文献速递主动脉疾病内脏/肾动脉与血管畸形

同步辐射精准评估肾上支架丝所致主动脉分支开口截面积减少

Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires.
PubMed 文献 · J Endovasc Ther · 2017-09-19

本研究在人体主动脉体模中置入Zenith支架,比较同步辐射与CTA测量肾上支架丝所致主动脉分支开口截面积减少的能力。同步辐射显示单丝占开口面积<10%、双丝<20%,而CTA测量分别为24%-25%和40%-48%,且CTA显著高估支架丝直径。结论认为CTA空间分辨率不足,同步辐射评估更准确。

要点

  • 同步辐射测量单丝和双丝所致分支开口截面积减少分别为<10%和<20%。
  • CTA测量单丝和双丝所致截面积减少分别为24%-25%和40%-48%。
  • CTA显著高估支架丝直径(1.15±0.01至1.57±0.02 mm vs 实际0.47±0.01 mm)。

核心结论

  • 同步辐射在评估主动脉支架丝及分支开口截面积减少方面优于CTA。

研究设计与样本

  • 在人体主动脉体模中置入Zenith支架,模拟腹主动脉瘤治疗,比较同步辐射与64排CTA的测量结果。

主要发现

  • 同步辐射显示单丝占肾动脉开口截面积<10%,双丝占肾内脏分支开口<20%。
  • CTA测量单丝和双丝所致截面积减少分别为24%-25%和40%-48%。
  • 同步辐射测得支架丝直径为0.38±0.01至0.53±0.03 mm,接近实际值0.47±0.01 mm。
  • CTA测得支架丝直径为1.15±0.01至1.57±0.02 mm,明显高估。

临床怎么用

  • 提示CTA在评估支架丝所致分支开口狭窄时可能高估狭窄程度,临床解读需谨慎。

局限性

  • 研究为体模实验,非人体在体研究,结果外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To compare in vivo the use of synchrotron radiation to computed tomography angiography (CTA) for the measurement of cross-sectional area (CSA) reduction of the aortic branch ostia caused by suprarenal stent-graft wires. METHODS: This study was performed with a Zenith stent-graft placed in a phantom of the human aorta to simulate treatment of abdominal aortic aneurysm. Synchrotron radiation scans were performed using beam energies between 40 and 100 keV and spatial resolution of 19.88 μm per pixel. CSA reduction of the aortic branch ostia by suprarenal stent wires was calculated based on these exposure factors and compared with measurements from CTA images acquired on a 64-row scanner with slice thicknesses of 1.0, 1.5, and 2.0 mm. RESULTS: Images acquired with synchrotron radiation showed <10% of the CSA occupied by stent wires when a single wire crossed a renal artery ostium and <20% for 2 wires crossing a renovisceral branch ostium. The corresponding areas ranged from 24% to 25% for a single wire and from 40% to 48% for double wires crossing the branch ostia when measured on CT images. The stent wire was accurately assessed on synchrotron radiation with a diameter between 0.38±0.01 and 0.53±0.03 mm, which is close to the actual size of 0.47±0.01 mm. The wire diameter measured on CT images was greatly overestimated (1.15±0.01 to 1.57±0.02 mm). CONCLUSION: CTA has inferior spatial resolution that hinders accurate assessment of CSA reduction. This experiment demonstrated the superiority of synchrotron radiation over CTA for more accurate assessment of aortic stent wires and CSA reduction of the aortic branch ostia.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Sun Z, Ng CKC. Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires. J Endovasc Ther. 2017. PMID: 28922970. doi:10.1177/1526602817732315