无症状患者颈动脉支架置入术中栓塞滤器碎片负荷的定量分析与预测因素

Quantitative analysis and predictors of embolic filter debris load during carotid artery stenting in asymptomatic patients.
PubMed 文献 · J Vasc Surg · 2018-03-01

研究纳入2008至2016年278例无症状颈动脉狭窄>70%行颈动脉支架置入术(CAS)的患者,定量分析远端滤器捕获的栓塞碎片(EFD)负荷并识别预测因素。74%病例滤器存在碎片,平均负荷为滤器面积的10%;术后发生缺血性神经事件者平均EFD负荷显著高于无事件者(26.7%±19.0% vs 8.5%±13.5%,P<.001)。年龄>75岁、同侧缺血性脑病灶、低回声斑块和斑块长度>15 mm是EFD负荷>12.5%的独立预测因素。

要点

  • 无症状CAS患者中74%滤器可检测到栓塞碎片,平均负荷为滤器面积的10%。
  • 术后发生卒中或短暂性脑缺血发作者平均EFD负荷显著高于无事件者。
  • 年龄>75岁、同侧缺血性脑病灶、低回声斑块及斑块长度>15 mm为EFD负荷>12.5%的独立预测因素。

核心结论

  • 大多数接受CAS的无症状颈动脉狭窄患者保护滤器中可检测到栓塞碎片。
  • 年龄>75岁、同侧缺血性脑病灶、低回声斑块和斑块长度>15 mm应作为术中临床相关栓塞碎片负荷的独立预测因素加以考虑。

研究设计与样本

  • 前瞻性数据库纳入2008年至2016年所有无症状颈动脉狭窄>70%接受CAS的患者,共278例。
  • 所有患者均使用远端滤器保护装置,术后滤器经福尔马林固定并用立体显微镜及Image-Pro Plus软件进行形态计量分析。
  • 采用多元logistic回归评估解剖和临床变量与EFD负荷的关联。

主要发现

  • 278例中211例(76%)置入开环支架,67例(24%)置入闭环支架;总体技术成功率和临床成功率均为99%,无围手术期死亡。
  • 卒中发生率为1.8%( major n=1[0.4%];minor n=4[1.4%]),短暂性脑缺血发作发生率为5%(n=14)。
  • 滤器定量分析显示74%病例(n=207)存在EFD,平均EFD负荷为滤器面积的10%(中位数1;范围0-80)。
  • EFD负荷<10%者203例(73%),11%-20%者39例(14%),21%-30%者14例(5%),>31%者22例(8%)。
  • CAS后发生任何类型缺血性神经事件(卒中及短暂性脑缺血发作)的患者平均EFD负荷显著高于无事件病例(26.7%±19.0% vs 8.5%±13.5%;P<.001)。

关键数值与推荐

  • 观察性频率分布分析确定EFD负荷>12.5%为与临床相关围手术期缺血事件关联的最佳截断值(敏感性78%;特异性77%;曲线下面积0.81)。
  • 年龄>75岁(OR 2.56;P=.003)、同侧缺血性脑病灶(OR 2.09;P=.047)、术前双功超声低回声斑块(OR 6.05;P<.001)和斑块长度>15 mm(OR 1.79;P=.049)是EFD负荷>12.5%的独立预测因素。

临床怎么用

  • 作者认为年龄>75岁、同侧缺血性脑病灶、低回声斑块和斑块长度>15 mm应作为术中临床相关栓塞碎片负荷的独立预测因素加以考虑。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The objective of this study was to perform a quantitative analysis and to identify predictors of embolic filter debris (EFD) load during carotid artery stenting (CAS) in asymptomatic patients. METHODS: All patients with asymptomatic carotid stenosis >70% undergoing CAS between 2008 and 2016 were included in a prospective database. A distal filter protection device was used in all patients. At the end of the procedure, the filter was fixed in formalin and then analyzed with a stereomicroscope. Morphometric analysis was performed with Image-Pro Plus software (Media Cybernetics, Rockville, Md). The total area of the filter membrane and the area covered by particulate material were quantified. The quantity of membrane occupied by debris was expressed as percentage of covered surface area. Anatomic and clinical variables were evaluated for their association with EFD load using multiple logistic regression. RESULTS: Among the 278 patients undergoing CAS, an open-cell stent was implanted in 211 patients (76%); 67 patients (24%) received a closed-cell stent. Overall technical success and clinical success were both 99%; no perioperative death was reported. Stroke rate was 1.8% (major, n = 1 [0.4%]; minor, n = 4 [1.4%]); transient ischemic attacks occurred in 5% of cases (n = 14). The quantitative analysis of the filter revealed that EFD was present in 74% of cases (n = 207). The mean EFD load was 10% of the filter surface (median, 1; range, 0-80); it was <10% in 203 patients (73%), between 11% and 20% in 39 patients (14%), between 21% and 30% in 14 patients (5%), and >31% in 22 (8%). Patients with any type of ischemic neurologic event after CAS (stroke and transient ischemic attack) had a significantly higher mean EFD load compared with uneventful cases (26.7% ± 19.0% vs 8.5% ± 13.5%; P < .001). The observational frequency distribution analysis identified the presence of >12.5% EFD load as the optimal cutoff for the association with clinically relevant perioperative ischemic events (sensitivity, 78%; specificity, 77%; area under the curve, 0.81). The multivariate analysis demonstrated that age >75 years (odds ratio [OR], 2.56; P = .003), pre-existing ipsilateral ischemic cerebral lesions (OR, 2.09; P = .047), hypoechogenic plaque on the preoperative duplex ultrasound examination (OR, 6.05; P < .001), and plaque length >15 mm (OR, 1.79; P = .049) were independent predictors of EFD load >12.5%. CONCLUSIONS: The majority of asymptomatic carotid stenoses treated with CAS have detectable embolic debris in the protecting filter. Age >75 years, pre-existing ipsilateral cerebral ischemic lesions, hypoechogenic plaque, and plaque length >15 mm should be taken into consideration as independent predictors of clinically relevant embolic debris during the procedure.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Piazza M, Squizzato F, Chincarini C, et al. Quantitative analysis and predictors of embolic filter debris load during carotid artery stenting in asymptomatic patients. J Vasc Surg. 2018. PMID: 29503003. doi:10.1016/j.jvs.2017.09.055