研究设计与样本

  • 单中心回顾性研究,纳入2007年11月至2016年3月122例主动脉弓病变患者(平均年龄50.4±12.7岁,范围29-80;92例男性)。
  • 101例采用单烟囱技术,21例采用双烟囱技术,重建无名动脉21条、左颈总动脉114条、左锁骨下动脉8条。

主要发现

  • 术后4天1例(0.8%)死于消化性溃疡穿孔;术中13例(10.7%)出现Ia型内漏,其中3例为双烟囱技术。
  • 6例(4.9%)发生II型内漏,3例左锁骨下动脉内漏使用封堵器治疗;1例(0.8%)双烟囱患者发生烟囱支架移位并追加支架。
  • 中位随访32.3个月,1例(0.8%)于术后3个月死于卒中;其余120例烟囱支架通畅。
  • 2例(1.7%)因远端主动脉夹层行二次TEVAR;随访中9例无症状Ia型内漏和1例II型内漏持续存在,1例Marfan综合征患者II型内漏在52个月时自愈。

临床怎么用

  • 对于不适合开放手术的主动脉弓病变,可考虑TEVAR联合烟囱技术重建弓上分支。
  • 双烟囱技术应严格选择病例,警惕内漏和支架移位风险。
  • 术后需规律CT随访(0.5、3、6、12个月及每年)以监测内漏和支架通畅情况。

局限性

  • 单中心回顾性研究,样本量有限,缺乏对照组。
  • 随访时间中期,长期结果有待进一步观察。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To summarize a single-center experience using the single/double chimney technique in association with thoracic endovascular aortic repairs (TEVAR) for aortic arch pathologies. METHODS: From November 2007 to March 2016, 122 patients (mean age 50.4±12.7 years, range 29-80; 92 men) with aortic arch pathologies underwent TEVAR combined with single (n=101) or double (n=21) chimney grafts to reconstruct the supra-aortic branches: 21 innominate arteries, 114 left common carotid arteries, and 8 left subclavian arteries (LSA). Pathologies included type B aortic dissection (n=47), aortic arch dissection (n=49), retrograde type A aortic dissection (n=8), thoracic aortic aneurysm (n=7), penetrating aortic arch ulcer (n=9), and post-TEVAR type I endoleak (n=2). Follow-up examinations included computed tomography at 0.5, 3, 6, and 12 months and yearly thereafter. RESULTS: The aortic stent-grafts were deployed in zone 0 (n=21), zone 1 (n=93), and zone 2 (n=8). One (0.8%) of the 122 patients died at 4 days due to a perforated peptic ulcer. Type Ia endoleaks were found intraoperatively in 13 (10.7%) patients, including 3 with the double chimney technique. Type II endoleaks occurred in 6 (4.9%) patients; 3 were treated with duct occluders in the LSA. Postoperative chimney graft migration occurred in 1 (0.8%) patient with double chimneys; additional stent-grafts were deployed in both chimneys. Median follow-up was 32.3 months, during which 1 (0.8%) patient died after a stroke at 3 months. Chimney stent-graft patency was observed in the remaining 120 patients. Two (1.7%) secondary TEVARs were performed for distal aortic dissection. Nine asymptomatic type Ia endoleaks and 1 type II endoleak persisted in follow-up; a type II endoleak in 1 patient with Marfan syndrome sealed in 52 months. CONCLUSION: TEVAR with the chimney technique provides a safe, minimally invasive alternative with good chimney graft patency and low postoperative mortality during midterm follow-up. The double chimney technique should be used judiciously owing to its potential complications.