术中超声造影在腹主动脉瘤腔内修复术中早期诊断内漏的价值
CEUS检出25例内漏(42%),2DSA检出11例(18%),两者一致性为fair agreement。CEUS可检出更多内漏,尤其II型,并可辅助术中瘤囊栓塞。
要点
- 术中CEUS检测内漏的敏感性高于2DSA,尤其对II型内漏。
- CEUS与2DSA检测内漏的一致性仅为fair agreement。
- CEUS可辅助术中瘤囊栓塞,技术成功率高。
核心结论
- 术中CEUS在EVAR中可行,且能检出比2DSA更多的内漏,尤其是II型内漏。
- 在II型内漏中,CEUS可能作为术中瘤囊栓塞引导的辅助工具。
研究设计与样本
- 前瞻性研究,连续纳入2017年1月至2018年4月接受择期EVAR的患者。
- 支架释放后行2DSA(正位和侧位),随后在无菌条件下行术中CEUS。
- 血管外科医生对2DSA结果设盲进行CEUS评估。
主要发现
- 2DSA检出11例内漏(18%;1例IA型,10例II型),CEUS检出25例内漏(42%;2例IA型,23例II型)。
- 2DSA与CEUS在39例(65%)中结果一致(32例无内漏,7例II型内漏)。
- CEUS检出17例2DSA未发现的内漏(28%;2例IA型,15例II型)。
- 2DSA检出3例CEUS未发现的内漏(5%;3例II型)。
- 检测任何类型内漏和II型内漏的Cohen's kappa分别为0.255和0.250(均为fair agreement)。
临床怎么用
- 术中CEUS可作为EVAR术中检测内漏的辅助手段,尤其有助于发现II型内漏。
- 在II型内漏需行瘤囊栓塞时,CEUS可提供引导。
局限性
- 研究样本量有限,需进一步研究评估术中CEUS诊断的可靠性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The aim of this study was to evaluate the feasibility and utility of intraoperative contrast-enhanced ultrasound (CEUS) for early detection of endoleaks (ELs) during endovascular abdominal aortic aneurysm repair (EVAR) compared with completion digital subtraction angiography. METHODS: Patients undergoing elective EVAR from January 2017 to April 2018 were consecutively enrolled in this prospective study. After endograft deployment, two-digital subtraction angiography (2DSA) with orthogonal C-arm angulations (anteroposterior and sagittal view) were routinely performed. After the endovascular treatment of clear, high-flow type I/III ELs detected by 2DSA, intraoperative CEUS was carried out in sterile conditions on the surgical field before guidewire removal. Presence and type of EL were evaluated with 2DSA and CEUS. CEUS was performed with the vascular surgeon blinded to the 2DSA findings. The primary end point was the level of agreement between 2DSA and CEUS to detect any type of EL and type II EL. Agreement between two diagnostic methods was calculated using Cohen's kappa. The secondary end point was utility of CEUS for intraoperative adjunctive procedure guidance. RESULTS: Sixty patients were enrolled (mean age, 78 ± 6 years; 90% male). 2DSA revealed 11 ELs (18%; 1 type IA, 10 type II), and CEUS 25 ELs (42%; 2 type IA, 23 type II). 2DSA and CEUS were in agreement in 39 cases (65%; 32 no ELs, 7 type II ELs). CEUS detected 17 ELs not identified by 2DSA (28%; 2 type IA, 15 type II); 2DSA detected three ELs not identified by CEUS (5%; 3 type II). In one case, 2DSA and CEUS detected type II and type IA ELs, respectively. For EL and type II EL detection, Cohen's kappa was 0.255 and 0.250, respectively (both "fair agreement"). Intraoperative adjunctive sac embolization was performed under CEUS control in 4 cases and technical success was 100%. CONCLUSIONS: Intraoperative CEUS during EVAR is feasible and can detect a greater number of ELs than 2DSA, in particular type II ELs. Further studies are necessary to assess the reliability of this intraoperative diagnostic examination. In type II ELs, CEUS may represent an additional, useful tool for intraoperative sac embolization guidance.
引用
- Bianchini Massoni C, Perini P, Fanelli M, et al. Intraoperative contrast-enhanced ultrasound for early diagnosis of endoleaks during endovascular abdominal aortic aneurysm repair. J Vasc Surg. 2019. PMID: 31147132. doi:10.1016/j.jvs.2019.02.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。