文献速递主动脉疾病

应用心电图门控计算机断层扫描体外定量烟囱EVAR支架组合中内漏间隙形成与烟囱移植物受压

In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography.
PubMed 文献 · J Endovasc Ther · 2018-03-16

本研究在硅胶动脉瘤模型中置入三种烟囱EVAR支架组合(Endurant与Advanta V12、Viabahn、BeGraft),在生理流条件下应用心电图门控增强CT观察心动周期内几何变化。结果显示三种组合的内漏间隙体积和烟囱移植物几何形态在心动周期中均显著变化(p<0.001),Endurant-Advanta组合内漏间隙体积最大且变化最明显,Endurant-Viabahn组合内漏间隙较小且变异少但支架受压更明显。

要点

  • 三种烟囱EVAR组合的内漏间隙体积和移植物几何在心动周期中均显著变化(p<0.001)。
  • Endurant-Advanta组合内漏间隙体积最大(232.8 mm³)且收缩期-舒张期体积变化最大(20.7 mm³)。
  • Endurant-Viabahn组合内漏间隙较小、体积变异少,但支架受压更明显且不影响烟囱移植物表面积。

核心结论

  • 内漏间隙体积和支架受压是心动周期中动态重塑的现象。
  • 三种组合在心动周期中均观察到显著差异,自膨式(Endurant-Viabahn)烟囱EVAR组合内漏间隙较小且心动周期中内漏间隙体积变异较少,但支架受压更明显且不影响烟囱移植物表面积。

研究设计与样本

  • 体外研究:将三种烟囱EVAR支架组合(Endurant与Advanta V12、Endurant与Viabahn、Endurant与BeGraft)置入硅胶动脉瘤模型并施加生理流。
  • 采用心电图门控增强CT在心动周期10个时间点由2名观察者独立评估内漏间隙体积、移植物几何等指标。

主要发现

  • 三种组合的内漏间隙体积和烟囱移植物几何在心动周期中均显著变化(p<0.001)。
  • 所有组合均观察到内漏间隙和内漏。
  • Endurant-Advanta组合内漏间隙体积最大(232.8 mm³),收缩期与舒张期之间体积变化最大(20.7 mm³)。
  • 上述数值分别较Endurant-Viabahn组合高2.7倍和3.0倍,较Endurant-BeGraft组合高1.7倍和1.6倍。
  • Endurant-Viabahn组合D比值最高(右侧1.26-1.35;左侧1.33-1.48),Endurant-BeGraft组合最低(右侧1.11-1.17;左侧1.08-1.15)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To assess the dynamic behavior of chimney grafts during the cardiac cycle. METHODS: Three chimney endovascular aneurysm repair (EVAR) stent-graft configurations (Endurant and Advanta V12, Endurant and Viabahn, and Endurant and BeGraft) were placed in silicone aneurysm models and subjected to physiologic flow. Electrocardiography (ECG)-gated contrast-enhanced computed tomography was used to visualize geometric changes during the cardiac cycle. Endograft and chimney graft surface, gutter volume, chimney graft angulation over the center lumen line, and the D-ratio (the ratio between the lengths of the major and minor axes) were independently assessed by 2 observers at 10 time points in the cardiac cycle. RESULTS: Both gutter volumes and chimney graft geometry changed significantly during the cardiac cycle in all 3 configurations (p<0.001). Gutters and endoleaks were observed in all configurations. The largest gutter volume (232.8 mm3) and change in volume (20.7 mm3) between systole and diastole were observed in the Endurant-Advanta configuration. These values were 2.7- and 3.0-fold higher, respectively, compared to the Endurant-Viabahn configuration and 1.7- and 1.6-fold higher as observed in the Endurant-BeGraft configuration. The Endurant-Viabahn configuration had the highest D-ratio (right, 1.26-1.35; left, 1.33-1.48), while the Endurant-BeGraft configuration had the lowest (right, 1.11-1.17; left, 1.08-1.15). Assessment of the interobserver variability showed a high correlation (intraclass correlation >0.935) between measurements. CONCLUSION: Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant-Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Overeem SP, Donselaar EJ, Boersen JT, et al. In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography. J Endovasc Ther. 2018. PMID: 29544371. doi:10.1177/1526602818762399

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。