文献速递主动脉疾病

混合手术室内血管腔内动脉瘤修复术中职业辐射剂量的操作与步骤分析

Procedure and step-based analysis of the occupational radiation dose during endovascular aneurysm repair in the hybrid operating room.
PubMed 文献 · J Vasc Surg · 2018-01-01

本研究在混合手术室内对65例接受90枚支架移植物的血管腔内动脉瘤修复术患者测量支持人员的累积职业X线辐射剂量,并分析术中不同步骤的剂量分布。结果发现开窗/分支型血管腔内修复术的最大操作者有效剂量最高,为0.345 mSv,数字减影血管造影采集及目标血管再通等步骤是辐射剂量风险增加的预测因素。

要点

  • 开窗/分支型血管腔内修复术的最大操作者有效剂量中位数为0.345 mSv,高于其他术式。
  • 非复杂介入中最大操作者剂量占参考剂量百分比为4.4%,EVAR最低为3.5%。
  • 股动脉入路时最大操作者屏蔽可使相对职业辐射风险最小化。

核心结论

  • 在股动脉入路时对最大操作者进行屏蔽可使相对职业辐射风险最小化。
  • 数字减影血管造影图像采集、目标血管再通、肠系膜上动脉或腹腔干再通以及分支目标血管再通是职业辐射剂量风险增加的预测因素,原因包括患者辐射剂量增加和操作者屏蔽选项减少。

研究设计与样本

  • 本研究在混合手术室内对65例接受90枚支架移植物的血管腔内动脉瘤修复术患者进行术中辐射剂量测量,术式包括EVAR、分叉型EVAR、胸主动脉EVAR、髂动脉分支装置部署、主髂单支架桥和开窗/分支型EVAR(F/BrEVAR)。
  • X线成像使用Philips Allura FD20 Clarity系统,职业辐射剂量(估计有效剂量E,单位mSv)使用DoseAware Xtend系统个人剂量计测量。
  • 主要结局为每名工作人员的中位累积E(ESTAFF),按介入术式和EVAR/F/BrEVAR步骤或支架分层;次要结局为支持人员吸收E相对于C臂参考剂量计测量E的百分比(ESTAFF%)。
  • 所有结局以中位数和四分位间距表示,除非另有说明。

主要发现

  • 虚拟最大操作者(MO)的职业有效剂量(EMO)在下列术式中分别为:主髂单支架桥0.055(0.029-0.082)mSv(n=6),胸主动脉EVAR 0.084(0.054-0.141)mSv(n=14),分叉型EVAR 0.036(0.026-0.068)mSv(n=38),髂动脉分支装置部署0.054(0.035-0.126)mSv(n=8),F/BrEVAR 0.345(0.235-0.757)mSv(n=24)。
  • 每个肾目标血管的中位EMO为0.025(0.012-0.062)mSv,非肾内脏目标血管为0.146(0.07-0.315)mSv。
  • 在所有非复杂介入中,EMO%为4.4%(2.7%-7.3%),EVAR的中位率最低为3.5%(2.5%-5%)。
  • F/BrEVAR术式的EMO%中位率最高,为8.2%(5.0%-14.4%)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: This study measured the cumulative occupational X-ray radiation dose received by support staff during endovascular aortic procedures and during additional intraoperative steps in the hybrid operating room. METHODS: Radiation dose measurements were performed during interventions on 65 patients receiving 90 stent grafts during endovascular aneurysm repair (EVAR), bifurcated EVAR, thoracic EVAR, iliac branched device deployment, aortouni-iliac stenting, and fenestrated/branched EVAR (F/BrEVAR). X-ray imaging was acquired using the Philips Allura FD20 Clarity System (Philips Medical Systems, Best, The Netherlands). The occupational radiation dose (also referred to as the estimated effective dose, E, measured in millisieverts) was measured with the DoseAware Xtend system (Philips Medical Systems) personal dosimeters. E was reported per staff member (ESTAFF), where "staff" was a generic term for each staff member included in the study: the first operator (FO), the second operator (ESO), a virtual maximum operator (MO), and all additional supporting staff, including the sterile nurse, nonsterile nurse, anaesthesiologist, and radiation technician. The primary outcome was the median cumulative ESTAFF (or EFO, EMO, and so on), which was presented as the median cumulative dose per intervention and stratified for several within-interventional EVAR and F/BrEVAR steps or stents. The second outcome was the percentage of the absorbed E by a supporting staff member in relation to the E measured by the reference badge attached on the C-arm (ESTAFF% or EFO%, EMO%, and so on). All outcomes are presented as median with interquartile range, unless stated differently. RESULTS: The occupational effective dose in millisieverts of the MO (EMO) was 0.055 (0.029-0.082) for aortouni-iliac stenting (n = 6), 0.084 (0.054-0.141) during thoracic EVAR (n = 14), 0.036 (0.026-0.068) during bifurcated EVAR (n = 38), 0.054 (0.035-0.126) during iliac branched device deployment (n = 8), and 0.345 (0.235-0.757) during F/BrEVAR (n = 24). The median EMO in millisieverts was 0.025 (0.012-0.062) per renal target vessel (TV) and 0.146 (0.07-0.315) for a nonrenal visceral TV. During all noncomplex interventions, the EMO% was 4.4% (2.7%-7.3%), with the lowest median rate at 3.5% (2.5%-5%) for EVAR. The highest median rate EMO% was found for F/BrEVAR procedures: 8.2% (5.0%-14.4%). CONCLUSIONS: With maximum operator shielding during femoral access, relative occupational radiation risk can be minimized. However, digital subtraction angiography image acquisition, recanalization of TVs, recanalization of superior mesenteric artery or celiac artery, and recanalization of branched TVs are predictors for increased occupational radiation dose risks caused by increased radiation doses to the patient and reduced options for shielding of the operator.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. de Ruiter QMB, Jansen MM, Moll FL, et al. Procedure and step-based analysis of the occupational radiation dose during endovascular aneurysm repair in the hybrid operating room. J Vasc Surg. 2018. PMID: 29801556. doi:10.1016/j.jvs.2017.09.043

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。