TEVAR中气体栓塞分布取决于锚定区:搏动流模型研究
在玻璃主动脉搏动流模型中释放10枚胸主动脉覆膜支架,5枚位于2区、5枚位于3区,记录主动脉上分支及主动脉内气泡总量。2区组右侧锁骨下动脉和右颈总动脉气泡量更大,3区组降主动脉气泡量显著更高;两组主动脉上分支总气泡量无差异。
要点
- 2区释放时右锁骨下动脉和右颈总动脉气泡量显著高于3区。
- 3区释放时降主动脉气泡量显著高于2区。
- 释放过程中持续出现小气泡,大气泡多见于近端支架释放时及近端释放后。
核心结论
- 在2区和3区释放管状胸主动脉覆膜支架时,气体释放至所有主动脉上分支和降主动脉。
- 2区释放时右侧主动脉上分支气泡量更高,3区释放时降主动脉气泡量显著更高。
研究设计与样本
- 在具有I型主动脉弓的玻璃主动脉流模型中释放10枚相同的在研管状胸主动脉覆膜支架,2区和3区各5枚。
- 搏动泵产生5 L/min流量,收缩压和舒张压(±5%)分别为105和70 mm Hg。
- 锁骨下动脉流量(±5%)为300 mL/min,椎动脉为220 mL/min,颈总动脉(CCA)为400 mL/min。
- 记录每个主动脉上分支和主动脉内释放的气体总量。
主要发现
- 2区组和3区组测得的气体平均量分别为0.82±0.23 mL和0.94±0.28 mL(p=0.49)。
- 2区组与3区组相比,右锁骨下动脉释放的气体量更大(0.07±0.02 vs 0.02±0.02 mL,p<0.01),右颈总动脉也更大(0.30±0.8 vs 0.18±07 mL,p=0.04)。
- 右椎动脉和所有左侧主动脉上分支的气体平均量在两组间无差异。
- 3区组降主动脉释放的气体量显著高于2区组(0.48±0.12 vs 0.13±0.08 mL,p<0.01)。
- 释放过程中持续观察到小气泡,大气泡更常见于近端支架移植物释放时及近端释放后。
临床怎么用
- 2区TEVAR释放时需关注右侧主动脉上分支(右锁骨下动脉、右颈总动脉)的气体栓塞风险。
- 3区TEVAR释放时需关注降主动脉的气体栓塞风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Purpose: To analyze the distribution of air bubbles in the supra-aortic vessels during thoracic stent-graft deployment in zones 2 and 3 in an aortic flow model. Materials and Methods: Ten identical, investigational, tubular, thoracic stent-grafts were deployed in a glass aortic flow model with a type I arch: 5 in zone 2 and 5 in zone 3. A pulsatile pump generated a flow of 5 L/min with systolic and diastolic pressures (±5%) of 105 and 70 mm Hg, respectively. The flow rates (±5%) were 300 mL/min in the subclavian arteries, 220 mL/min in the vertebral arteries, and 400 mL/min in the common carotid arteries (CCAs). The total amounts of air released in each supra-aortic branch and in the aorta were recorded. Results: The mean amounts of air measured were 0.82±0.23 mL in the zone-2 group and 0.94±0.28 mL in the zone-3 group (p=0.49). In the zone-2 group compared with zone 3, the amounts of released air were greater in the right subclavian artery (0.07±0.02 vs 0.02±0.02 mL, p<0.01) and right CCA (0.30±0.8 vs 0.18±07 mL, p=0.04). There were no differences between the groups concerning the mean amounts of air measured in the right vertebral and all left-side supra-aortic branches. The amount of air released in the descending aorta was significantly higher in the zone-3 group vs the zone-2 group (0.48±0.12 vs 0.13±0.08 mL, p<0.01). Small bubbles were observed continuously during deployment, whereas large bubbles appeared more commonly during deployment of the proximal stent-graft end and after proximal release of the stent-graft. Conclusion: Air is released into all supra-aortic branches and the descending aorta during deployment of tubular thoracic stent-grafts in zones 2 and 3 in an aortic flow model. Higher amounts of air were observed in right-side supra-aortic branches during deployment in zone 2, whereas significantly greater amounts of air were observed in the descending aorta during deployment in zone 3.
引用
- Makaloski V, Rohlffs F, Trepte C, et al. Distribution of Air Embolization During TEVAR Depends on Landing Zone: Insights From a Pulsatile Flow Model. J Endovasc Ther. 2019. PMID: 31088321. doi:10.1177/1526602819849931
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。