本文报告一例75岁男性Ovation支架置入后发生II型内漏,经 translumbar 囊穿刺注射弹簧圈和组织胶栓塞后,出现支架无支撑部分因胶水外压而严重管腔狭窄,经皮跨动脉对吻球囊扩张成功挽救。提示该并发症罕见但需警惕。

要点

  • Ovation支架无支撑部分在胶水栓塞后可能发生外压性狭窄。
  • 经皮跨动脉对吻球囊扩张可成功挽救该并发症。
  • II型内漏胶水栓塞后需警惕支架受压导致管腔狭窄。

核心结论

  • 胶水栓塞治疗II型内漏后Ovation支架外压性狭窄是罕见但可能带来问题的并发症。

研究设计与样本

  • 病例报告:1例75岁男性,既往缺血性心脏病史,Ovation支架行EVAR后出现II型内漏。

主要发现

  • 经 translumbar 囊穿刺联合弹簧圈和组织胶成功栓塞II型内漏。
  • 最终造影见Ovation支架无支撑部分因胶水外压而严重管腔狭窄。
  • 经皮跨动脉对吻球囊扩张成功挽救狭窄。

临床怎么用

  • 对Ovation支架行II型内漏胶水栓塞后,应通过造影警惕支架外压性狭窄。
  • 发现支架外压性狭窄时,可考虑经皮跨动脉对吻球囊扩张进行挽救。

局限性

  • 单例病例报告,结论的外推性有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To describe a case of extrinsic compression of the Ovation stent-graft following glue embolization for type II endoleak. CASE REPORT: A 75-year-old man with a past history of ischemic heart disease and endovascular aneurysm repair with an Ovation stent-graft was admitted for treatment of type II endoleaks from the right L2 and left L4 lumbar arteries with egress via the inferior mesenteric and right L4 lumbar arteries, respectively. Successful embolization was performed via a translumbar sac puncture with a combination of coils and histoacryl glue. On final angiography severe lumen narrowing of the unsupported portion of the Ovation stent-graft was seen owing to extrinsic compression by the glue. This was successfully salvaged with percutaneous transarterial kissing balloon angioplasty. CONCLUSION: Aortic lumen narrowing caused by extrinsic compression of an Ovation stent-graft following glue embolization of type II endoleak is an unusual and potentially problematic complication.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Ni W, Leong S, Chng JK, et al. Extrinsic Compression of the Ovation Stent-Graft Following Glue Embolization for Type II Endoleak: An Unusual Complication. J Endovasc Ther. 2018. PMID: 29409379. doi:10.1177/1526602818757013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。