文献速递主动脉疾病

胸主动脉腔内修复术中血管内超声与中心线CT测定主动脉直径的比较

Comparison of intravascular ultrasound- and centerline computed tomography-determined aortic diameters during thoracic endovascular aortic repair.
PubMed 文献 · J Vasc Surg · 2017-06-22

CTA与IVUS测定的主动脉直径高度相关(R=0.62;P<.001),但Bland-Altman分析显示IVUS测值较CTA大,平均差异3.09 mm,差异标准差4.56 mm。在成角主动脉段一致性最低,平均差异可达7.96 mm,差异标准差可达8.27 mm。

要点

  • IVUS与中心线CTA测定的主动脉直径存在显著差异,IVUS测值偏大。
  • 在成角主动脉段(左锁骨下动脉起源远端2-7 cm)一致性最差。
  • 仅用CTA进行支架尺寸选择需谨慎,尤其近端锚定区成角时。

核心结论

  • IVUS与中心线CTA可能提供显著不同的主动脉直径测量值,尤其在成角主动脉段。

主要发现

  • CTA与IVUS测定的主动脉直径高度相关(R=0.62;P<.001)。
  • Bland-Altman分析显示IVUS测值较CTA大,平均差异3.09 mm,差异标准差4.56 mm。
  • 在左锁骨下动脉基部(位置0)一致性较高,平均差异-2.69 mm,差异标准差4 mm。
  • 在成角主动脉段(左锁骨下动脉起源远端2-7 cm)一致性最低,平均差异可达7.96 mm,差异标准差可达8.27 mm。

临床怎么用

  • 在TEVAR术中,尤其近端锚定区成角时,应谨慎仅依据CTA进行支架尺寸选择,可考虑IVUS辅助测量。

局限性

  • 回顾性研究,样本量较小(26例,最终20例分析)。
  • 仅纳入非创伤性主动脉病变患者,结果可能不适用于其他人群。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Accurate sizing of stent grafts during thoracic endovascular aortic repair (TEVAR) is critical for a successful outcome. Centerline measurements using three-dimensional reconstruction of preoperative computed tomography angiography (CTA) is the current standard for stent graft sizing. However, this technique is predicated on an idealized straightened aorta and does not account for the variability in the aortic diameter during the cardiac cycle or the overall status of the patient's volume. Intravascular ultrasound (IVUS) offers real-time cross-sectional imaging of the aorta orthogonal to the support wire, thereby providing an adjunctive method for aortic diameter determination at the time of TEVAR. METHODS: A retrospective review was performed on all patients who underwent TEVAR for nontraumatic aortic pathology from July 2015 to December 2015. Preoperative CTA images were reconstructed on a dedicated three-dimensional workstation. CTA centerline aortic diameter measurements were performed in major and minor axes at 1-cm intervals from the left subclavian origin to 20 cm distally. The IVUS images were acquired intraoperatively through 1-cm stepwise pullback along the aorta from the left subclavian origin to 20 cm. IVUS aortic diameters were measured at the maximum phase during the cardiac cycle. The average values of major and minor axes diameters from both modalities were calculated at each location for comparison. Linear regression analysis was performed to evaluate correlation, and Bland-Altman plots assessed agreement between different imaging modalities. RESULTS: During the study period, 26 patients underwent TEVAR. Of these, 20 patients had adequate CTA and IVUS images, providing 355 paired measurements. There was a high correlation between CTA- and IVUS-determined aortic diameters (R = 0.62; P < .001). However, Bland-Altman analysis showed that, compared with CT, IVUS resulted in larger aortic diameters, with the mean difference of 3.09 mm. There was a significant variability between IVUS and CTA with the standard deviation of difference (SD diff) of 4.56 mm. When stratified by the aortic position, a high degree of agreement was observed at the base of the left subclavian (position 0), with a mean difference of -2.69 mm and an SD diff of 4 mm. The agreement was the lowest at the angulated aortic segments (2 cm to 7 cm distal to the subclavian origin) with a mean difference up to 7.96 mm and an SD diff up to 8.27 mm. CONCLUSIONS: IVUS imaging and centerline CTA may provide significantly different aortic diameter measurements, particularly in angulated aortic segments. Caution must be taken when sizing a stent graft using CTA alone, particularly in an angulated proximal landing zone.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Han SM, Elsayed RS, Ham SW, et al. Comparison of intravascular ultrasound- and centerline computed tomography-determined aortic diameters during thoracic endovascular aortic repair. J Vasc Surg. 2017. PMID: 28648482. doi:10.1016/j.jvs.2017.03.445

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。