这项多中心回顾性观察研究纳入461例择期EVAR患者,43.8%存在至少一项装置特异性IFU违反。Kaplan-Meier和Cox模型显示,任何IFU违反与移植物相关不良事件显著相关(HR 1.8,95%CI 1.05-3.1),颈部特异性违反关联更强(HR 2.2,95%CI 1.2-4.0)。

要点

  • 43.8%的择期EVAR患者存在至少一项装置特异性IFU违反。
  • 任何IFU违反与GRAE风险独立相关(HR 1.8,95%CI 1.05-3.1)。
  • 颈部特异性IFU违反与GRAE关联更强(HR 2.2,95%CI 1.2-4.0)。

核心结论

  • EVAR患者中任何装置特异性IFU违反与随访期间移植物相关不良事件显著相关。
  • 颈部特异性IFU违反与GRAE的关联较任何IFU违反更强。

研究设计与样本

  • 多中心回顾性观察研究,纳入2005至2014年接受择期EVAR的461例腹主动脉瘤患者,使用三种不同装置。
  • 收集人口学、解剖及随访数据,分析装置特异性IFU违反与GRAE的关系。

主要发现

  • 43.8%的患者存在至少一项IFU违反。
  • IFU违反者外周血管病比例更高(12.4% vs 7.3%),男性比例更低(78.7% vs 90.3%)。
  • 最常见IFU违反为颈部直径偏差(15.2%)和髂动脉直径偏差(21.4%)。
  • 总体GRAE发生率为12.8%。
  • Kaplan-Meier分析显示IFU违反与GRAE显著相关(log-rank P = .031),Cox模型调整后HR 1.8(95%CI 1.05-3.1)。

临床怎么用

  • 对存在IFU偏离的EVAR候选患者应谨慎评估。
  • 应重视颈部解剖条件对EVAR术后GRAE风险的影响。

局限性

  • 本研究为回顾性观察性研究。
  • 随访时间中位数较短(IFU违反者1.9年,无违反者2.1年)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: A variety of devices exist for endovascular aneurysm repair (EVAR). Device-specific instructions for use (IFU) detail anatomic constraints to application and deployment of devices and are developed from rigorous bench testing. Nonadherence to IFU occurs frequently to avoid open surgery. The purpose of this study was to determine if IFU violations are associated with increased risk of graft-related adverse events (GRAEs) during follow-up. METHODS: This multicenter retrospective observational study included patients undergoing elective endovascular repair for abdominal aneurysmal disease with three different devices. Demographics, anatomic data, and follow-up data were collected on patients from 2005 to 2014. IFU violations were device specific and included neck diameter, length, and angulation and iliac artery diameter and length. GRAEs included a composite outcome of reintervention, migration, endoleak (type II excluded), rupture, limb occlusion, sac growth, and aneurysm-related mortality during the follow-up period. Kaplan-Meier survival and Cox proportional hazards modeling were performed. Any IFU violations as well as neck-specific IFU violations were analyzed. RESULTS: In 461 patients undergoing EVAR, 43.8% had at least one IFU violation. Patients with IFU violations were more likely to have peripheral vascular disease (12.4% vs 7.3%) and were less likely to be male (78.7% vs 90.3%). The most frequent IFU violations included diameter deviations of the neck (15.2%) and of the iliac artery (21.4%). Overall, the GRAE rate was 12.8%. Median follow-up time was 1.9 and 2.1 years for patients with and without an IFU violation, respectively. Kaplan-Meier survival revealed a significant association between the presence of an IFU violation and GRAEs (log-rank, P = .031). When adjusted for clinical variable through Cox hazard modeling, the association remained significant (hazard ratio 1.8; 95% confidence interval, 1.05-3.1). When neck-specific violations were considered independently, Kaplan-Meier survival (log-rank, P = .003) and Cox modeling (hazard ratio, 2.2; 95% confidence interval, 1.2-4.0) revealed a significant association between neck-specific IFU violation and GRAEs. CONCLUSIONS: A total of 43.8% of patients undergoing EVAR had a device-specific IFU violation, indicating that implanters are pushing the boundaries of device capabilities. Our study identified that any IFU violation was significantly associated with GRAEs over time. Caution should be applied to patients being considered for EVAR when IFU deviations exist.