新型CGuard MicroNet覆盖「一码通」颈动脉支架的初步临床结果与体外测试
本研究评估一种可自适应不同血管直径的MicroNet覆盖颈动脉支架,连续纳入30例症状性颈内动脉狭窄患者(平均年龄72.1±7.7岁,26例男性,狭窄程度86.3%±6.4%),使用「一码通」CGuard支架(长度30或40 mm)。支架在全部30例中成功植入,围手术期无并发症,30天内无神经事件。体外测试显示支架在5.5至9.0 mm扩张直径范围内径向力近乎等效。
要点
- 「一码通」CGuard支架在30例症状性颈内动脉狭窄患者中成功植入,围手术期无并发症。
- 30天内无神经事件发生。
- 体外测试显示支架在5.5至9.0 mm扩张直径范围内径向力近乎等效。
核心结论
- 新型自适应MicroNet覆盖支架在5.5至9.0 mm扩张直径范围内具有高顺应性和近乎等效的径向力,初步临床结果显示可安全植入参考直径在此范围内的颈内动脉。
研究设计与样本
- 连续纳入30例症状性颈内动脉狭窄患者(平均年龄72.1±7.7岁,26例男性,狭窄程度86.3%±6.4%),使用「一码通」CGuard支架(长度30或40 mm)。
- 体外测试使用分段头端径向力测试装置测定支架的慢性外向力,支架在5.0 mm直径下释放,测量至10.0 mm最大扩张。
主要发现
- 支架在全部30例患者中成功植入,围手术期无并发症,包括30天内无神经事件。
- 按支架长度归一化的慢性外向力显示近乎等效的径向力结果:9.0 mm处最小径向力为0.195 N/mm,5.5 mm处最大径向力为0.330 N/mm,差异微小。
临床怎么用
- 对于参考直径在5.5至9.0 mm范围内的症状性颈内动脉狭窄患者,可考虑使用该「一码通」MicroNet覆盖支架进行植入。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Purpose: To evaluate a MicroNet-covered stent designed for the carotid artery with the new ability to adjust to different vessel diameters. Materials and Methods: Thirty consecutive patients (mean age 72.1±7.7 years; 26 men) with symptomatic stenosis (86.3%±6.4%) of the internal carotid artery were treated with the new self-adjusting nitinol stent, which has a self-expanding, open-cell design covered by an outer conformable layer (MicroNet). The only stent used was the "One-Size-Fits-All" CGuard stent with lengths of 30 or 40 mm. In bench testing, the chronic outward force of the One-Size-Fits-All stent was determined with a segmented head radial force test device. The stent was deployed directly into the test device at a diameter of 5.0 mm, and the chronic outward force was measured up to 10.0 mm, the maximum expansion of the stent. Results: The stent was successfully implanted in all 30 patients without periprocedural complications, including no neurological events within 30 days. The chronic outward force normalized by stent length demonstrated a near-equivalent radial force outcome: The stent displayed only a minor difference between the minimal radial force at 9.0 mm (0.195 N/mm) and the maximal radial force at 5.5 mm (0.330 N/mm). Conclusion: The new self-adjusting, MicroNet-covered stent has high conformability combined with an almost equivalent radial force at expansion diameters ranging from 5.5 to 9.0 mm. The first clinical results demonstrate that the new One-Size-Fits-All stent can be safely implanted in internal carotid arteries with reference diameters within this range.
引用
- Wissgott C, Brandt-Wunderlich C, Kopetsch C, et al. Initial Clinical Results and In Vitro Testing of the New CGuard MicroNet-Covered "One-Size-Fits-All" Carotid Stent. J Endovasc Ther. 2019. PMID: 31060430. doi:10.1177/1526602819849078
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。