EVAR中使用外扩髂支动脉直径≥20mm与晚期1b型内漏高风险相关
该回顾性分析纳入2006至2012年接受EVAR且有髂支支架类型及至少3年CT随访或早期发生1b型内漏的患者,比较远端直径≥20mm与<20mm髂支的晚期1b型内漏风险。
要点
- 髂支远端直径≥20mm者晚期1b型内漏风险是<20mm者的5.3倍。
- 平均随访37.7个月时,≥20mm组1b型内漏发生率为18%。
- ROC曲线提示19mm可作为预测1b型内漏的截断值。
核心结论
- 使用远端直径≥20mm的髂支的患者,晚期(平均37个月)1b型内漏风险是使用<20mm者的5倍。
研究设计与样本
- 回顾性分析2006至2012年在本机构接受EVAR的患者,纳入有髂支支架类型信息且至少3年增强CT随访或早期发生1b型内漏者。
- 共分析239条髂支(来自129例患者),分为远端直径<20mm组(178条)和≥20mm组(61条)。
主要发现
- 全组随访期间发生18例1b型内漏(7.5%),平均发生时间37.7个月(范围4-96个月)。
- ROC曲线分析提示髂支直径19mm为预测1b型内漏概率增高的截断值。
关键数值与推荐
- 髂支远端直径≥20mm定义为Group II,<20mm定义为Group I。
- 髂支平均oversizing为17.2%(IQR 14)。
- 无即刻1b型内漏发生。
临床怎么用
- 对于髂总动脉扩张需使用髂支的患者,应警惕远端直径≥20mm时晚期1b型内漏风险增高,需加强影像学随访。
局限性
- 回顾性单中心分析,样本量有限,仅纳入有完整随访信息的患者,可能存在选择偏倚。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Re-interventions are still the Achilles' heel of endovascular aneurysm repair (EVAR). Ectatic common iliac arteries (CIA) can be treated with flared iliac limbs but a dilated artery used as sealing zone could increase the risk of a late type 1b endoleak. The aim of the present analysis was to evaluate the risk of late type 1b endoleak after EVAR in patients with CIA limbs ≥20 mm compared with those <20 mm. METHODS: A retrospective analysis was performed of patients treated by EVAR at the study institution between 2006 and 2012, including patients with available information about the type of iliac stent grafts and a minimum follow-up (FU) of 3 years with contrast enhanced CT, or those that had developed a type 1b endoleak earlier. The cohort was divided into two groups: Group I included iliac limbs with a distal diameter <20 mm, and Group II included all iliac limbs with a distal diameter ≥20 mm. RESULTS: Of 692 limbs (346 patients), 239 limbs (34.5%) from 129 patients were included in the analysis. Mean CT FU was 53 months, 178 had an iliac stent graft diameter <20 mm (Group I), and 61 a diameter of ≥20 mm (Group II). Mean oversizing for iliac limbs was 17.2% (IQR 14) and there was no case of immediate type 1b endoleak. For the whole group, 18 type 1b endoleaks occurred during FU (7.5%) after a mean of 37.7 months (range 4-96). Eleven of 61 (18%) iliac limbs in Group II developed a type 1b endoleak during FU in contrast to 7/178 (3.9%) in Group I (OR 5.3, 95% CI 1.98-14.59, p = 0,001). The ROC curve analysis indicated a limb diameter of 19 mm as a cutoff value for a higher probability of developing a type 1b endoleak. CONCLUSION: Patients treated with iliac limbs ≥20 mm had a fivefold higher risk of late (mean 37 months) type 1b endoleak compared with patients treated with a distal iliac limb diameter <20 mm.
引用
- Gray D, Shahverdyan R, Reifferscheid V, et al. EVAR with Flared Iliac Limbs has a High Risk of Late Type 1b Endoleak. Eur J Vasc Endovasc Surg. 2017. PMID: 28687427. doi:10.1016/j.ejvs.2017.05.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。