半自动测量三维覆膜至肾动脉距离以判定EVAR后内移植物位置
本研究建立并验证了一种在主动脉曲线上计算内移植物覆膜与肾动脉间三维距离的方法,并在81例择期EVAR患者首次术后CT上测量最短距离(SFD)与对侧距离(CFD)。自动与手动计算的一致性极佳(ICC>0.892,p<0.001),仅44%的移植物相对最低肾动脉位于目标位置。
要点
- 三维CT重建可准确评估近端瘤颈内内移植物的位置与倾斜角度。
- 本组仅44%病例相对最低肾动脉达到目标位置。
- 自动与手动测量SFD、CFD的观察者内与观察者间一致性均极佳。
核心结论
- 基于三维CT重建的新方法可准确评估内移植物在近端主动脉瘤颈内的位置与倾斜度。
- 本系列中仅44%的内移植物相对最低肾动脉置于目标位置。
研究设计与样本
- 方法学研究,在81例择期EVAR患者首次术后CT扫描上测量SFD与CFD。
- SFD按位置分为目标位置(肾动脉远端0-3 mm)、高位(部分覆盖肾动脉)与低位(肾动脉远端>3 mm)。
- 数据以中位数(四分位距,IQR)报告。
主要发现
- SFD中位数1.4 mm(IQR -0.9,3.0),CFD中位数8.0 mm(IQR 3.9,14.2)。
- 目标位置占44%,高位占30%,低位占26%。
- 内移植物朝向较高肾动脉的倾斜角中位数2.5°(IQR -5.5°,13.9°)。
- 自动与手动计算SFD、CFD的观察者内与观察者间一致性极佳(ICC>0.892,p<0.001)。
临床怎么用
- 术后可用三维CT重建半自动测量覆膜至肾动脉距离,辅助评估EVAR释放精度。
- 测量结果可区分目标、高位与低位释放,为随访与再干预决策提供依据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To report a methodology for 3-dimensional (3D) assessment of the stent-graft deployment accuracy after endovascular aneurysm repair (EVAR). METHODS: A methodology was developed and validated to calculate the 3D distances between the endograft fabric and the renal arteries over the curve of the aorta. The shortest distance between one of the renal arteries and the fabric (SFD) and the distance from the contralateral renal artery to the fabric (CFD) were determined on the first postoperative computed tomography (CT) scan of 81 elective EVAR patients. The SFDs were subdivided into a target position (0-3 mm distal to the renal artery), high position (partially covering the renal artery), and low position (>3 mm distal to the renal artery). Data are reported as the median (interquartile range, IQR). RESULTS: Intra- and interobserver agreements for automatic and manual calculation of the SFD and CFD were excellent (ICC >0.892, p<0.001). The median SFD was 1.4 mm (IQR -0.9, 3.0) and the median CFD was 8.0 mm (IQR 3.9, 14.2). The target position was achieved in 44%, high position in 30%, and low position in 26% of the patients. The median slope of the endograft toward the higher renal artery was 2.5° (IQR -5.5°, 13.9°). CONCLUSION: The novel methodology using 3D CT reconstructions enables accurate evaluation of endograft position and slope within the proximal aortic neck. In this series, only 44% of endografts were placed within the target position with regard to the lowermost renal artery.
引用
- Schuurmann RCL, Overeem SP, Ouriel K, et al. A Semiautomated Method for Measuring the 3-Dimensional Fabric to Renal Artery Distances to Determine Endograft Position After Endovascular Aneurysm Repair. J Endovasc Ther. 2017. PMID: 28689482. doi:10.1177/1526602817719882
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。