Sn-EVAS术后肾动脉及snorkel支架的心脏搏动与呼吸形变
本研究对10例Sn-EVAS患者行心电门控、呼吸分辨CT血管造影,量化肾动脉及snorkel支架因心脏搏动和呼吸产生的形变,并与16例未治疗AAA及11例Sn-EVAR患者比较。结果显示呼吸对肾动脉形变影响大于心脏搏动,左侧更明显;Sn-EVAR和Sn-EVAS中支架端弯曲及最大曲率变化显著大于未治疗AAA,提示可能影响支架机械耐久性。
要点
- 呼吸所致肾动脉形变大于心脏搏动,左侧约为右侧3倍。
- Sn-EVAR和Sn-EVAS中呼吸致最大曲率变化约为未治疗AAA的10倍。
- snorkel支架刚度抑制肾分支成角但加重支架远端动脉弯曲。
核心结论
- 心脏和呼吸影响可能挑战Sn-EVAS中snorkel支架的机械耐久性,呼吸可能是支架端血栓风险的主要因素,尤其左肾动脉。
研究设计与样本
- 对10例Sn-EVAS患者(平均年龄75±6岁,8例男性)行心电门控、呼吸分辨CT血管造影,基于三维几何模型量化肾动脉和支架成角与曲率变化。
- 呼吸所致运动与16例既往报道的未治疗AAA患者及11例Sn-EVAR患者进行比较。
主要发现
- 支架端肾动脉弯曲呼吸影响大于心脏影响(6°±7° vs -1°±2°,p<0.025)。
- 呼吸对左肾动脉形变约为右侧3倍。
- 最大曲率变化呼吸大于心脏(0.49±0.29 vs 0.24±0.17 cm-1,p<0.025),snorkel肾支架心脏与呼吸最大曲率变化相似(0.14±0.10 vs 0.19±0.09 cm-1,p=0.142)。
- 三组比较:未治疗AAA有显著肾分支成角,Sn-EVAR和Sn-EVAS无;Sn-EVAR和Sn-EVAS有显著支架端弯曲;呼吸致最大曲率变化Sn-EVAR和Sn-EVAS约为未治疗AAA的10倍。
临床怎么用
- 关注Sn-EVAS和Sn-EVAR术后snorkel支架的机械耐久性及支架端血栓风险,尤其左肾动脉。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Purpose: To quantify deformations of renal arteries and snorkel stents after snorkel endovascular aneurysm sealing (Sn-EVAS) resulting from cardiac pulsatility and respiration and compare these deformations to patients with untreated abdominal aortic aneurysms (AAA) and snorkel endovascular aneurysm repair (Sn-EVAR). Materials and Methods: Ten Sn-EVAS patients (mean age 75±6 years; 8 men) were scanned with cardiac-gated, respiration-resolved computed tomography angiography. From 3-dimensional geometric models, changes in renal artery and stent angulation and curvature due to cardiac pulsatility and respiration were quantified. Respiration-induced motions were compared with those of 16 previously reported untreated AAA patients and 11 Sn-EVAR patients. Results: Renal artery bending at the stent end was greater for respiratory vs cardiac influences (6°±7° vs -1°±2°, p<0.025). Respiration caused a 3-fold greater deformation on the left renal artery as compared with the right side. Maximum curvature change was higher for respiratory vs cardiac influences (0.49±0.29 vs 0.24±0.17 cm-1, p<0.025), and snorkel renal stents experienced similar maximum curvature change due to cardiac pulsatility and respiration (0.14±0.10 vs 0.19±0.09 cm-1, p=0.142). When comparing the 3 patient cohorts for respiratory-induced deformation, there was significant renal branch angulation in untreated AAAs, but not in Sn-EVAR or Sn-EVAS, and there was significant bending at the stent end in Sn-EVAR and Sn-EVAS. Maximum curvature change due to respiration was ~10-fold greater in Sn-EVAR and Sn-EVAS compared to untreated AAAs. Conclusion: The findings suggest that cardiac and respiratory influences may challenge the mechanical durability of snorkel stents of Sn-EVAS; similarly, however, respiration may be the primary culprit for tissue irritation, increasing the risk for stent-end thrombosis, especially in the left renal artery. The bending stiffness of snorkel stents in both the Sn-EVAR and Sn-EVAS cohorts damped renal branch angulation while it intensified bending of the artery distal to the snorkel stent. Understanding these device-to-artery interactions is critical as they may affect mechanical durability of branch stents and quality and durability of treatment.