三维超声半自动定量颈动脉斑块体积的算法验证

Semiautomatic quantification of carotid plaque volume with three-dimensional ultrasound imaging.
PubMed 文献 · J Vasc Surg · 2017-03-06

本研究在30例无症状≥50%颈动脉狭窄患者的三维超声数据集上,比较半自动斑块勾画算法与人工勾画的准确性、可重复性和耗时。结果显示半自动算法与人工勾画相关性好(r=0.76),几何相似性高(DSC 0.85),可重复性更优,能在95%置信度下检测4%-6%的斑块体积变化,全算法耗时约14分钟。

要点

  • 半自动算法与人工勾画斑块体积相关性好(r=0.76),几何相似性高(DSC 0.85)。
  • 全算法可重复性优于人工,能检测低至4%-6%的斑块体积变化(95%置信度)。
  • 全算法耗时约14分钟,中间阶段仅需7分钟但可重复性略低。

核心结论

  • 半自动算法可快速、准确、可重复地定量三维超声颈动脉斑块体积,适用于临床和纵向研究。

研究设计与样本

  • 研究纳入30例无症状≥50%颈动脉狭窄患者的三维超声数据集,由两名观察者实施半自动分割算法并与人工勾画比较。

主要发现

  • 半自动算法与人工勾画斑块体积的Pearson相关系数为0.76(P < .001),DSC为0.85,MHD为0.48 mm。
  • 观察者间变异:ICC 0.9,CV 8.22%,MDC 5.57%,SEM 1.45%,DSC 0.88,MHD 0.43 mm。
  • 观察者内变异:ICC 0.9,CV 8%,MDC 3.62%,SEM 1.31%,DSC 0.89,MHD 0.37 mm。
  • 中间阶段算法与全算法结果匹配(r=0.76,DSC 0.84,MHD 0.52 mm),但可重复性较低(观察者间ICC 0.81,CV 11.7%)。
  • 全算法耗时约14分钟,中间阶段约7分钟。

关键数值与推荐

  • 半自动算法可在95%置信度下可靠检测低至4%至6%的斑块体积变化。

临床怎么用

  • 该算法可用于临床环境中快速定量颈动脉斑块体积,并适用于追踪斑块进展的纵向研究。

局限性

  • 中间阶段算法虽快(7分钟)但可重复性和可靠性低于全算法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Vessel wall volume (VWV) assessed by three-dimensional duplex ultrasound (3DUS) imaging provides a more comprehensive measure of plaque burden than conventional two-dimensional measures of diameter stenosis. We previously demonstrated that manual outlining of the arterial lumen-intima boundary and outer wall boundary can be performed reliably on images obtained with a commercially available 3D-DUS transducer. Manual segmentation, however, is time consuming (∼45 minutes), limiting its clinical translation. We have developed a semiautomatic algorithm (manual selection of the carotid bifurcation image with subsequent automatic plaque outlining) to outline carotid plaques on 3DUS data sets. In this study, we investigated the accuracy, reproducibility, reliability, and time taken by this algorithm. METHODS: 3DUS data sets from 30 patients with asymptomatic ≥50% carotid stenosis underwent manual outlining of lumen-intima boundary and outer wall boundary to measure VWV. Two observers implemented a semiautomatic segmentation algorithm. The algorithm's accuracy was compared with manual outlining using the Pearson correlation coefficient. The Dice similarity coefficient (DSC) and modified-Hausdorff distance (MHD) were used to quantify the geometric similarity of the outlines. We also compared results after an intermediate stage of the algorithm vs the complete algorithm. Reproducibility and the least amount of detectable change in plaque volume were computed for each method. Intraobserver and interobserver metrics for each method were computed using the intraclass correlation coefficient (ICC), coefficient of variability (CV), minimum detectable change (MDC), and standard error of measurement (SEM) of the VWV. RESULTS: Plaque volume estimates obtained from the semiautomatic algorithm were accurate compared with manual outlining. The Pearson correlation coefficient was 0.76 (P < .001), and measurements were geometrically similar (DSC, 0.85; MHD, 0.48 mm). The algorithm was more reproducible and reliable and could detect smaller changes in plaque volume on repeat imaging (low interobserver variability: ICC, 0.9; CV, 8.22%; MDC, 5.57%; SEM, 1.45%; DSC, 0.88; MHD, 0.43 mm). Intraobserver variability was even lower (ICC, 0.9; CV, 8%; MDC, 3.62%; SEM, 1.31%; DSC, 0.89; MHD, 0.37 mm). Plaque volume estimates at the intermediate stage of the algorithm matched results from the full algorithm (Pearson correlation coefficient, 0.76; DSC, 0.84; MHD, 0.52 mm). The intermediate approach, however, was less reliable than the full algorithm (interobserver: ICC, 0.81; CV, 11.7%; MDC, 9.58%; SEM, 3.46%; DSC, 0.88; MHD, 0.42 mm; intraobserver: ICC, 0.87; CV, 8.6%; MDC, 4.55%; SEM, 1.64%; DSC, 0.89; MHD, 0.38 mm). The full algorithm required ∼14 minutes to implement. However, a quick (7 minutes) and accurate assessment of VWV can be obtained by running only the intermediate stage of the algorithm, although with a loss in repeatability and reliability. CONCLUSIONS: We present a unique algorithm to perform semiautomatic quantification of carotid plaque volume using 3DUS imaging. It is quick (mean time, 14 minutes), accurate, repeatable, and implementable in a clinical environment and in longitudinal studies tracking plaque progression. It reliably detects plaque volume changes as low as 4% to 6% with 95% confidence.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Khan AA, Koudelka C, Goldstein C, et al. Semiautomatic quantification of carotid plaque volume with three-dimensional ultrasound imaging. J Vasc Surg. 2017. PMID: 28274755. doi:10.1016/j.jvs.2016.11.033