Endologix覆膜支架置入后波浪现象——内漏模仿者的回顾性综述
波浪现象与不良事件、动脉瘤增大或内漏存在无关,应视为常见良性表现。
要点
- 波浪现象不增加并发症、死亡、动脉瘤增大或内漏风险。
- 需识别波浪现象以避免误诊为内漏。
核心结论
- 波浪现象似乎与发病率、死亡率增加、动脉瘤尺寸增大或内漏存在无关。
- 波浪现象应被认识为Endologix支架的常见良性表现。
研究设计与样本
- 回顾性研究,纳入2002年1月至2013年9月在本机构接受Endologix支架行腹主动脉瘤腔内修复并有影像随访的患者。
主要发现
- 平均波浪厚度4.2 mm(范围1-13 mm)。
- 波浪现象患者动脉瘤囊前后径和横径均显著缩小(P=0.014和P=0.009)。
临床怎么用
- 在Endologix支架置入后影像随访中应识别波浪现象,避免误判为内漏。
- 对于波浪现象可进行系列影像监测,多数稳定或改善。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Billowing is a unique computed tomography (CT) finding of the Endologix stent graft where a rim of contrast is observed outside the stent's endoskeleton but still contained within the graft cover. The purpose of this study is to evaluate the incidence and clinical outcome of billowing following an Endologix stent-graft placement. METHODS: All patients who underwent abdominal aortic aneurysm (AAA) repair with endovascular placement of an Endologix stent graft from January 2002 to September 2013 and who were followed with imaging were included in the study. The images were evaluated for the incidence of billowing. Demographics and clinical outcomes were also analyzed. RESULTS: One hundred twenty-two patients received an Endologix stent graft for AAA repair at our institution. One hundred patients qualified for this study. Eighty-four patients had billowing with the majority (70 patients) having crescentic billowing, and the remaining 14 patients having focal outpouching. Average billowing thickness measured 4.2 mm (range 1-13). Sixty-three patients with billowing had multiple follow-up CT angiograms to allow serial monitoring. Twelve patients had resolution of billowing, 11 patients had reduction in billowing size, 14 patients had no change in billowing size, and 26 patients had increase in billowing size. Patients with billowing had a statistically significant reduction in aneurysm sac size in anteroposterior and transverse dimensions (P = 0.014 and 0.009, respectively). Five cases of billowing were misdiagnosed as endoleak. Nineteen true endoleaks were identified, of which 13 cases were associated with billowing. CONCLUSIONS: Billowing does not seem to be associated with increased morbidity, mortality, increase in aneurysm size, or presence of endoleak. Billowing should be recognized as a common and benign finding of the Endologix stent.
引用
- Helo N, Chang AC, Hyun C, et al. Retrospective Review of Billowing Phenomenon-A Mimic of Endoleak Following Placement of Endologix Covered Stent for the Treatment of Abdominal Aortic Aneurysm. Ann Vasc Surg. 2017. PMID: 28688878. doi:10.1016/j.avsg.2017.06.127
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。