编辑精选——Nellix支架系统腔内动脉瘤封堵术后中期迁移与器械失败:单中心经验
该单中心回顾性综述纳入161例接受Nellix腔内动脉瘤封堵术的患者,其中原发腹主动脉瘤115例;原发腹主动脉瘤患者4年无器械失败率为42%,失败多发生于植入2年后,共发生8例继发性瘤囊破裂和14例移植物取出。作者认为Nellix封堵失败较预期更常见并可导致主动脉破裂,术后监测至关重要。
要点
- 原发腹主动脉瘤患者接受Nellix腔内动脉瘤封堵术后4年无器械失败率为42%。
- 器械失败多发生在Nellix植入术后2年以上。
- 共发生8例继发性瘤囊破裂(发病率2.4/100人年)和14例移植物取出。
核心结论
- Nellix腔内动脉瘤封堵术后封堵失败较预期更常见,并可导致主动脉破裂。
- Nellix支架移植术后监测对于识别失败特征至关重要。
研究设计与样本
- 剑桥大学医院基金会信托所有接受Nellix治疗患者的回顾性综述。
- 共161例患者接受Nellix治疗,适应证包括原发腹主动脉瘤115例、破裂腹主动脉瘤4例、其他主动脉移植物挽救18例、原发髂动脉瘤6例、肾旁腹主动脉瘤烟囱腔内动脉瘤封堵18例。
- 器械失败定义为继发性瘤囊破裂、移植物取出、针对1型内漏的进一步外科手术或支架移植物大迁移伴主动脉瘤囊内压力增高。
主要发现
- 原发腹主动脉瘤患者共发生42例移植物失败。
- 原发腹主动脉瘤患者4年无移植物失败率为42%。
- 失败大多发生在Nellix植入术后2年以上。
- 共发生8例继发性瘤囊破裂,发病率为2.4/100人年。
- 共发生14例移植物取出。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Endovascular aneurysm sealing (EVAS) with the Nellix stent graft system is a novel concept in the management of abdominal aortic aneurysm (AAA) that aims to reduce the prevalence of all endoleaks following endovascular repair. There are few data describing the longer-term durability of this approach. The aim was to report the longer-term outcomes following EVAS in a single centre. METHODS: This is a retrospective review of all patients that underwent Nellix at Cambridge University Hospitals Foundation Trust. Factors that are described as device failure include secondary sac rupture, graft explantation, further surgical procedures for Type 1 endoleak, or major migration of the stent grafts with pressurisation of the aortic sac. RESULTS: A total of 161 patients have been treated with Nellix. The indications included primary AAA (n = 115), ruptured AAA (n = 4), salvage of other aortic grafts (n = 18), primary iliac aneurysm (n = 6), and chimney EVAS (ChEVAS) for pararenal AAA (n = 18). In total there have been 42 graft failures in patients treated with EVAS for primary AAA. The 4 year freedom from graft failure was 42% in patients treated for primary AAA. Failures mostly occurred more than 2 years post-Nellix implant. There were eight secondary sac ruptures (incidence 2.4 per 100 person years) and there have been 14 graft explants. CONCLUSIONS: Failure of aneurysm sealing following treatment with Nellix has been more common than anticipated and can cause aortic rupture. Post-operative surveillance of Nellix stent grafts is crucial to identify features of failure.
引用
- Harrison SC, Winterbottom AJ, Coughlin PA, et al. Editor's Choice - Mid-term Migration and Device Failure Following Endovascular Aneurysm Sealing with the Nellix Stent Graft System - a Single Centre Experience. Eur J Vasc Endovasc Surg. 2018. PMID: 30190039. doi:10.1016/j.ejvs.2018.06.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。