新型机械分级间歇序贯加压装置增强静脉血流:一项初步研究
主要终点为腘静脉和股静脉峰值静脉流速。结果显示WIC低档和高档的峰值流速均高于基线及商用IPC装置(P < .05),且患者无不适。
要点
- WIC装置低档和高档的峰值静脉流速均显著高于基线及IPC装置(P < .05)。
- WIC装置在CEAP 3–6级静脉功能不全患者中耐受性良好,无患者报告不适。
- WIC装置可能克服传统IPC的便携性和依从性限制,但需进一步研究验证溃疡愈合获益。
核心结论
- WIC装置在CEAP 3–6级静脉功能不全患者中显著增强腘静脉和股静脉的峰值静脉流速,且使用方便、舒适。
主要发现
- WIC装置低档和高档的腘静脉及股静脉平均峰值静脉流速均高于基线(P < .05)。
- 在5例患者中,WIC装置低档和高档的峰值静脉流速均高于商用IPC装置(P < .05)。
- 无患者报告WIC装置或IPC装置治疗期间的不适。
关键数值与推荐
- 主要终点为峰值静脉流速(厘米/秒),测量部位为腘静脉和股静脉。
- 测量条件包括基线、WIC低档和WIC高档;5例患者额外测量商用IPC装置。
临床怎么用
- WIC装置可作为CEAP 3–6级静脉功能不全患者增强静脉血流的潜在选择,尤其适用于传统IPC不耐受或依从性差者。
- 需进一步研究确定WIC改善静脉血流和依从性是否转化为更高的静脉溃疡愈合率。
局限性
- 未报告溃疡愈合等临床结局,随访时间未提及。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Guidelines as well as multiple RTCs support the use of intermittent pneumatic compression (IPC) for the treatment of venous leg ulcers when conservative measures fail. Unfortunately, the clinical usefulness of IPC is significantly limited by the physical limitations of pneumatic motors, which leads to bulky devices with slow inflation cycles, uncomfortable sleeves, lack of patient mobility, and ultimately poor patient compliance with therapy. A novel mechanical device for lower leg graded intermittent sequential compression was designed to address these limitations of IPC therapy for venous leg ulcer treatment by providing rapid compression cycles in a truly wearable device that offers the additional benefit of monitoring compression dose and patient compliance. The wearable intermittent compression (WIC) device was hypothesized to provide improved augmentation of venous flow compared with both baseline and standard IPC therapy. METHODS: Ten patients with Clinical, Etiologic, Anatomic and Pathophysiologic class 3 to 6 venous insufficiency were recruited under institutional review board approval. The primary end point for the study was augmentation of venous blood flow as measured by peak venous velocity. Patients underwent measurement of peak venous velocity in centimeters per second at the popliteal and femoral veins for the following conditions: (1) baseline, (2) WIC device on a low setting, and (3) WIC device on a high setting. In five patients, an additional measurement of peak venous velocity in centimeters per second at the popliteal and femoral veins was completed while wearing a commercially available IPC device. RESULTS: Both low and high settings of the WIC device resulted in higher average peak venous velocities when compared with both baseline and the IPC device (P < .05). No patients reported discomfort with either the WIC device or the IPC device during therapy. CONCLUSIONS: The WIC device significantly increases the augmentation of venous flow as measured by peak venous velocity in both the popliteal and femoral veins in patients with Clinical, Etiologic, Anatomic and Pathophysiologic class 3 to 6 venous insufficiency. In addition, the WIC device was found to be easy to use and comfortable during therapy. Future studies are planned to determine if the WIC improvements in venous flow augmentation and patient compliance will lead to higher rates of venous ulcer healing.
引用
- Wall J, Johnson E, Johnson B, et al. A pilot study of venous flow augmentation using a novel mechanical graded intermittent sequential compression device for venous insufficiency. J Vasc Surg Venous Lymphat Disord. 2019. PMID: 30612969. doi:10.1016/j.jvsv.2018.10.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。