文献速递主动脉疾病

体外评估主动脉覆膜支架在远端锚定区的释放准确性

In Vitro Evaluation of Aortic Stent Graft Deployment Accuracy in the Distal Landing Zone.
PubMed 文献 · Eur J Vasc Endovasc Surg · 2018-09-07

该体外研究将三种胸主动脉覆膜支架以三种方式植入两例患者CT血管造影重建的1:1 3D打印主动脉(直型与迂曲型),评估远端锚定区释放准确性。反向释放(由远及近)距目标血管为0 mm(IQR 0;0),优于近端锚定于动脉瘤的3 mm(IQR 0;8)和近端锚定于另一支架的2 mm(IQR 0;5);迂曲主动脉楔形贴壁为18 mm(IQR 15;20),直型为0 mm(IQR 0;0)(p < .001)。结论认为反向释放与更准确的远端锚定区释放相关,远端主动脉迂曲是覆膜支架覆盖远端锚定区全周径的重要障碍。

要点

  • 反向释放距目标血管0 mm(IQR 0;0),近端锚定于动脉瘤为3 mm(IQR 0;8),近端锚定于另一支架为2 mm(IQR 0;5)。
  • 距目标血管测量达5 mm的发生率分别为45%、30%和0%。
  • 迂曲主动脉楔形贴壁18 mm(IQR 15;20),直型为0 mm(IQR 0;0)(p < .001)。

核心结论

  • 反向释放覆膜支架与更准确的远端锚定区释放相关。
  • 远端主动脉迂曲是覆膜支架覆盖远端锚定区全周径的重要障碍。

研究设计与样本

  • 体外研究,基于两例患者CT血管造影扫描以1:1比例3D打印两具主动脉:直型与迂曲型,远端主动脉迂曲度分别为1.006和1.078。
  • 对三种覆膜支架(E-vita THORACIC 3G、Relay Plus、Valiant Captivia)各以三种方式植入10次:近端锚定于动脉瘤内、近端锚定于另一覆膜支架内、反向释放(经模拟顺行入路)。
  • 目标为释放至目标血管上缘正上方,直视下用卡尺评估距目标血管距离与楔形贴壁。

主要发现

  • 距目标血管距离:近端锚定于动脉瘤为3 mm(IQR 0;8),近端锚定于另一支架为2 mm(IQR 0;5),反向释放为0 mm(IQR 0;0)。
  • 距目标血管测量达5 mm的发生率分别为45%、30%和0%。
  • 覆膜支架植入后总体楔形贴壁:直型主动脉为0 mm(IQR 0;0),迂曲主动脉为18 mm(IQR 15;20)(p < .001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Thoracic endovascular aortic repair technology focuses on accurately deploying the stent graft in the proximal landing zone in the proximal to distal direction. The aim of this in vitro study was to evaluate the accuracy of stent graft deployment in the distal landing zone. METHODS: It was hypothesised that a reverse implantation mechanism (in a distal to proximal deployment direction, when the operator starts to open the endoprosthesis from distal to proximal), might enhance landing accuracy in the distal LZ. The aim was to investigate this hypothesis by implanting stent grafts into the 3D printed aortas with the currently available deployment mechanism. Based on two human patients' computed tomography angiography scans, two aortas were 3D printed at 1:1 scale: "straight" and "crooked" aortas with distal aortic tortuosity of 1.006 and 1.078, respectively. They were used in order to test three endoprostheses (E-vita THORACIC 3G, Relay Plus, Valiant Captivia) 10 times by implanting them in three ways: proximal landing in the aneurysm, proximal landing in another stent graft, reverse implantation (via simulated antegrade access). The aim was to land just above the target vessel's upper edge. The distance to the target vessel and wedge apposition were assessed under a direct view using caliper. RESULTS: The distance to the target vessel was 3 mm (IQR 0; 8) if the stent graft landed proximally in aneurysm, 2 mm (IQR 0; 5) if it landed proximally in another stent graft, and 0 mm (IQR 0; 0) when reverse implantation was applied. The distance to the target vessel measuring 5 mm or occurred in 45%, 30%, and 0%, respectively. Overall the median wedge apposition after stent graft implantation was 0 mm (IQR 0; 0) in the "straight" versus 18 mm (IQR 15; 20) in the "crooked" aorta (p < .001). CONCLUSIONS: Reverse stent graft deployment is associated with more accurate landing in the distal landing zone. Distal aortic tortuosity constitutes an important impediment to covering the distal LZ's entire circumference with a stent graft.

来源:Eur J Vasc Endovasc Surg,查看原文。版权归原作者所有。

引用

  1. Berezowski M, Kondov S, Beyersdorf F, et al. In Vitro Evaluation of Aortic Stent Graft Deployment Accuracy in the Distal Landing Zone. Eur J Vasc Endovasc Surg. 2018. PMID: 30197286. doi:10.1016/j.ejvs.2018.07.034

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。