respectively. The risk factor of grade 3 penetration was indwelling time ≥30 months on binary logistic regression analysis (odds ratio, 4.395; 95% confidence interval, 1.179-16.385; P = .027). CONCLUSIONS: Regardless of type of IVC filter, the incidence of strut penetration was high. The risk of adjacent organ involvement increases over time. Retrievable IVC filters need close follow-up and retrieval as soon as they are no longer needed.

来源:J Vasc Surg Venous Lymphat Disord,查看原文。版权归原作者所有。

引用

  1. Yang SS, Yun WS. Long-term computed tomography follow-up results of strut penetration of inferior vena cava filters. J Vasc Surg Venous Lymphat Disord. 2019. PMID: 31231055. doi:10.1016/j.jvsv.2019.03.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。