文献速递主动脉疾病

Nellix近端套接延伸治疗晚期支架向尾侧移位的可行性与技术要点

Feasibility and Technical Aspects of Proximal Nellix-in-Nellix Extension for Late Caudal Endograft Migration.
PubMed 文献 · J Endovasc Ther · 2016-11-19

该研究在体外和临床中评估Nellix-in-Nellix近端延伸治疗EVAS后支架尾侧移位继发Ia型内漏的可行性。体外发现暴露endobag长度10-20mm密封不良,30-40mm可实现造影密封;5例临床患者均成功完成手术,无手术相关并发症,中位随访12个月无内漏。

要点

  • Nellix-in-Nellix近端延伸可用于治疗尾侧移位继发的Ia型内漏。
  • 体外显示暴露endobag长度30-40mm才能实现血管造影密封。
  • 临床5例均需烟囱移植物创建近端锚定区,手术全部成功。

核心结论

  • Nellix-in-Nellix近端延伸可治疗尾侧移位的Nellix支架及继发Ia型内漏,但技术不同于初次EVAS。

研究设计与样本

  • 体外研究评估12mm球囊后扩张后内径及不同暴露长度(10、20、30、40mm)的贴壁性,并用硅胶模型和搏动流泵模拟。
  • 临床部分为2中心5例患者(中位年龄74岁,范围73-83),因尾侧移位(左侧中位9mm,范围7-15;右侧中位7mm,范围0-11)继发Ia型内漏接受近端Nellix延伸。

主要发现

  • 体外:12mm球囊后扩张后Nellix-in-Nellix延伸内径一致;暴露endobag 10和20mm密封不良伴内漏,30和40mm实现造影密封。
  • 体外:第二个Nellix的填充线压力高于预期。
  • 临床:5例均需烟囱移植物创建足够近端锚定区,Nellix-in-Nellix手术全部成功,无手术相关并发症。
  • 临床:中位12个月随访无内漏;2例分别因支架内狭窄和烟囱移植物受压再次干预。

关键数值与推荐

  • 初次EVAS中位聚合物填充体积42.5mL(范围25-71)。
  • 再次干预中位时间15个月(范围13-32)。

临床怎么用

  • 对于EVAS后尾侧移位继发Ia型内漏,可考虑近端Nellix-in-Nellix延伸,但需注意暴露endobag长度至少30mm以确保密封。
  • 临床操作中常需联合烟囱移植物以获得足够近端锚定区,并注意第二个Nellix填充线压力可能升高。

局限性

  • 临床样本量小(5例),随访时间有限(中位12个月)。
  • 体外模型可能不能完全模拟体内条件。
  • 缺乏专用近端延伸装置,技术操作与初次EVAS不同。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To describe the feasibility and technical aspects of a proximal Nellix-in-Nellix extension to treat caudal stent-graft migration after endovascular aneurysm sealing (EVAS) in the in vitro and in vivo settings. METHODS: In vitro studies were designed (1) to assess inner diameters of Nellix-in-Nellix extensions after postdilation with 12-mm balloons and (2) to test wall apposition in tubes with different diameters using a Nellix-in-Nellix stent-graft that extended out of the original Nellix stent-graft lumen by 10, 20, 30, and 40 mm. Simulated-use experiments were performed using silicone models in conjunction with a pulsatile flow pump. In the clinical setting, 5 patients (median age 74 years, range 73-83) presented at 2 centers with type Ia endoleak secondary to caudal Nellix stent-graft migration measuring a median 9 mm (range 7-15) on the left and 7 mm (range 0-11) on the right. Median polymer fill volume at the initial EVAS procedure was 42.5 mL (range 25-71). The median time to reintervention with a proximal Nellix extension was 15 months (range 13-32). RESULTS: In vitro, the inner diameters of the Nellix-in-Nellix extensions were consistent after postdilation. Cases with 10 and 20 mm of exposed endobag resulted in a poor seal with endoleak, while cases with 30 and 40 mm of exposed endobag length exhibited angiographic seal. Fill line pressures of the second Nellix were higher than expected. In the 5 clinical cases, chimney grafts were required in each case to create an adequate proximal landing zone. The Nellix-in-Nellix procedure was successful in all patients. There were no procedure-related complications, and no endoleaks were observed during a median 12-month follow-up. Reinterventions were performed in 2 patients because of in-stent stenosis and chimney graft compression, respectively. CONCLUSION: Proximal Nellix-in-Nellix extension can be used to treat caudally migrated Nellix stent-grafts and to treat the consequent type Ia endoleak, but the technique differs from primary EVAS. The development of dedicated proximal extensions is desirable.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Donselaar EJ, Holden A, Zoethout AC, et al. Feasibility and Technical Aspects of Proximal Nellix-in-Nellix Extension for Late Caudal Endograft Migration. J Endovasc Ther. 2016. PMID: 27864459. doi:10.1177/1526602816677037

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。