Endurant与Excluder支架长期结果对比

Long-term results after standard endovascular aneurysm repair with the Endurant and Excluder stent grafts.
PubMed 文献 · J Vasc Surg · 2019-05-27

单中心回顾性研究纳入277例标准EVAR患者(Endurant 156例,Excluder 121例),随访中位5.8年。

要点

  • 两种支架7年主要临床成功率和瘤颈相关事件无显著差异。
  • Endurant组瘤颈扩张更明显,且是肾功能显著下降的独立预测因子。
  • Endurant组更常用于宽瘤颈(>28mm占27.3% vs 1.7%)。

核心结论

  • 两种晚期支架均可获得持久稳定的结果。
  • Endurant支架与更高的瘤颈扩张率和更快的eGFR下降相关。
  • 仔细的术前规划和基于解剖的个体化器械选择比支架型号本身更重要。

研究设计与样本

  • 单中心回顾性研究,纳入2004年7月至2011年12月接受标准EVAR的完整退行性肾下腹主动脉瘤患者277例(Endurant 156例,Excluder 121例)。
  • 主要终点为主要临床成功(无需额外或二次外科/腔内干预的临床成功)。

临床怎么用

  • 两种支架均可用于标准EVAR,但Endurant在宽/成角瘤颈中应用更多且瘤颈扩张和肾功能下降风险较高。
  • 术前应仔细评估瘤颈解剖并个体化选择器械,而非仅依据支架型号。
  • 对植入Endurant的患者应加强随访监测瘤颈扩张和肾功能变化。

局限性

  • 单中心回顾性研究,选择偏倚可能存在。
  • 两组基线瘤颈解剖存在显著差异,可能影响结果比较。
  • 需要更长期随访研究确定瘤颈扩张和肾功能下降的临床相关性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Many endografts are currently available for standard endovascular repair of infrarenal abdominal aortic aneurysms. Comparison of long-term outcomes between devices might aid in this decision process, but comparative data are scarce. The purpose of this study was to report long-term clinical outcomes of two commercially available endoprosthesis, the Endurant (Medtronic Vascular, Inc, Minneapolis, Minn) and the Excluder (W. L. Gore & Associates, Flagstaff, Ariz) stent grafts. METHODS: Patients undergoing standard endovascular repair from July 2004 to December 2011 in a single institution with the Endurant or the Low-Porosity Excluder endografts were eligible. Only patients treated for intact degenerative abdominal infrarenal aneurysms were included. All measurements were performed on center-lumen line reconstructions obtained on dedicated software. The primary end point was primary clinical success, defined as clinical success without the need for an additional or secondary surgical or endovascular procedure. Neck-related events (a composite of type IA endoleak, neck-related secondary intervention, or migration of >5 mm), neck morphology changes, renal function, and overall survival were secondary end points. RESULTS: The study included 277 patients (156 Endurants; 121 Excluders). The median follow-up was 5.8 years (range, 0.1-12.4 years) and did not differ between groups (P = .18). Patients treated with the Endurant stent graft had wider (neck diameter of >28 mm, 27.3% vs 1.7% [P < .001]; neck diameter of 27 mm, [interquartile range (IQR), 24-29 mm] for Endurant and 24 mm [IQR, 22-25 mm] for Excluder; P < .001) and more angulated necks (β-angle of >60°, 26.7% vs 12.5%; P = .004). Oversizing was greater in the Endurant group (16% [IQR, 12%-22%] vs 13% [IQR, 8%-17%], respectively; P < .001). Patients were treated outside device instructions for use regarding proximal neck: 16.7% in the Endurant and 17.3% in the Excluder group (P = .720). The 7-year primary clinical success was 54.7% for the Endurant and 58.1% for the Excluder groups (P = .53). Freedom from neck-related events at 7 years was 76.7% for the Endurant and 78.8% for Excluder group (P = .94). The Endurant stent graft (hazard ratio [HR], 2.7; 95% confidence interval [CI], 1.3-5.8; P = .009) was an independent predictor of significant renal function decline. Neck dilatation was greater in Endurant-implanted patients (13% [95% CI, 2%-22%] vs 4% [95% CI, 0%-10%]; P < .001). Overall survival at 7 years was 61.4% in the Endurant and 50.3% (n = 50; standard error, 0.047) in the Excluder group (P = .39). CONCLUSIONS: This study reveals that durable and sustainable results can be obtained with either of these late generation devices. This finding suggests that careful planning and a tailored device selection taking into account the patient's anatomy are more relevant determinants than the graft model itself to obtain clinical success. The Endurant endoprosthesis seems to be associated with a higher rate of neck dilatation and faster decrease in the estimated glomerular filtration rate, but further studies with longer follow-up are necessary to determine the clinical relevance of these findings.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Oliveira-Pinto J, Oliveira NFG, Bastos-Gonçalves FM, et al. Long-term results after standard endovascular aneurysm repair with the Endurant and Excluder stent grafts. J Vasc Surg. 2019. PMID: 31147134. doi:10.1016/j.jvs.2019.03.039

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。