文献速递主动脉疾病

超声多普勒在腹主动脉瘤腔内修复术后检测移植物血栓形成和内漏中的作用

The Role of Duplex Ultrasound in Detecting Graft Thrombosis and Endoleak after Endovascular Aortic Repair for Abdominal Aneurysm.
PubMed 文献 · Ann Vasc Surg · 2018-05-19

该回顾性研究纳入2个中心2000年1月1日至2016年12月31日因择期肾下腹主动脉瘤行腔内修复术的920例患者,评估双功能超声(DUS)检测内漏和移植物血栓形成的作用,以CT血管造影(CTA)为标准,DUS检测内漏的敏感度93.2%±5.8%、特异度98.8%±1.1%、准确率97.8%;DUS测量瘤囊最大横径较CTA低估2.5 mm(95%CI 2.25-2.75 mm)。

要点

  • DUS检测内漏以CTA为标准敏感度93.2%±5.8%、特异度98.8%±1.1%、准确率97.8%。
  • DUS与CTA测量瘤囊最大横径的Kappa一致性系数为0.91,DUS较CTA低估2.5 mm(95%CI 2.25-2.75 mm)。
  • DUS检出全部移植物血栓形成和大部分内漏,是可靠的检查方法。

核心结论

  • 与CTA相比,DUS是识别EVAR术后全部移植物血栓形成和大部分内漏的可靠检查。
  • DUS与CTA测量的瘤囊最大横径相关性良好,DUS较CTA轻度低估动脉瘤直径。

研究设计与样本

  • 回顾性收集2个血管外科中心2000年1月1日至2016年12月31日连续接受择期肾下腹主动脉瘤EVAR治疗的患者资料。
  • 共纳入920例患者(女性104例,11.3%),技术成功910例(98.9%)。
  • 每年进行CTA和DUS检查,两次检查间隔小于30天,配对比较瘤囊最大横径、内漏和移植物血栓形成的识别。

主要发现

  • 30天时死亡5例(0.5%),无移植物血栓形成,3例小的II型内漏未需任何附加手术。
  • 平均随访64.8个月(范围1-120.3个月),10年生存率和无并发症生存率分别为78.4%±1.9%和48.3%±1.7%。
  • 随访期间记录到额外76例内漏和8例移植物血栓形成。
  • 以CTA为标准,DUS检测内漏的敏感度93.2%±5.8%、特异度98.8%±1.1%、准确率97.8%,阳性预测值94.5%±5.2%,阴性预测值98.3%±1.4%。
  • 以真实临床诊断为标准,DUS检测内漏的敏感度89.4%±6.9%、特异度98.5%±1.3%、准确率96.9%,阳性预测值93.1%±5.8%,阴性预测值97.7%±1.6%。

关键数值与推荐

  • DUS与CTA测量瘤囊最大横径的Kappa一致性系数为0.91。
  • DUS较CTA低估动脉瘤直径,平均直径差2.5 mm(95%CI 2.25-2.75 mm)。
  • 统计学显著性水平为P值<0.05。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To assess the role of duplex ultrasound (DUS) in detecting endoleaks (ELs) and graft thrombosis (GT) in a cohort of patients submitted to endovascular aortic repair (EVAR) for elective infrarenal abdominal aortic aneurysm (AAA) in 2 centers. METHODS: Data, of all consecutive patients treated in 2 operative units of vascular surgery, from January 01, 2000 to December 31, 2016, were retrospectively collected and evaluated. Follow-up data were analyzed to evaluate survival and device-related complications, both at 30-day and in the midterm. The results of computed tomography angiography (CTA) and DUS which were performed yearly, with a time interval between the 2 examinations lower than 30 days, were paired in terms of maximum transverse diameters of the aneurysmal sac, identification of EL, and of GT. Sensibility (SE), specificity (SP), positive predictive value (PPV), negative predictive value (NPV), and accuracy (ACC) of the DUS were calculated with respect to data of each paired CTA and to data of the real clinical diagnosis of either EL or GT. A Student's t-test was performed to compare the aneurysmal sac diameters measured with DUS to those of CTA. "Kappa" coefficient of agreement was also calculated. A P value < 0.05 was the level reference for statistical significance. RESULTS: A total of 920 patients (104 female, 11.3%) underwent EVAR for elective infrarenal AAA. Technical success was achieved in 910 cases (98.9%). At 30 days, there were 5 deaths (0.5%), no GT, and 3 small type II ELs which did not require any adjunctive procedure. Mean follow-up was 64.8 months (range 1-120.3 months). Survival and freedom from complications were estimated to be 78.4% ± 1.9% and 48.3% ± 1.7%, respectively, at 10 years. Seventy-six more ELs and 8 GTs were recorded. Considering the CTA as the gold standard, SE, SP, and ACC of DUS in the detection of the EL were 93.2% ± 5.8%, 98.8% ± 1.1%, and 97.8%, respectively, with a PPV of 94.5% ± 5.2% and a NPV of 98.3% ± 1.4%. Considering, however, the real diagnosis of the EL, SE, SP, and ACC of DUS were slightly lower (89.4% ± 6.9%, 98.5% ± 1.3%, and 96.9%, respectively), as well as PPV and NPV (93.1% ± 5.8% and 97.7% ± 1.6%, respectively). K coefficient of agreement between the measurements of the sac maximum transverse diameter recorded at DUS and CTA was 0.91, with a little underestimation of the aneurysm diameter at DUS if compared with CTA (mean diameter difference of 2.5 mm, 95% confidence interval: 2.25-2.75 mm). CONCLUSIONS: DUS has proven to be a reliable examination in identifying all GT and most ELs after EVAR, compared with CTA. A good correlation was also observed between the measurements of the sac maximum transverse diameter recorded at DUS and CTA, with a little underestimation of the aneurysm diameter at DUS compared with CTA.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Mazzaccaro D, Farina A, Petsos K, et al. The Role of Duplex Ultrasound in Detecting Graft Thrombosis and Endoleak after Endovascular Aortic Repair for Abdominal Aneurysm. Ann Vasc Surg. 2018. PMID: 29787854. doi:10.1016/j.avsg.2018.03.040

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。