急诊病例中t-Branch现成多分支内移植物部署后的准确方向评估

Accurate Orientation of the t-Branch Off-the-Shelf Multibranched Endograft After Deployment in Urgent Cases.
PubMed 文献 · J Endovasc Ther · 2018-06-05

本研究回顾性分析42例急诊胸腹主动脉瘤患者置入t-Branch后的方向偏差,技术成功率为93%(39/42),7例(17%)至少一根目标血管方向偏差,女性与腹腔干和肠系膜上动脉旋转偏差增加相关,t-Branch对较大方向偏差表现出「宽容」特性。

要点

  • 急诊t-Branch置入技术成功率为93%(39/42)。
  • 7例(17%)患者术后至少一根目标血管方向偏差,未影响手术时间、目标血管血运重建或早期分支通畅率。
  • 女性与腹腔干和肠系膜上动脉术后旋转偏差增加相关,也与各目标血管分支起点与标准构型间旋转偏差增加相关。

核心结论

  • 急诊t-Branch置入实现了目标血管与内移植物分支位置的准确对位,无旋转偏差。
  • 性别可能影响装置的方向。
  • t-Branch对较大方向偏差表现出「宽容」特性,不影响手术时间、目标血管血运重建或早期分支通畅率。

研究设计与样本

  • 回顾性分析2014年1月至2017年6月连续42例急诊胸腹主动脉瘤患者(平均年龄73.3±7岁;25例男性),均置入t-Branch。
  • 主要目标为量化目标血管术前钟面位置,并确定术前与术后测量之间以及术后测量与标准分支构型之间的旋转偏差。
  • 次要目标为识别影响方向偏差(>2个钟面小时)且可能影响结局的因素。
  • 比较早期(学习曲线;n=18)与晚期(n=24)经验,以及遵守(n=23)与不遵守(n=19)使用说明(IFU)的结果。

主要发现

  • 技术成功率为93%(39/42)。
  • 遵守IFU与不遵守IFU的患者之间旋转偏差无显著差异。
  • 7例(17%)患者术后至少一根目标血管方向偏差,与总手术时间、透视时间、对比剂体积、辐射剂量或遵守IFU无关。
  • 女性与腹腔干(p=0.044)和肠系膜上动脉(SMA;p=0.006)术后测量旋转偏差增加相关。
  • 女性还与部署后分支起点与t-Branch标准构型之间各目标血管旋转偏差增加相关:腹腔干(p=0.005)、SMA(p=0.001)、右肾动脉(p=0.037)、左肾动脉(p=0.003)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate the orientation of the standardized off-the-shelf multibranched t-Branch after implantation in urgent thoracoabdominal aortic aneurysm (TAAA) repairs, to characterize the impact of branch malorientation on procedural success, and to identify any predictive factors associated with malorientation. METHODS: A retrospective analysis was conducted of 42 consecutive patients (mean age 73.3±7 years; 25 men) with urgent TAAA presentation treated with the t-Branch from January 2014 to June 2017. The primary objective was to quantify the preoperative clock position of the target vessels and determine any rotational deviation between the pre- and postoperative measurements and between the postoperative measurements and the standard branch configuration. Secondary outcomes were to identify factors influencing malorientation (>2 clock face hours) that could affect outcome. Results were compared for early (learning curve; n=18) vs late (n=24) experience and for adherence to (n=23) vs nonobservance of (n=19) the instructions for use (IFU). RESULTS: Technical success was 93% (39/42). No significant difference in rotational deviation was identified between patients treated within the IFU vs outside the IFU. Seven (17%) patients had at least one target vessel maloriented after the procedure, which was not associated with total procedure time, fluoroscopy time, contrast volume, radiation dose, or adherence to the IFU. Female gender was associated with increased rotational deviation in postprocedure measurement for the celiac trunk (p=0.044) and superior mesenteric artery (SMA; p=0.006). Female gender was also associated with increased rotational deviation between the branch origin after deployment and the standard configuration of the t-Branch for every target vessel [celiac trunk (p=0.005), SMA (p=0.001), right renal artery (p=0.037), and left renal artery (p=0.003)]. CONCLUSION: The implantation of the t-Branch device in urgent cases achieved accurate apposition without rotational deviation between the target vessels and the position of the endograft branches. Gender may have an impact on orientation of the device. The t-Branch appears to have a "forgiving" nature for higher malorientation with no effect on procedure time, target vessel revascularization, or early branch patency.

来源:J Endovasc Ther,查看原文。版权归原作者所有。