BlueDop血管专家仪在外周动脉疾病诊断中的准确性
本研究为横断面队列研究,纳入217名疑似间歇性跛行患者共430条肢体,比较BlueDop血管专家仪(BVE)与手动踝臂指数(ABI)/趾臂指数(TBI)的诊断准确性。以手动ABI为参考,BVE的敏感度88.5%、特异度80.7%;BVE评估显著快于手动方法(p<0.01)。
要点
- BVE与手动ABI相比敏感度88.5%、特异度80.7%,阴性预测值88.1%。
- BVE诊断仅基于波形者占71%,基于ABI与波形组合者占29%。
- BVE评估速度显著快于手动方法(p<0.01),但需更多人群验证。
核心结论
- BVE在检测或排除PAD方面与手动ABI相比表现出良好的敏感度和特异度。
- 在常规临床实践应用前,需在不同临床人群和环境中与其他设备进一步比较。
研究设计与样本
- 横断面队列研究,纳入转诊至护士主导的间歇性跛行门诊的患者。
- 共纳入217名参与者、430条肢体,总体PAD患病率为46%。
- BVE测量与手动静息ABI/TBI方法比较,分别采用传统截断值<0.9和<0.7。
主要发现
- 与手动ABI相比,BVE的敏感度88.5%、特异度80.7%、阳性预测值81.3%、阴性预测值88.1%。
- 与手动TBI相比(仅38条肢体),BVE的敏感度45.5%、特异度88%、阳性预测值62.5%、阴性预测值78.6%。
- 无BVE诊断仅基于ABI,71%仅基于波形,29%基于ABI与波形组合。
- BVE评估显著快于手动方法(p<0.01)。
关键数值与推荐
- 手动ABI传统截断值<0.9,手动TBI传统截断值<0.7。
- BVE与手动ABI比较的敏感度88.5%、特异度80.7%、PPV 81.3%、NPV 88.1%。
- BVE与手动TBI比较的敏感度45.5%、特异度88%、PPV 62.5%、NPV 78.6%(仅38条肢体)。
临床怎么用
- BVE可作为疑似间歇性跛行患者中快速筛查PAD的辅助工具。
- BVE诊断多基于波形(71%),解读时应结合波形与ABI组合信息。
- 在常规临床推广前,需在不同临床人群和环境中进一步验证。
局限性
- 与手动TBI的比较仅纳入38条肢体,样本量小。
- 单中心研究,转诊至护士主导门诊的人群可能限制外推性。
- 需与其他设备及不同临床环境比较后方可推荐常规使用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Manual ankle-brachial pressure index (ABI) assessment requires significant time and expertise, precluding its use in primary care. Automated vascular assessment tools may offer a simple and efficient alternative for detecting peripheral arterial disease (PAD). However, further evidence is needed before they can be recommended. This study considers the diagnostic accuracy of the BlueDop Vascular Expert (BVE) in people with suspected intermittent claudication. METHODS: We performed a cross-sectional cohort study including people referred to our nurse-led intermittent claudication clinic. BVE measurements were compared to the manual resting ABI/TBI method, using traditional cut-off values of <0.9 and <0.7, respectively. RESULTS: 217 participants and 430 limbs were included, with an overall PAD prevalence of 46%. Sensitivity, specificity, positive (PPV) and negative predictor values (NPV) for the BVE were 88.5%, 80.7%, 81.3% and 88.1% respectively, when compared to the manual ABI. When compared to the manual TBI, the sensitivity, specificity, PPV and NPV for the BVE were 45.5%, 88%, 62.5% and 78.6% respectively, though this analysis included only 38 limbs. No BVE diagnoses were based on ABI alone, 71% were based on waveform alone and 29% were based on a combination of ABI and waveform. The BVE assessment was significantly quicker to perform than the manual method (p<0.01). CONCLUSION: The BVE demonstrated favourable sensitivity and specificity for detecting or excluding PAD compared to the manual ABI. Further work is needed comparing this device to others, across a range of clinical populations and settings before it can be considered in routine clinical practice.
引用
- Prosser J, Hatfield J, Roe T, et al. Diagnostic accuracy of the BlueDop Vascular Expert for detecting peripheral arterial disease in people with suspected intermittent claudication referred to secondary care: A cross-sectional study. Ann Vasc Surg. 2026. PMID: 42874854. doi:10.1016/j.avsg.2026.09.054
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。