肿胀麻醉与体位对腔内激光消融激光尖端连接距离的影响
该回顾性研究纳入2008至2014年单术者收治的症状性静脉曲张行腔内激光治疗的患者,术中分别在肿胀麻醉前和Trendelenburg体位激光发射前测量激光尖端至隐股或隐腘连接处的距离。结果显示62%患者连接距离增加,23%减少,仅23%无变化,部分需调整激光尖端位置以维持0.75至2 cm的目标距离,建议激光激活前最终测量连接距离。
要点
- 肿胀麻醉和Trendelenburg体位可改变激光尖端与连接处的距离。
- 62%患者连接距离增加,其中17%需推进激光尖端。
- 23%患者连接距离减少,其中58%需回撤激光尖端。
核心结论
- 肿胀麻醉浸润和Trendelenburg体位会影响激光尖端放置,推荐在激活激光前进行最终连接距离测量以维持安全且最佳的距离。
研究设计与样本
- 回顾性分析前瞻性维护数据库,纳入2008年2月至2014年2月单术者在教学医院血管单元收治的行腔内激光治疗的症状性静脉曲张患者。
- 术中测量两次激光尖端至隐股或隐腘连接处的距离:肿胀麻醉前和Trendelenburg体位激光发射前。
主要发现
- 62%病例(490例患者;大隐静脉348例,小隐静脉142例)连接距离增加,其中17%(84例;大隐静脉56例,小隐静脉28例)需推进激光尖端以维持0.75至2 cm的目标连接距离。
- 23%患者(185例;大隐静脉155例,小隐静脉30例)连接距离减少,其中58%(大隐静脉79例,小隐静脉28例)需回撤激光尖端至目标连接距离。
- 23%患者(185例)连接距离无变化。
关键数值与推荐
- 目标连接距离为0.75至2 cm。
临床怎么用
- 在腔内激光消融中,肿胀麻醉和Trendelenburg体位后应重新测量激光尖端至连接处的距离,必要时调整尖端位置以确保安全有效的治疗。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The correct positioning of the laser tip at the saphenofemoral or saphenopopliteal junction during endovenous laser therapy is paramount to ensure a safe and effective procedure. The aim of this study was to demonstrate how patient positioning and tumescence infiltration can affect this safe junctional distance. METHODS: A retrospective review of a prospectively maintained database was carried out for all patients who received endovenous laser treatment for symptomatic varicose veins between February 2008 and February 2014 in one surgeon's practice in a teaching hospital vascular unit. The junctional distance of the laser tip from the saphenofemoral or saphenopopliteal junction was measured two times during the procedure: before tumescence and before laser deployment with the patient in a Trendelenburg position. RESULTS: Junctional distance was found to have increased in 62% cases (490 patients; great saphenous vein [GSV], 348; small saphenous vein [SSV], 142). Of these, 17% (84) required the laser tip to be advanced (GSV, 56; SSV, 28) to maintain a desired junctional distance of 0.75 to 2 cm. In 185 patients (23%), the junctional distance was noted to have been reduced (GSV, 155; SSV, 30), with 58% (GSV, 79; SSV, 28) requiring the laser tip to be withdrawn to the desired junctional distance; 23% of patients (185) had no change in the junctional distance. CONCLUSIONS: This study has demonstrated the effect of tumescence infiltration and Trendelenburg positioning on laser tip placement, and thus a final junctional measurement before activation of the laser is recommended to maintain a safe and optimal junctional distance.
引用
- Al Shakarchi J, Kay M, Kuyumdzhiev S, et al. Effect of tumescent anesthesia and patient positioning on laser tip junctional distance during endovenous laser ablation. J Vasc Surg Venous Lymphat Disord. 2017. PMID: 29097173. doi:10.1016/j.jvsv.2017.08.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。