不同下腔静脉滤器模型间径向扩张力的差异与已报道的穿孔率相符
本研究用三重复测量Greenfield、Cook Celect和Cook Platinum三种下腔静脉滤器的径向扩张力,发现各模型径向扩张力显著不同(P < .001),扩张过程中力的分布速率也显著不同(P < .001)。结论认为滤器模型间径向扩张力差异与已报道的穿孔率一致,Cook Celect滤器留置>90天时穿孔发生率高于Greenfield滤器。
要点
- 三种下腔静脉滤器模型的径向扩张力显著不同(P < .001)。
- 最小腔静脉直径下径向扩张力最大为Cook Platinum,其次为Cook Celect和Greenfield。
- Cook Celect滤器留置>90天时穿孔发生率高于Greenfield滤器。
核心结论
- 下腔静脉滤器模型间径向扩张力的差异与已报道的穿孔率一致。
- Cook Celect下腔静脉滤器留置>90天时穿孔发生率高于Greenfield滤器。
- Cook Celect滤器在最小腔静脉直径下径向扩张力显著更高,扩张过程中力分散更大。
研究设计与样本
- 体外研究,每组三个重复,分析Greenfield(Boston Scientific,Marlborough,Mass)、Cook Celect(Cook Medical,Bloomington,Ind)和Cook Platinum滤器的径向扩张力。
- 在受控扩张过程中实时采集滤器对测量装置施加的内在力。
- 重复样本取平均值,使用SAS软件(SAS Institute,Cary,NC)计算协方差分析确定显著性。
主要发现
- 每个滤器模型产生的径向扩张力显著不同(P < .001),扩张过程中力的分布速率也显著不同(P < .001)。
- 最小腔静脉直径下径向扩张力最大为Cook Platinum滤器,其次为Cook Celect和Greenfield滤器。
- 扩张过程中径向力分散最大为Cook Celect,其次为Cook Platinum和Greenfield滤器。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Inferior vena cava (IVC) filters are used in patients at risk for pulmonary embolism who cannot be anticoagulated. Unfortunately, these filters are not without risk, and complications include perforation, migration, and filter fracture. The most prevalent complication is filter perforation of the IVC, with incidence varying among filter models. To our knowledge, the mechanical properties of IVC filters have not been evaluated and are not readily available through the manufacturer. This study sought to determine whether differences in mechanical properties are similar to differences in documented perforation rates. METHODS: The radial expansion forces of Greenfield (Boston Scientific, Marlborough, Mass), Cook Celect (Cook Medical, Bloomington, Ind), and Cook Platinum filters were analyzed with three replicates per group. The intrinsic force exerted by the filter on the measuring device was collected in real time during controlled expansion. Replicates were averaged and significance was determined by calculating analysis of covariance using SAS software (SAS Institute, Cary, NC). RESULTS: Each filter model generated a significantly different radial expansion force (P < .001), and force was distributed at significantly different rates (P < .001) during expansion. The largest radial expansion force at minimal caval diameter was seen in the Cook Platinum filter, followed by the Cook Celect and Greenfield filters. Radial force dispersion during expansion was greatest in the Cook Celect, followed by the Cook Platinum and Greenfield filters. CONCLUSIONS: Differences in radial expansion forces among IVC filter models are consistent with documented perforation rates. Cook Celect IVC filters have a higher incidence of perforation compared with Greenfield filters when they are left in place for >90 days. Evaluation of Cook Celect filters yielded a significantly higher radial expansion force at minimum caval diameter, with greater force dispersion during expansion.