新型即用型技术修复III/IV型胸腹主动脉瘤:Gore Excluder联合Viabahn分支
本研究描述一种在缺乏预制分支支架时,使用Gore C3 Excluder和Viabahn覆膜支架即用型修复III/IV型胸腹主动脉瘤的新技术。8例患者(平均78岁)接受修复,技术可行且安全,短期结果良好,无重大并发症,分支均通畅。
要点
- 该即用型技术可用于无预制分支装置时的III/IV型胸腹主动脉瘤修复。
- 8例患者中无重大心脏、呼吸、肾、神经或伤口并发症。
- 早期和内漏均经额外血管内或开放手术处理,分支均保持通畅。
核心结论
- 该即用型血管内技术修复III和IV型胸腹主动脉瘤可行且安全,短期结果良好。
研究设计与样本
- 前瞻性登记研究,纳入自2015年1月起接受该技术修复的III/IV型胸腹主动脉瘤患者,共8例(男性6例,平均年龄78岁)。
主要发现
- 所有患者使用Gore C3装置,其中5例修复3个内脏分支,3例修复4个内脏分支。
- 4例患者腹腔干被牺牲,1例患者一侧肾动脉被牺牲。
- 平均透视时间88.7分钟,平均对比剂用量92.3毫升。
- 中位住院时间7天,其中重症监护3天;无重大心脏、呼吸、肾、神经或伤口并发症。
- 3例早期内漏经额外血管内技术(2例)或开放手术结扎(1例)处理;2例晚期内漏(1例II型稳定,1例III型需髂支重新衬里);末次影像(平均6.8个月)所有分支和肢体通畅。
临床怎么用
- 在无预制分支装置时,可考虑采用Gore C3 Excluder联合Viabahn分支的即用型技术修复III/IV型胸腹主动脉瘤。
局限性
- 样本量小(8例),随访时间短(平均6.8个月),缺乏长期结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The aim of this study is to describe the use of a novel off-the-shelf technique to repair type III and type IV thoracoabdominal aortic aneurysms (TAAAs) in the absence of available prefabricated branched devices. METHODS: All patients undergoing endovascular repair of type III and IV TAAAs using this technique were included from a prospectively maintained registry at a regional aortic referral center. The proximal bifurcated Gore C3 Excluder device is positioned in the descending thoracic aorta with the contralateral gate 2-3 cm above the celiac artery. From an axillary approach, the contralateral gate renovisceral branches are sequentially cannulated and simultaneously stented using Viabahn covered stents. In cases were the celiac artery could not be excluded, a parallel stent (snorkel) was added adjacent to the proximal endograft. All branches are simultaneously balloon dilated to ensure proximal gutter seal in the contralateral gate. Via the ipsilateral limb, the device can then be extended with a flared iliac extension and/or additional bifurcated device to obtain seal in the distal aorta (previous open repair) or common iliac arteries. RESULTS: Eight patients (male = 6, mean 78 years of age) were treated in this manner since January 2015. All patients underwent repair using Gore C3 device with 3 (n = 5) or 4 (n = 3) renovisceral branches. The celiac artery was sacrificed in 4 patients and 1 renal artery in 1 patient. Mean fluoroscopy time was 88.7 min with a mean of 92.3 cc contrast utilized. Median length of stay was 7 days with 3 days spent in the intensive care unit. No major cardiac, respiratory, renal, neurologic, or wound complications occurred. Three patients had early endoleaks treated with additional endovascular techniques (n = 2) or open surgical ligation (n = 1) during the index hospitalization. Two late endoleaks were identified; 1 type II with stable sac size and 1 type III requiring iliac limb relining. All limbs and branches remain patent at the time of the last imaging study (mean 6.8 months). CONCLUSIONS: We present an endovascular technique for repair of type III and IV TAAAs which appears to be both feasible and safe with good short-term outcomes.
引用
- Wooster M, Tanious A, Jones RW, et al. A Novel Off-the-Shelf Technique for Endovascular Repair of Type III and IV Thoracoabdominal Aortic Aneurysms Using the Gore Excluder and Viabahn Branches. Ann Vasc Surg. 2017. PMID: 28689952. doi:10.1016/j.avsg.2017.06.155