文献速递主动脉疾病

非圆柱形瘤颈形态对血管内主动脉瘤修复术后结局的发生率及影响:Gore全球血管内主动脉治疗注册研究

The incidence and effect of noncylindrical neck morphology on outcomes after endovascular aortic aneurysm repair in the Global Registry for Endovascular Aortic Treatment.
PubMed 文献 · J Vasc Surg · 2018-05-24

基于Gore全球血管内主动脉治疗注册研究(GREAT)的回顾性分析,纳入2010年8月至2016年10月接受EXCLUDER内移植物治疗的患者3077例,评估非圆柱形瘤颈形态的发生率及其对再干预的影响。非圆柱形瘤颈在女性中更常见(17.4% vs 12.6%,P < .001)且与更严重的肾下角度相关(33.8° vs 28.4°,P < .001),需要更多主动脉延伸袖套(P = .004),但未增加再干预风险;女性性别和AAA直径与再干预需求增加相关。

要点

  • 非圆柱形瘤颈患者中女性比例更高(17.4% vs 12.6%,P < .001),肾下角度更大(33.8° vs 28.4°,P < .001)。
  • 非圆柱形瘤颈患者需要更多主动脉延伸袖套(P = .004),但瘤颈形态不是再干预的显著预测因素。
  • 女性性别和最大AAA直径是后续再干预(总体、器械相关、内漏特异)的显著风险因素。

核心结论

  • 非圆柱形瘤颈形态在女性中更常见,与主动脉延伸袖套使用增加相关,但未增加再干预风险。
  • 女性性别和AAA直径与再干预需求增加相关。

研究设计与样本

  • 回顾性分析Gore全球血管内主动脉治疗注册研究(GREAT)数据,纳入2010年8月至2016年10月接受EXCLUDER内移植物治疗且有近端主动脉锚定区直径测量的患者3077例。
  • 非圆柱形瘤颈定义为近端锚定区前15 mm内直径变化≥2 mm,提示锥形、圆锥形或沙漏形形态。
  • 采用Cox多因素回归分析任何再干预、器械相关再干预及内漏特异再干预。

主要发现

  • 3077例患者中,圆柱形瘤颈1765例,非圆柱形瘤颈1312例。
  • 非圆柱形瘤颈组女性比例显著高于圆柱形组(17.4% vs 12.6%,P < .001),肾下角度更严重(33.8° vs 28.4°,P < .001)。
  • 非圆柱形瘤颈组14.7%的患者和圆柱形组11.2%的患者在EXCLUDER说明书解剖纳入标准外植入(P = .004)。
  • 非圆柱形瘤颈患者需要更多主动脉延伸袖套(P = .004)。
  • 圆柱形组平均随访21.2 ± 17.5个月,非圆柱形组17.8 ± 15.8个月(P < .001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The Gore Global Registry for Endovascular Aortic Treatment (GREAT) was designed to evaluate real-world outcomes after treatment with Gore aortic endografts used in a real-world, global setting. We retrospectively analyzed the GREAT data to evaluate the incidence and effects of noncylindrical neck anatomy in patients undergoing endovascular aortic aneurysm repair. METHODS: The present analysis included patients with data in the GREAT who had been treated with the EXCLUDER endograft from August 2010 to October 2016. A noncylindrical neck was defined when the proximal aortic landing zone diameter had changed ≥2 mm over the first 15 mm of the proximal landing zone, indicating a tapered, conical, or hourglass morphology. Cox multivariate regression analyses were performed for any reintervention (including reinterventions on aortic branch vessels), device-related reinterventions, and reintervention specifically for endoleak. Independent binary (cylindrical vs noncylindrical necks) and continuous (percentage of neck diameter change) variables were assessed. The abdominal aortic aneurysm (AAA) diameter, proximal neck length, maximal infrarenal neck angle, gender, and use of aortic extender cuffs were also assessed. RESULTS: Of 3077 GREAT patients with available proximal aortic landing zone diameter measurements available, 1765 were found to have cylindrical necks and 1312 had noncylindrical necks. The noncylindrical neck cohort had a significantly greater proportion of women (17.4% vs 12.6%; P < .001) and more severe infrarenal angulation (33.8° vs 28.4°; P < .001). A total 14.7% of noncylindrical neck patients and 11.2% cylindrical neck patients underwent implantation outside of the EXCLUDER instructions for use regarding the anatomic inclusion criteria (P = .004). The procedural characteristics were similar between the two cohorts; however, noncylindrical neck patients required significantly more aortic extender cuffs (P = .004). The average follow-up was 21.2 ± 17.5 months and 17.8 ± 15.8 months for the cylindrical and noncylindrical cohorts, respectively (P < .001). The Cox multivariate regression models demonstrated female gender and maximum AAA diameter were significant risk factors for subsequent reintervention (overall, device-related, and endoleak-specific). Women were 2.2 times as likely to require device-related intervention during the follow-up period compared with men (P < .001). Neck shape morphology was not a significant predictor, except for device-related intervention, for which cylindrical necks (binary definition) resulted in a slightly elevated risk (1.5 times; P = .03). CONCLUSIONS: Noncylindrical neck morphology was more common in women and was associated with an increased use of aortic extender cuffs but did not increase the risk of intervention. Female gender and AAA diameter were associated with an increased need for reintervention.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Shutze W, Suominen V, Jordan W, et al. The incidence and effect of noncylindrical neck morphology on outcomes after endovascular aortic aneurysm repair in the Global Registry for Endovascular Aortic Treatment. J Vasc Surg. 2018. PMID: 29803683. doi:10.1016/j.jvs.2018.03.394

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。