文献速递主动脉疾病

影像融合中的二维-三维配准方法在标准EVAR术中准确且有助于减少所用对比剂、辐射和手术时间

The 2D-3D Registration Method in Image Fusion Is Accurate and Helps to Reduce the Used Contrast Medium, Radiation, and Procedural Time in Standard EVAR Procedures.
PubMed 文献 · Ann Vasc Surg · 2018-03-06

该研究评估影像融合(IF)技术中二维-三维配准方法在血管腔内动脉瘤修复术(EVAR)中的准确性和有效性,回顾性分析2011至2015年本机构血管腔内主动脉数据库。结果显示配准误差中位数在右肾动脉起始处为1.8 mm、左肾动脉起始处为1.0 mm,42%的病例达到完全视觉准确。使用IF的105例患者相比47例对照,对比剂用量显著减少(中位数58 mL,P < 0.0001),但在总体辐射暴露和手术时间上无显著差异。

要点

  • 二维-三维配准IF技术可准确勾画目标血管,中位配准误差为右肾动脉1.8 mm、左肾动脉1.0 mm。
  • IF组对比剂用量中位数58 mL,显著少于对照组(P < 0.0001)。
  • 在最有经验的两位术者亚组中,IF组辐射剂量更低且手术时间缩短。

核心结论

  • 二维-三维配准的IF引导能准确勾画目标血管,并可能有助于执行EVAR手术,显著减少对比剂用量,且在术者对EVAR和IF更熟悉时还能减少辐射并缩短手术时间。

研究设计与样本

  • 回顾性审查本机构2011至2015年间接受EVAR的血管腔内主动脉数据库,比较安装三维IF计算机断层扫描系统前后。
  • 评估准确性的14例血管腔内手术中,右肾动脉起始处中位配准误差为1.8 mm,左肾动脉起始处为1.0 mm,42%的病例达到完全视觉准确。
  • 105例患者术中采用二维-三维配准的IF技术(IF组),47例未使用该技术作为对照。

主要发现

  • IF组对比剂用量中位数为58 mL,显著少于对照组(P < 0.0001)。
  • IF组与对照组术中辐射暴露相似,剂量面积乘积中位数分别为2,343.7 cGy cm2与3,219 cGy cm2(P = 0.457)。
  • 由最有经验的两位术者操作的IF亚组辐射剂量低于对照亚组(中位数1,087 cG cm2 vs. 2,705.3 cG cm2,P = 0.012)。
  • 两组手术时间和术中辐射时间无显著差异(P = 0.117和0.106),IF亚组透视时间亦无显著差异(中位数6.3 min vs. 9.5 min,P = 0.067),但两位最有经验的术者使用IF时手术时间缩短(P = 0.002)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: This study aimed to evaluate the accuracy and the effectiveness of 2D-3D registration method of image fusion (IF) technology in endovascular aneurysm repair (EVAR). METHODS: We performed a review of our institutional endovascular aortic database of patients who had undergone EVAR between 2011 and 2015 before and after the installation of a 3D IF computed tomography system in our hybrid operating room. RESULTS: The accuracy was assessed in 14 endovascular procedures and showed a median registration error of 1.8 mm at the origin of the right renal artery and 1.0 mm at the origin of the left renal artery and a complete visual accuracy in 42% of the cases. EVAR was performed using the intraoperative IF technique with a 2D-3D registration method in 105 patients (group IF), whereas 47 patients done without served as controls. The IF group had a significantly reduced amount of used contrast compared with controls with a median of 58 mL and P < 0.0001. The intraoperative exposition to radiation was similar between the 2 groups with a median dose area product of 2,343.7 cGy cm2 in the IF group and 3,219 cGy cm2 among the controls (P = 0.457). The radiation dose in the sub group IF (including patients operated by the 2 most experienced surgeons) was lower than that in sub controls (median, 1,087 cG cm2 vs. 2,705.3 cG cm2, P = 0.012). The procedure time and the time of intraoperative radiation did not differ between the study groups (P = 0.117 and 0.106, respectively), as did not fluoroscopy time in the sub group IF (median, 6.3 min, vs. 9.5 min, P = 0.067), but for the 2 most experienced surgeons, the procedural time was shortened when using IF (P = 0.002). CONCLUSIONS: The 2D-3D registration method of IF guidance is accurate to delineate the vessels of interest and could help the execution of the EVAR procedures with a significantly reduced amount of contrast medium and also with reduced radiation and shorter procedural duration when surgeons are more familiar with EVAR and IF.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Ahmad W, Obeidi Y, Majd P, et al. The 2D-3D Registration Method in Image Fusion Is Accurate and Helps to Reduce the Used Contrast Medium, Radiation, and Procedural Time in Standard EVAR Procedures. Ann Vasc Surg. 2018. PMID: 29522869. doi:10.1016/j.avsg.2018.01.098