文献速递主动脉疾病

Zenith开窗主动脉修复术后需积极监测内漏与组件间分离

Aggressive Surveillance Is Needed to Detect Endoleaks and Junctional Separation between Device Components after Zenith Fenestrated Aortic Reconstruction.
PubMed 文献 · Ann Vasc Surg · 2019-01-24

本文回顾性分析了单中心110例Cook Zenith开窗支架(ZFEN)植入病例,平均随访1.5年,临床显著组件间分离发生率为4.5%,IIIa型内漏为2.7%。分离均与Ib型内漏并存,平均诊断时间40个月,部分病例进展至破裂或主动脉相关死亡,故建议术中最大化支架重叠并积极监测。

要点

  • ZFEN术后组件间分离可导致IIIa型内漏、瘤囊扩大甚至破裂。
  • 平均诊断时间40个月,需长期积极影像监测以早期发现支架重叠丢失。
  • 术中应最大化开窗主体与远端分叉支架的主动脉内重叠长度。

核心结论

  • ZFEN平台存在开窗主体与远端分叉支架间分离并继发IIIa型内漏的既定并发症,需术中最大化重叠并积极监测与治疗。

研究设计与样本

  • 回顾性分析单中心前瞻性维护的ZFEN开窗EVAR数据库,纳入2012年10月至2017年12月植入的110例ZFEN,平均随访1.5年。

主要发现

  • 临床显著组件间分离发生率为4.5%,IIIa型内漏发生率为2.7%。
  • 5例分离患者均并存Ib型内漏,3例出现瘤囊扩大,1例无显著内漏且瘤囊缩小。
  • 组件间分离平均诊断时间为40个月;2例监测中发现并再干预,3例完成分离,其中2例破裂急诊干预、1例主动脉相关死亡。

关键数值与推荐

  • 临床显著组件间分离发生率4.5%,IIIa型内漏发生率2.7%,平均随访1.5年,分离平均诊断时间40个月。

临床怎么用

  • 术中应最大化开窗主体与远端分叉支架的主动脉重叠,术后积极影像监测支架重叠丢失并及时治疗。

局限性

  • 为单中心回顾性研究,样本量110例,平均随访仅1.5年,可能低估晚期分离发生率。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Junctional separation and resulting type IIIa endoleak is a well-known problem after EVAR (endovascular aneurysm repair). This complication results in sac pressurization, enlargement, and eventual rupture. In this manuscript, we review the incidence of this late finding in our experience with the Cook Zenith fenestrated endoprosthesis (ZFEN, Bloomington, IN). METHODS: A retrospective review was performed of a prospectively maintained institutional ZFEN fenestrated EVAR database capturing all ZFENs implanted at a large-volume, academic hospital system. Patients who experienced junctional separation between the fenestrated main body and distal bifurcated graft (with or without type IIIa endoleak) at any time after initial endoprosthesis implantation were subject to further evaluation of imaging and medical records to abstract clinical courses. RESULTS: In 110 ZFENs implanted from October 2012 to December 2017 followed for a mean of 1.5 years, we observed a 4.5% and 2.7% incidence of clinically significant junctional separation and type IIIa endoleak, respectively. Junctional separation was directly related to concurrent type Ib endoleak in all 5 patients. Three patients presented with sac enlargement. One patient did not demonstrate any evidence of clinically significant endoleak and had a decreasing sac size during follow-up imaging. The mean time to diagnosis of modular separation in these patients was 40 months. Junctional separation was captured in surveillance in 2 patients and reintervened upon before manifestation of endoleak. However, the remaining 3 patients completed modular separation resulting in rupture and emergent intervention in 2 and an aortic-related mortality in the other. CONCLUSIONS: Junctional separation between the fenestrated main and distal bifurcated body with the potential for type IIIa endoleak is an established complication associated with the ZFEN platform. Therefore, we advocate for maximizing aortic overlap during the index procedure followed by aggressive surveillance and treatment of stent overlap loss captured on imaging.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Wang SK, Lemmon GW, Gupta AK, et al. Aggressive Surveillance Is Needed to Detect Endoleaks and Junctional Separation between Device Components after Zenith Fenestrated Aortic Reconstruction. Ann Vasc Surg. 2019. PMID: 30684629. doi:10.1016/j.avsg.2018.09.038

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。