文献速递主动脉疾病

Nellix装置腔内动脉瘤封闭术后第一年CT影像演变

Evolution of Computed Tomography Imaging the First Year after Endovascular Sealing of Infrarenal Aortic Aneurysms Using the Nellix Device.
PubMed 文献 · Ann Vasc Surg · 2017-09-12

纳入3家医院50例肾下腹主动脉瘤患者,分析150次CTA,评估Nellix EVAS术后30天与1年的正常影像演变。术后瘤径先轻度增加后回到术前值,主动脉容积分布发生改变:聚合物容积轻度但显著增加,附壁血栓容积呈下降趋势;髂动脉成角先减小后恢复,瘤颈变直持续存在,未见内漏,装置位置稳定。

要点

  • EVAS术后CTA表现独特,聚合物袋容积与放射性密度变化不同于经典EVAR,判读需经验。
  • 术后1年内主动脉成角、容积与放射性密度存在可预期的正常变化,瘤颈与髂动脉直径不变。
  • 随访中未见内漏且装置位置稳定,影像变化本身不应误判为并发症。

核心结论

  • EVAS术后CT表现具有独特性,其正常演变涉及主髂成角、容积与放射性密度变化,判读需结合经验。
  • 术后1年内未见内漏,装置位置稳定,瘤颈与髂动脉直径保持不变。

研究设计与样本

  • 多中心回顾性影像分析,纳入3家医院50例接受Nellix EVAS治疗的肾下腹主动脉瘤患者。
  • 使用专用软件按预设解剖节段变量分析共150次CTA扫描,评估术后30天与1年的影像表现。

主要发现

  • 术后30天AAA直径轻度增加但无统计学意义,1年后回到术前值。
  • 血栓容积30天为85.6 ± 49.1 mL、1年为78.8 ± 35.5 mL(P < 0.242);聚合物容积分别为68.2 ± 34.1 mL与71.9 ± 35.2 mL(P < 0.001)。
  • β角术后减小(P = 0.06),髂动脉成角术后减小(P < 0.001)且1年后恢复,而瘤颈变直持续存在。
  • 聚合物填充袋中部放射性密度随时间显著下降、边缘升高(P < 0.05);血栓放射性密度第一年显著升高(P < 0.05)。

临床怎么用

  • EVAS术后随访判读CTA时,应认识聚合物袋容积与放射性密度变化及主髂成角演变属正常表现,避免误判为内漏或装置失败。
  • 关注瘤颈变直持续存在与髂动脉成角恢复等特征,结合装置位置稳定与无内漏综合评估。

局限性

  • 样本量50例、随访1年,为影像演变观察,未报告临床结局与长期随访数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The Nellix endovascular aneurysm sealing (EVAS) system is an alternative endovascular treatment option for infrarenal aortic aneurysms (AAAs), with a unique appearance on computed tomography angiography (CTA). Normal evolution of post-EVAS CTA appearance follow-up is still largely unknown and important to timely detect eventual complications. The objective is to assess the normal appearance of CTA images 30 days and 1 year after EVAS in 50 consecutive patients. METHODS: Fifty patients treated with Nellix EVAS for an infrarenal AAA were included from 3 hospitals. Using dedicated software, a total of 150 CTA scans were analyzed by predetermined variables per anatomical segment. RESULTS: Thirty days post-EVAS, there was a slight, but not statistically significant, increase in AAA diameter that returned to the preoperative value after 1 year. A shift in total aortic volume distribution was observed without changing aortic diameter, including a trend toward a decreased thrombus volume (85.6 ± 49.1 mL and 78.8 ± 35.5 mL at 30 days and 1 year, respectively, P < 0.242) and a slight, but statistically significant, increase in polymer volume (68.2 ± 34.1 mL and 71.9 ± 35.2 mL at 30 days and 1 year, respectively, P < 0.001). The β-angle (P = 0.06) and iliac artery angulation (P < 0.001) decreased after implant. The latter returned to its original state after 1 year, whereas the neck straightening remained. Over time, there was a significant decrease in radiodensity in the middle of the polymer-filled endobags with an increase at its edges (P < 0.05). Thrombus radiodensity significantly increased over the first year (P < 0.05). Diameters of the infrarenal neck and common iliac arteries remained unchanged, no endoleaks were observed, and the position of the device was stable. CONCLUSIONS: Change of CT appearance after EVAS is unique, and as such, the judgment of these images requires experience. The appearance of the endobags in respect to volume and radiodensity differ from classic EVAR. Normal changes over time are observed in aortoiliac angulation, volumes, and radiodensities.