杂交手术室EVAR术中患者辐射风险即时图表
该回顾性研究分析74例EVAR手术中的16889次X线曝光,用多变量log线性混合模型预测术中辐射剂量率,发现透视时辐射协议是最强预测因子,DSA时C臂角度和BMI影响最大,并据此制作患者风险图表以预测达到剂量阈值前的允许曝光时间。
要点
- 透视辐射协议从低到中等使剂量率增加200%,低到正常增加410%
- DSA每增加30°C臂角度剂量率增加47%,BMI每增加5点增加58%
- BMI 30、45°旋转、800cm²视野下中等透视协议预测DRAK为0.39 mGy/s,达2Gy阈值需85.5分钟
核心结论
- X线辐射剂量率随BMI、协议、C臂位置和采集方式持续波动。
- 即时患者风险图表可可视化剂量波动并预测达到随访剂量阈值的时间。
研究设计与样本
- 回顾性分析74例EVAR手术,包含16889次X线曝光,使用固定C臂成像设备。
- 构建四个多变量log线性混合模型,以患者为随机效应。
主要发现
- 透视中DRAK和DRDAP最强预测因子为辐射协议,低到中等增加200%,低到正常增加410%。
- DSA中DRAK和DRDAP最强预测因子为C臂角度(每30°增加47%)和BMI(每5点增加58%)。
- BMI 30、45°旋转、800cm²视野下中等透视协议预测DRAK为0.39 mGy/s,达2Gy阈值需85.5分钟。
- 相同设置下切换至2帧/秒DSA协议,预测DRAK为6.6 mGy/s,达2Gy阈值需5.0分钟。
关键数值与推荐
- 透视DRAK阈值2 Gy,DRDAP阈值500 Gycm²。
- 辐射协议从低到中等DRAK和DRDAP增加200%,低到正常增加410%。
- DSA每30°C臂角度DRAK和DRDAP增加47%,BMI每5点增加58%。
临床怎么用
- 可根据患者BMI、C臂角度和透视协议使用风险图表预测达到剂量阈值前的允许曝光时间。
- 术中应关注辐射协议选择、C臂角度和BMI对剂量率的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To determine which patient and C-arm characteristics are the strongest predictors of intraoperative patient radiation dose rates (DRs) during endovascular aneurysm repair (EVAR) procedures and create a patient risk chart. METHODS: A retrospective analysis was performed of 74 EVAR procedures, including 16,889 X-ray runs using fixed C-arm imaging equipment. Four multivariate log-linear mixed models (with patient as a random effect) were constructed. Mean air kerma DR (DRAK, mGy/s) and the mean dose area product DR (DRDAP, mGycm2/s) were the outcome variables utilized for fluoroscopy as differentiated from digital subtraction angiography (DSA). These models were used to predict the maximum radiation duration allowed before a 2-Gy skin threshold (for DRAK) or a 500-Gycm2 threshold (for DRDAP) was reached. RESULTS: The strongest predictor of DRAK and DRDAP for fluoroscopy imaging was the radiation protocol, with an increase of 200% when changing from "low" to "medium" and 410% from "low" to "normal." The strongest predictors of DRAK and DRDAP for DSA were C-arm angulation, with an increase of 47% per 30° of angulation, and body mass index (BMI), with an increase of 58% for every 5-point increase in BMI. Based on these models, a patient with a BMI of 30 kg/m2, combined with 45° of rotation and a field size of 800 cm2 in the medium fluoroscopy protocol has a predicted DRAK of 0.39 mGy/s (or 85.5 minutes until the 2-Gy skin threshold is reached). While using comparable settings but switching the acquisition to a DSA with a "2 frames per second" protocol, the predicted DRAK will be 6.6 mGy/s (or 5.0 minutes until the 2-Gy threshold is reached). CONCLUSION: X-ray radiation DRs are constantly fluctuating during and between patients based on BMI, the protocols, C-arm position, and the image acquisitions that are used. An instant patient risk chart visualizes these radiation dose fluctuations and provides an overview of the expected duration of X-ray radiation, which can be used to predict when follow-up dose thresholds are reached during abdominal endovascular procedures.