文献速递主动脉疾病

现成开窗支架治疗急性B型主动脉夹层的可行性研究

A Feasibility Study of Off-the-Shelf Scalloped Stent-Grafts in Acute Type B Aortic Dissection.
PubMed 文献 · J Endovasc Ther · 2017-08-17

本研究基于70例急性Stanford B型主动脉夹层患者的CTA回顾性分析,评估现成开窗支架在TEVAR中保留左锁骨下动脉血流的适用性。结果显示80%患者需覆盖LSA,其中41%适用开窗支架,三种尺寸的现成支架可治疗36%需覆盖LSA的患者。

要点

  • 急性B型夹层TEVAR中80%需覆盖左锁骨下动脉。
  • 需覆盖LSA者中41%适用现成开窗支架。
  • 三种尺寸的现成支架可治疗36%需覆盖LSA的患者。

核心结论

  • 急性B型主动脉夹层中左锁骨下动脉受累常见,现成开窗支架可能为保留LSA血流提供简单方案。

研究设计与样本

  • 单中心回顾性影像学研究,分析70例 consecutive 急性Stanford B型主动脉夹层患者的CTA。

主要发现

  • 56例(80%)患者需覆盖LSA。
  • 需覆盖LSA的患者中,23例(41%)适用现成开窗支架。
  • 适用组中位主动脉直径31 mm(范围26-37),LSA开口中位长度13 mm(范围10-20),中位宽度15 mm(范围11-24)。
  • 三种不同尺寸的现成支架(最大开窗)可充分治疗56例中的20例(36%)。

关键数值与推荐

  • 近端锚定区<15 mm(距逆向剥离壁层)需覆盖LSA。
  • 开窗支架适用条件:内膜破口距LSA开口至少20 mm。

临床怎么用

  • 在急诊/紧急情况下,现成开窗支架可能简化TEVAR中LSA血流的保留,降低大脑后部、脊髓和左臂灌注不良风险。

局限性

  • 单中心、回顾性、基于影像的研究,样本量有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate the applicability of an off-the-shelf scalloped stent-graft to preserve left subclavian artery (LSA) flow in thoracic endovascular aortic repair (TEVAR) for acute type B aortic dissection. METHODS: The computed tomography angiograms (CTA) of 70 consecutive patients (median age 64 years; 44 men) with acute Stanford type B aortic dissection were retrospectively analyzed to identify patients in whom a short proximal landing zone (<15 mm from the retrogradely dissected wall layers) would require LSA overstenting during TEVAR. A scalloped stent-graft was deemed possible in those patients with the intimal entry tear located at least 20 mm distant from the LSA ostium. RESULTS: The LSA needed to be covered in 56 (80%) patients. Of these, an off-the-shelf scalloped stent-graft would have been applicable in 23 (41%) patients. In the latter group, the median aortic diameter was 31 mm (range 26-37), the median length of the LSA ostium was 13 mm (range 10-20), and the median width of the LSA ostium was 15 mm (range 11-24). Three differently sized off-the-shelf stent-grafts with the largest scallop possible could have adequately treated 20 (36%) of the 56 patients in the acute phase. CONCLUSION: In this single-center imaging-based study, involvement of the LSA in the setting of acute type B aortic dissection was seen in 80% of patients treated with TEVAR. Three off-the-shelf stent-grafts would suffice to treat one-third of these acute type B aortic dissections and may offer a relatively simple solution to preserve LSA flow, thereby lowering the risk of malperfusion of the (posterior) cerebrum, spinal cord, and left arm in an urgent/emergent setting.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. van der Weijde E, Bakker OJ, Kamman AV, et al. A Feasibility Study of Off-the-Shelf Scalloped Stent-Grafts in Acute Type B Aortic Dissection. J Endovasc Ther. 2017. PMID: 28814171. doi:10.1177/1526602817726329

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。