B型主动脉夹层患者对现成胸主动脉单分支支架的解剖适用性

Anatomic suitability for "off-the-shelf" thoracic single side-branched endograft in patients with type B aortic dissection.
PubMed 文献 · J Vasc Surg · 2019-06-24

通过三维中心线重建术前CTA评估解剖条件,仅16例(28%)满足胸主动脉单分支支架(TSSB)的全部解剖要求,主要限制为分支血管间长度不足。

要点

  • 仅28%需Zone 2 TEVAR的B型夹层患者符合TSSB解剖标准。
  • 49%患者因弓部分支间长度不足无法避免覆盖近端分支。
  • 未来器械设计需考虑左颈总动脉与左锁骨下动脉间短距离。

核心结论

  • 仅28%需Zone 2 TEVAR的B型夹层患者满足TSSB全部解剖要求。
  • 左颈总动脉与左锁骨下动脉间距离短是主要限制因素。

研究设计与样本

  • 对术前CTA进行三维中心线重建,评估主动脉直径、分支间距、目标分支直径及破口位置。
  • 仅纳入符合Zone 2 TEVAR标准的57例患者进行分析。

主要发现

  • 57例(66%)需要Zone 2 TEVAR。
  • TEVAR指征包括灌注不良12例、动脉瘤15例、持续疼痛22例、破裂3例、难控高血压5例、其他3例。
  • 平均随访19个月(范围1-72个月)。
  • 仅16/57例(28%)满足全部解剖要求。
  • 仅49%患者有足够的分支间长度以避免覆盖近端分支。

关键数值与推荐

  • TSSB解剖适用性定义为满足所有解剖要求,仅28%患者符合。
  • 分支间长度不足导致49%患者无法避免覆盖近端分支。

临床怎么用

  • 在考虑使用TSSB装置行Zone 2 TEVAR前,应通过CTA三维重建评估分支间距离等解剖条件。
  • 多数B型夹层患者因分支间距短而不适合现成TSSB装置,需个体化选择。

局限性

  • 单中心回顾性研究,样本量有限。
  • 结果基于特定TSSB装置解剖要求,外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Treatment of type B aortic dissections with thoracic endovascular aortic repair (TEVAR) has been adopted in many centers with the goal of covering the proximal entry tear. Coverage of the left subclavian artery (LSCA) is commonly required to achieve a dissection-free proximal seal zone. A novel thoracic single side-branched (TSSB) endograft device offers a potential off-the-shelf option to achieve total endovascular incorporation of LSCA during zone 2 TEVAR. The aim of this study was to determine what percentage of patients with type B aortic dissection who require zone 2 TEVAR meet the anatomical requirements for this device. METHODS: All consecutive patients undergoing TEVAR for type B aortic dissections at a single institution from 2006 to 2016 were evaluated. Three-dimensional centerline reconstruction of preoperative computed tomography angiography was performed to identify the diameter of the aorta, distances between branch vessels, diameter of the target branch vessel, and location of the primary entry tear. Only patients who met criteria for zone 2 TEVAR were included in the analysis. The primary outcome was percentage of patients that meet all anatomical requirements for TSSB. Individual criteria were evaluated independently, and results were stratified by dissection chronicity. RESULTS: Eighty-seven patients who underwent TEVAR for Stanford type B aortic dissections were reviewed. Fifty-seven (66%) would have required zone 2 TEVAR. Indications for TEVAR were malperfusion (12), aneurysm (15), persistent pain (22), rupture (3), uncontrolled hypertension (5), and other (3). Mean follow-up was 19 months (range, 1-72 months). Only 16 of the 57 patients (28%) met all the requirements for anatomic suitability. The primary contributor was that only 49% of patients had sufficient length between arch branches to prevent coverage of a proximal branch. CONCLUSIONS: Although the new TSSB device can allow for a more proximal seal zone and eliminate the need for open aortic arch debranching, only 28% of patients with type B dissection who required zone 2 TEVAR met all the anatomic requirements for this device. Future devices will need to account for the short distance between the left carotid and LSCA to be more broadly applicable.