胸主动脉腔内修复术中空气栓塞:输送系统推送器上的附加冲洗端口在实验环境中显著减少胸主动脉覆膜支架释放时排出的空气量
实验研究将20枚带附加冲洗端口的胸主动脉覆膜支架分为C、D两组,D组在生理盐水冲洗前额外用二氧化碳冲洗;C组释放气体平均0.51 mL,D组0.07 mL(p<0.001),C组也显著低于无附加端口的参考A组0.79 mL(p=0.002)。结论认为附加侧端口可显著减少释放空气,联合二氧化碳冲洗进一步减少气体释放。
要点
- 带附加冲洗端口的C组释放气体平均0.51 mL,D组联合二氧化碳冲洗为0.07 mL(p<0.001)。
- C组释放气体平均0.51 mL,显著低于无附加端口参考A组的0.79 mL(p=0.002)。
- 胸主动脉覆膜支架释放时会排出空气,附加侧端口可显著减少排出量。
核心结论
- 胸主动脉覆膜支架在释放过程中会排出空气,使用附加侧端口减少推送器内部空气填充空间可显著减少释放的空气量。
- 附加侧端口联合二氧化碳冲洗技术可将气体释放进一步降至小体积。
- 在临床环境中,这可能是降低胸主动脉腔内修复术中空气栓塞和卒中风险的有前景的方法。
研究设计与样本
- 实验研究:20枚带附加冲洗端口的胸主动脉覆膜支架被分为C和D两个等量组。
- 两组均经连接推送器的额外侧端口用20 mL生理盐水冲洗,并经常规冲洗端口用60 mL生理盐水冲洗。
- D组在生理盐水冲洗前额外经常规冲洗端口用二氧化碳冲洗。
- 所有支架均释放入连接于充水容器底部的弯曲塑料管中,使用校准装置记录并测量释放的气体。
- C组与既往发表的、无附加侧端口且仅用标准60 mL生理盐水冲洗的参考A组进行比较。
主要发现
- 支架释放时排出的气体量各不相同,C组平均释放气体0.51 mL,D组0.07 mL(p<0.001)。
- C组平均释放气体0.51 mL,显著低于参考A组的0.79 mL(p=0.002)。
关键数值与推荐
- C组平均释放气体0.51 mL vs D组0.07 mL(p<0.001)。
- C组平均释放气体0.51 mL vs 参考A组0.79 mL(p=0.002)。
临床怎么用
- 在临床环境中,使用附加侧端口联合二氧化碳冲洗技术可能是降低胸主动脉腔内修复术中空气栓塞和卒中风险的有前景的方法。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To investigate the influence of (1) an additional side port for flushing the hollow pusher in Zenith thoracic stent-graft delivery systems and (2) additional carbon dioxide flushing on the amount of air released during stent-graft deployment. METHODS: Twenty thoracic stent-grafts with an additional flush port to fill the hollow pusher were separated into 2 equal groups (C and D). Both groups were flushed with 20 mL of normal saline through the extra side port connected to the pusher and with 60 mL of saline through the regular flushing port. One group of grafts (group D) was additionally flushed with carbon dioxide through the regular flushing port prior to saline. All grafts were deployed into a curved plastic pipe attached to the bottom of a water-filled container. The released gas was recorded and measured using a calibrated setup. To evaluate the influence of the extra side port irrespective of the carbon dioxide flushing technique, group C was compared with a previously published reference group A without an extra side port that was flushed with the standard 60 mL of saline. RESULTS: Volumes of gas were released in various amounts from the stent-grafts during deployment. The average amount of released gas was 0.51 mL in group C and 0.07 mL in group D (p<0.001). The mean amount of gas from group C samples (0.51 mL) was also significantly lower (p=0.002) compared with the reference group (0.79 mL). CONCLUSION: Thoracic endografts release air during deployment. Reducing the air-filled space inside the pusher of the catheter assembly using an additional side port can significantly reduce the amount of released air. Using the extra side port in combination with the carbon dioxide flushing technique reduces gas release further to small volumes. In a clinical setting this could be a promising approach to lower the risk of air embolism and stroke during thoracic endovascular aortic repair.
引用
- Rohlffs F, Tsilimparis N, Trepte C, et al. Air Embolism During TEVAR: An Additional Flush Port on the Delivery System Pusher Significantly Reduces the Amount of Air Released During Deployment of a Thoracic Stent-Graft in an Experimental Setting. J Endovasc Ther. 2018. PMID: 29865945. doi:10.1177/1526602818780951