腹主动脉瘤腔内修复中部分肾动脉覆盖导致不良肾血流模式

Partial renal coverage in endovascular aneurysm repair causes unfavorable renal flow patterns in an infrarenal aneurysm model.
PubMed 文献 · J Vasc Surg · 2017-10-12

本研究利用体外腹主动脉瘤模型,通过二维粒子图像测速技术观察生理静息条件下脉冲流,比较无支架、支架无肾覆盖及支架部分肾覆盖(1.3 mm织物超出标记,16%面积覆盖)对肾动脉血流模式的影响。结果显示部分覆盖导致低且振荡的壁面剪切应力(WSS)区域增大,肾动脉平均回流面积百分比从对照的0.9%增至无覆盖的6.4%和部分覆盖的18.8%。结论提示部分肾覆盖可能促进动脉粥样硬化及肾动脉狭窄。

要点

  • 部分肾动脉覆盖导致肾动脉壁低且振荡的WSS,可能促进动脉粥样硬化。
  • 部分覆盖模型肾动脉回流面积百分比达18.8%,高于无覆盖的6.4%和对照的0.9%。
  • 临床需注意不同装置织物边缘与不透射线标记之间的偏移,避免意外部分覆盖。

核心结论

  • 在体外血流模型中,腹主动脉瘤腔内修复后部分肾动脉覆盖引起低且明显振荡的WSS,可能促进动脉粥样硬化及后续肾动脉狭窄。

研究设计与样本

  • 采用体外腹主动脉瘤模型,在生理静息条件下使用二维粒子图像测速技术可视化脉冲流。
  • 共三个模型:一个对照,两个植入Endurant支架(Medtronic Inc),其中一个无肾动脉覆盖,另一个部分覆盖(1.3 mm支架织物超出标记,16%面积覆盖)。
  • 分析肾动脉WSS的幅度与振荡、相对停留时间及回流。

主要发现

  • 两个支架模型沿肾动脉下壁均出现低且振荡的WSS区域,对照模型无此现象。
  • 部分覆盖模型存在长度7 mm的极低WSS(<0.1 Pa)区域,而无覆盖模型该区域长度为2 mm。
  • 肾动脉平均回流面积百分比从对照的0.9%增至无覆盖的6.4%和部分覆盖的18.8%。

临床怎么用

  • 临床行EVAR时应了解所用装置织物边缘与不透射线标记间的偏移,避免意外部分覆盖肾动脉。

局限性

  • 研究为体外模型,结果不能直接等同于体内情况。
  • 仅测试了一种支架装置(Endurant)及特定覆盖程度(1.3 mm,16%面积),其他装置或覆盖程度未评估。
  • 血流条件为生理静息状态,未模拟运动或其他生理变化。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To achieve an optimal sealing zone during endovascular aneurysm repair, the intended positioning of the proximal end of the endograft fabric should be as close as possible to the most caudal edge of the renal arteries. Some endografts exhibit a small offset between the radiopaque markers and the proximal fabric edge. Unintended partial renal artery coverage may thus occur. This study investigated the consequences of partial coverage on renal flow patterns and wall shear stress (WSS). METHODS: In vitro models of an abdominal aortic aneurysm were used to visualize pulsatile flow using two-dimensional particle image velocimetry under physiologic resting conditions. One model served as control and two models were stented with an Endurant endograft (Medtronic Inc, Minneapolis, Minn), one without and one with partial renal artery coverage with 1.3 mm of stent fabric extending beyond the marker (16% area coverage). The magnitude and oscillation of WSS, relative residence time, and backflow in the renal artery were analyzed. RESULTS: In both stented models, a region along the caudal renal artery wall presented with low and oscillating WSS, not present in the control model. A region with very low WSS (<0.1 Pa) was present in the model with partial coverage over a length of 7 mm compared with a length of 2 mm in the model without renal coverage. Average renal backflow area percentage in the renal artery incrementally increased from control (0.9%) to the stented model without (6.4%) and with renal coverage (18.8%). CONCLUSIONS: In this flow model, partial renal coverage after endovascular aneurysm repair causes low and marked oscillations in WSS, potentially promoting atherosclerosis and subsequent renal artery stenosis. Awareness of the device-dependent offset between the fabric edge and the radiopaque markers is therefore important in endovascular practice.