文献速递主动脉疾病

大直径(34-36mm)移植物行肾下EVAR与近端固定失败风险升高相关

Infrarenal endovascular aneurysm repair with large device (34- to 36-mm) diameters is associated with higher risk of proximal fixation failure.
PubMed 文献 · J Vasc Surg · 2018-06-28

单中心回顾性研究纳入2000-2016年500例择期肾下腹主动脉瘤EVAR患者,108例(21.6%)使用34-36mm大直径移植物。大直径组近端固定失败率24%显著高于常规直径组6.1%(P<.001),多因素分析显示Talent移植物和瘤颈直径≥29mm是独立危险因素。

要点

  • 大直径(34-36mm)移植物组近端固定失败率24%,显著高于常规直径组6.1%。
  • 瘤颈直径≥29mm是近端固定失败的独立危险因素(OR 2.50)。
  • 大直径瘤颈患者应考虑开窗/分支移植物或开放手术等更近端密封策略。

核心结论

  • 肾下瘤颈≥29mm需34-36mm直径移植物的标准EVAR与近端固定失败率升高独立相关。

研究设计与样本

  • 单中心前瞻性维护数据库的回顾性综述,纳入2000至2016年择期肾下腹主动脉瘤EVAR患者500例。
  • 仅纳入可选34-36mm近端直径的移植物,按装置直径分为大直径组(34-36mm,108例)和常规直径组(<34mm,392例)。
  • 主要结局为近端固定失败,定义为EVAR后IA型内漏和支架移植物移位>10mm的复合终点。

主要发现

  • 大直径组与常规直径组30天死亡率无差异(0.9% vs 0.95%;P=.960),1年死亡率亦无差异(9.0% vs 6.2%;P=.920)。
  • 近端固定失败在大直径组发生26/108例(24%),常规直径组24/392例(6.1%)(P<.001)。
  • IA型内漏在大直径组16例(14.8%),常规直径组13例(3.3%)(P<.001)。
  • 支架移植物移位>10mm在大直径组16例(14.8%),常规直径组15例(3.8%)(P<.001)。
  • 多因素分析中,Talent移植物(OR 4.50;95%CI 1.18-17.21)和瘤颈直径≥29mm(OR 2.50;95%CI 1.12-5.08)为近端固定失败的独立危险因素。

关键数值与推荐

  • 近端固定失败定义为IA型内漏和支架移植物移位>10mm的复合终点。
  • 瘤颈直径≥29mm为近端固定失败的独立危险因素(OR 2.50;95%CI 1.12-5.08)。
  • 使用Talent移植物为近端固定失败的独立危险因素(OR 4.50;95%CI 1.18-17.21)。

临床怎么用

  • 对于肾下瘤颈≥29mm需34-36mm直径移植物的患者,应考虑开窗或分支移植物等更近端密封策略而非标准EVAR。
  • 此类患者可能需要更严格的影像学随访以监测近端固定失败。

局限性

  • 单中心回顾性研究,样本来自单一机构。
  • 仅纳入可选34-36mm近端直径的移植物,可能存在选择偏倚。
  • 原文未报告其他局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Endovascular aneurysm repair (EVAR) has become the standard of care for infrarenal aneurysms. Endografts are commercially available in proximal diameters up to 36 mm, allowing proximal seal in necks up to 32 mm. We sought to further investigate clinical outcomes after standard EVAR in patients requiring large main body devices. METHODS: We performed a retrospective review of a prospectively maintained database for all patients undergoing elective EVAR for infrarenal abdominal aortic aneurysms at a single institution from 2000 to 2016. Only endografts with the option of a 34- to 36-mm proximal diameter were included. Requisite patient demographics, anatomic and device-related variables, and relevant clinical outcomes and imaging were reviewed. The primary outcome in this study was proximal fixation failure, which was a composite of type IA endoleak and stent graft migration >10 mm after EVAR. Outcomes were stratified by device diameter for the large-diameter device cohort (34-36 mm) and the normal-diameter device cohort (<34 mm). RESULTS: There were 500 patients treated with EVAR who met the inclusion criteria. A total of 108 (21.6%) patients received large-diameter devices. There was no difference between the large-diameter cohort and the normal-diameter cohort in terms of 30-day (0.9% vs 0.95%; P = .960) or 1-year mortality (9.0% vs 6.2%; P = .920). Proximal fixation failure occurred in 24 of 392 (6.1%) patients in the normal-diameter cohort and 26 of 108 (24%) patients in the large-diameter cohort (P < .001). There were 13 (3.3%) type IA endoleaks in the normal-diameter cohort and 16 (14.8%) in the large-diameter cohort (P < .001). Stent graft migration (>10 mm) occurred in 15 (3.8%) in the normal-diameter cohort and 16 (14.8%) in the large-diameter cohort (P < .001). After multivariate analysis, only the use of Talent (Medtronic, Minneapolis, Minn) endografts (odds ratio [OR], 4.50; 95% confidence interval [CI], 1.18-17.21) and neck diameter ≥29 mm (OR, 2.50; 95% CI, 1.12-5.08) remained significant independent risk factors for development of proximal fixation failure (OR, 3.99; 95% CI, 1.75-9.11). CONCLUSIONS: Standard EVAR in patients with large infrarenal necks ≥29 mm requiring a 34- to 36-mm-diameter endograft is independently associated with an increased rate of proximal fixation failure. This group of patients should be considered for more proximal seal strategies with fenestrated or branched devices vs open repair. Also, this group likely needs more stringent radiographic follow-up.