EVAR术中支架释放后即刻出现的IIIb型内漏一例
本文报道一例74岁男性肾下腹主动脉瘤患者,在择期EVAR术中新一代低剖面支架释放后即刻出现IIIb型内漏,经球囊阻断双侧支架分支并顺行主动脉造影确诊,采用转换支架联合髂动脉封堵器及股股交叉旁路成功治疗,术后4周超声无内漏。
要点
- IIIb型内漏可于支架释放后即刻出现,需与 lainnya 内漏类型鉴别。
- 双侧球囊阻断支架分支后顺行主动脉造影有助于定位IIIb型内漏。
- 近血流分叉处织物缺陷可采用主髂动脉支架联合股股交叉旁路作为开放手术替代方案。
核心结论
- 新一代低剖面支架释放后即刻出现的IIIb型内漏极罕见,可能源于支架装载或输送导管内受限时的织物损伤。
研究设计与样本
- 单例病例报告:74岁男性肾下腹主动脉瘤行择期EVAR。
主要发现
- 支架释放后即刻于血流分叉附近发现由织物缺陷导致的IIIb型内漏。
- 采用转换支架置入、左分支髂动脉封堵器闭塞及股股交叉旁路手术后内漏成功解决。
- 术后4周超声扫描未见内漏迹象。
关键数值与推荐
- 怀疑IIIb型内漏时,双侧球囊阻断支架双分支后顺行主动脉造影有助于识别漏口。
临床怎么用
- 术中怀疑IIIb型内漏时可行双侧球囊阻断加顺行主动脉造影明确诊断。
- 若织物缺陷靠近血流分叉,可考虑主髂动脉支架联合股股交叉旁路作为开放手术转换的替代方案。
局限性
- 单病例报告,不能确定织物缺陷具体发生于装载还是低压球囊操作过程。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Type IIIb endoleak usually occurs years after the initial endograft implantation, and the cause is the chronic fatigue of the endograft. This rare case describes a type IIIb endoleak, appearing immediately after deployment of a new generation low-profile stentgraft and highlights the diagnostic and treatment challenges associated with the type IIIb endoleak. A 74-year-old man underwent elective EVAR for an infrarenal abdominal aortic aneurysm. A type IIIb endoleak near to the flow divider due to a fabric defect was diagnosed. The endoleak was successfully treated by endovascular positioning of a converter stent graft followed by the occlusion of the left limb with an iliac occluder and a femoro-femoral crossover bypass surgery. The ultrasound scan after 4 weeks showed no sign of endoleak. Occurrence of a type IIIb endoleak immediately after deployment is extremely rare. Based on the convenience of the intraoperative procedure and the anatomic characteristics of the aneurysm, we assume that the fabric defect might have occurred during loading of the endograft and subsequent confinement in the delivery catheter. We cannot definitely rule out the possibility of fabric damage induced by low-pressure balloon instrumentation. In case of a suspicion of a type IIIb endoleak, bilateral balloon occlusion of both limbs followed by antergrade aortography will help to identify the leak. In case the defect is near to the flow divider, aortouniliac grafting followed by femoro-femoral crossover bypass surgery represents an alternative option to conversion to open surgical repair.
引用
- Moulakakis KG, Kakisis JD, Geroulakos G, et al. Intraoperative Type IIIb Endoleak after Endograft Deployment during EVAR. Ann Vasc Surg. 2017. PMID: 28479444. doi:10.1016/j.avsg.2017.04.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。