文献速递主动脉疾病

超大自膨式镍钛支架使用需谨慎:一例近肾腹主动脉瘤腔内治疗失败报告

A Word of Caution for Extra Large Self-expandable Nitinol Stents.
PubMed 文献 · Ann Vasc Surg · 2017-04-05

本文报告一例82岁男性近肾腹主动脉瘤患者,采用超肾固定自膨式腔内移植物联合超大自膨式高径向力镍钛支架预防近端内漏,术后CT显示镍钛支架严重内折并可能穿孔原移植物,6个月随访瘤囊增大5 mm、体积增加6%。作者警示超大口径自膨式镍钛支架用于预防目的时应极其谨慎。

要点

  • 超大自膨式镍钛支架用于预防近端内漏可能导致支架内折和移植物穿孔。
  • 短且成角的瘤颈形态下,应审慎选择其他器械或手术方式。
  • 支架oversizing应仔细考虑,不一定达到移植物选定尺寸。

核心结论

  • 使用超大自膨式镍钛支架进行预防性处理时应极其谨慎,尤其当瘤颈短且成角时。

研究设计与样本

  • 个案报告:1例82岁男性近肾腹主动脉瘤患者,接受超肾固定自膨式腔内移植物联合超大自膨式高径向力镍钛支架治疗。

主要发现

  • 术后CT显示镍钛支架严重内折,并可能穿孔原腔内移植物。
  • 6个月随访瘤囊增大5 mm,体积增加6%。

关键数值与推荐

  • 瘤颈特征:长度7 mm,前后成角高。
  • ASA风险评分IV。
  • 瘤体直径80 mm。

临床怎么用

  • 对于短且成角的近肾瘤颈,应优先考虑其他器械或手术方式而非超大自膨式镍钛支架。
  • 使用自膨式镍钛支架时,oversizing应仔细评估,不一定达到移植物选定尺寸。

局限性

  • 仅为单个病例报告,结论推广受限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To report an unsuccessful use of large nitinol stent to prevent proximal endoleak in endograft treatment for juxtarenal aortic aneurysm. CASE REPORT: An 82-year-old male presented coronary disease and severe chronic obstructive pulmonary disease requiring oxygen support. A large 80-mm juxtarenal aneurysm was found on routine urological examination. Neck features were as follows: 7 mm and high anterior-posterior angulation. ASA risk score IV was given. Patient wanted to be treated. We decided a 1-shot intervention based on self-expandable suprarenal fixation endograft with adjunctive extra large self-expandable (high radial force) nitinol stent. Control computed tomography scan demonstrates nitinol stent severe infolding and a possible perforation of the primary endograft. At 6-month follow-up, sac reveals enlargement of 5 mm and 6% volume increase. Patient does not want further treatments due to high risk of procedures. CONCLUSIONS: Extreme caution should be taken when using self-expandable extra large nitinol stents for preventive matters. Oversizing should be carefully considered, not necessarily reaching the endograft chosen size. When short and angulated neck morphology co-exists, other devices or procedures may be a better option.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Reyes Valdivia A, Duque Santos A, Alvarez Marcos F, et al. A Word of Caution for Extra Large Self-expandable Nitinol Stents. Ann Vasc Surg. 2017. PMID: 28389291. doi:10.1016/j.avsg.2016.12.010