文献速递主动脉疾病

EndoAnchor植入物治疗EVAR后Ia型内漏的持久性

Sustainability of Individual EndoAnchor Implants in Therapeutic Use to Treat Type Ia Endoleak After Endovascular Aneurysm Repair.
PubMed 文献 · J Endovasc Ther · 2019-03-25

本研究基于ANCHOR注册研究,纳入54例患者共360枚EndoAnchor植入物,评估用于治疗EVAR术中或迟发Ia型内漏后初始穿透良好的植入物在随访中的穿透深度和角度变化。结果显示,首次术后CTA上187枚穿透良好,末次CTA上97.4%维持良好,仅2.6%变为临界或无穿透,均无临床后果。

要点

  • 初始穿透良好的EndoAnchor植入物在约1年随访中97.4%维持良好位置。
  • 仅2.6%的植入物变为临界或无穿透,无临床后果。
  • 纵向角度中位数从85°变为84°(p=0.043),正交角度无显著变化。

核心结论

  • 初始穿透良好的EndoAnchor植入物在1年随访时持久性令人满意,绝大多数维持良好位置,低于3%变为临界或无穿透且无临床后果。

研究设计与样本

  • 基于ANCHOR全球注册研究(NCT01534819)的回顾性分析,纳入54例患者(44例男性)共360枚EndoAnchor植入物,入选标准为用于治疗术中或迟发Ia型内漏且至少2次术后CTA,排除使用辅助手术者。

主要发现

  • 首次术后CTA上EndoAnchor穿透良好187枚(51.9%)、临界69枚(19.2%)、缺失104枚(28.9%)。
  • 末次CTA上187枚初始良好者中182枚(97.4%)维持良好,5枚(2.6%)在4例患者中构型改变(4枚变临界、1枚无穿透),均无临床后果。
  • 初始和末次CTA上穿透良好植入物的正交角度中位数分别为92°(IQR 85,98)和90°(IQR 84,97),p=0.822;纵向角度中位数分别为85°(IQR 71,96)和84°(IQR 70,96),p=0.043。
  • 首次术后CTA有18例(33%)患者存在Ia型内漏,其中6例随时间缓解;中位随访13个月,未发现新发Ia型内漏。

临床怎么用

  • 对于EVAR后Ia型内漏接受EndoAnchor治疗且初始穿透良好的患者,中期随访中植入物位置可持续,支持该疗法的持久性。

局限性

  • 作者指出分析的EndoAnchor数量较少。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate changes in penetration depths and angles of EndoAnchor implants with initially good penetration after therapeutic use in endovascular aneurysm repair. MATERIALS AND METHODS: Patients were selected from the Aneurysm Treatment Using the Heli-FX Aortic Securement System Global Registry (ANCHOR; ClinicalTrials.gov identifier NCT01534819). Inclusion criteria were (1) EndoAnchor implantation to treat intraoperative or late type Ia endoleak and (2) at least 2 postoperative computed tomography angiography (CTA) scans. Exclusion criteria were the use of adjunct procedures. Based on these criteria, 54 patients (44 men) with 360 EndoAnchor implants were eligible for this analysis. Penetration depth of each EndoAnchor implant into the aortic wall was judged as (1) good (≥2-mm penetration), (2) borderline (<2 mm or when there was a gap between the endograft and the aortic wall), or (3) no penetration. The penetration depth and longitudinal angles of EndoAnchors with good penetration were investigated on the last available postprocedure CTA scan. Endoleaks were also analyzed. RESULTS: EndoAnchor penetration on the first postprocedure CTA scan was good in 187 (51.9%), borderline in 69 (19.2%), and missing in 104 (28.9%). On the last CTA scan, 182 (97.4%) of the 187 initially well-positioned EndoAnchors remained good. Five (2.6%) EndoAnchors in 4 patients changed configuration over time (4 became borderline and 1 became nonpenetrating), all without any clinical sequelae. The median orthogonal angles of the EndoAnchor implants with good penetration on the first and last CTA scans were 92° [interquartile range (IQR) 85, 98] and 90° (IQR 84, 97), respectively (p=0.822); for longitudinal angles, medians of 85° (IQR 71, 96) and 84° (IQR 70, 96) were found (p=0.043). Of the 18 (33%) patients who had a type Ia endoleak on the first postprocedure CTA, 6 resolved over time. Median follow-up was 13 months, during which no new type Ia endoleak was found. CONCLUSION: Despite the small number of EndoAnchors analyzed, this study showed that the sustainability of EndoAnchor implants with initially good penetration is satisfactory at 1-year follow-up. The vast majority of EndoAnchor implants with good penetration initially remained in good position; <3% of implants became borderline or nonpenetrating, without any clinical consequence.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. van Noort K, Vermeulen JJM, Goudeketting SR, et al. Sustainability of Individual EndoAnchor Implants in Therapeutic Use to Treat Type Ia Endoleak After Endovascular Aneurysm Repair. J Endovasc Ther. 2019. PMID: 30907259. doi:10.1177/1526602819837753

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。