Endostaple配置与放大率对内移植物移位力的影响:Aortoseal系统体外评估

Effect of Endostaple Configuration and Oversizing on Endograft Displacement Force: An In Vitro Evaluation of the Aortoseal Endostapling System.
PubMed 文献 · J Endovasc Ther · 2026-10-30

该体外研究使用硅胶管模型评估Aortoseal Endostapling系统不同激活钉配置(11种)与内移植物放大率(10%、20%、30%)对内移植物移位力(EDF)和拔出力(pull-out force)的影响。结果显示观察者间ICC为0.987,EDF范围8.8±0.15 N至39.6±2.96 N,主要受激活钉数量和最大未钉合周向角度(MUCA)影响,6钉配置EDF最高,且除1钉配置外拔出力均超过EDF。

要点

  • Aortoseal框架单独置入即可在10%和20%放大率下显著提高EDF。
  • 激活钉的周向分布(MUCA)是EDF的主要决定因素,EDF随MUCA增大而降低、随钉数增加而升高。
  • 体外模型中拔出力在所有配置(除1钉)下均超过EDF。

核心结论

  • 体外硅胶管模型中,Aortoseal框架置入本身在10%和20%放大率下较单纯移植物显著提高EDF。
  • 激活钉的周向分布是EDF的主要决定因素,EDF随MUCA增大而降低、随钉数增加而升高。
  • 10%放大率在多数配置下产生的EDF与20%放大率相当。

研究设计与样本

  • 体外研究,使用硅胶管模型评估Aortoseal Endostapling系统。
  • 测试11种激活钉配置与10%、20%、30%内移植物放大率的组合。
  • 评估指标为2 mm内移植物移位时的EDF和拔出力,观察者间一致性采用ICC(双向混合效应模型绝对一致性)。

主要发现

  • 观察者间ICC为0.987。
  • EDF范围8.8 N ± 0.15 N至39.6 N ± 2.96 N(所有配置和放大率条件下)。
  • Aortoseal框架单独置入(不激活钉)在10%和20%放大率下较无框架显著提高EDF。
  • 6钉激活配置始终产生最高EDF;30%放大率在多数配置下EDF较低。
  • 除1钉配置外,所有测试配置的拔出力均超过EDF。

关键数值与推荐

  • EDF范围8.8 N ± 0.15 N至39.6 N ± 2.96 N。
  • 观察者间ICC为0.987。
  • 6钉激活配置EDF最高;拔出力在除1钉配置外均超过EDF。

临床怎么用

  • 研究结果为术前规划提供机制性参考框架,尤其适用于解剖挑战性病例。
  • 周向钉分布(MUCA)和钉数量是影响固定效果的关键参数,应予以关注。
  • 10%与20%放大率的EDF相当,提示放大率选择需结合具体钉配置综合考量。

局限性

  • 本研究为体外硅胶管模型,不能完全反映体内生物力学环境。
  • 结论需进一步实验和临床评估验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Proximal endograft fixation remains a key determinant of long-term durability of endovascular aortic repair (EVAR). The Aortoseal Endostapling system consists of a nitinol frame with 6 welded endostaples at the proximal end, intended to enhance fixation and seal of an endograft by mimicking the hand-sewn anastomosis of open aortic repair. This study evaluates the effect of different distributions of activated endostaples on endograft displacement force (EDF), while considering various endograft oversizing options. METHODS: An in vitro silicone tube model was used to assess EDF at 2 mm endograft displacement and pull-out force across 11 activated staple configurations of the Aortoseal Endostapling system with 10%, 20%, and 30% endograft oversizing. Interobserver agreement was assessed using the intraclass correlation coefficient (ICC), based on a 2-way mixed-effects model for absolute agreement. RESULTS: Intraclass correlation coefficient for interobserver agreement was 0.987. Endograft displacement force ranged from 8.8 N ± 0.15 N to 39.6 N ± 2.96 N across all configurations and oversizing conditions. Incorporation of the Aortoseal frame alone, without staple activation, substantially increased EDF compared with no Aortoseal frame at 10% and 20% oversizing. Furthermore, EDF was mostly influenced by the number of activated staples and the maximum unstapled circumferential angle (MUCA). The 6-activated-staple configuration consistently yielded the highest EDF. The effect of oversizing on EDF was inconsistent across configurations at 10% and 20%, whereas 30% oversizing was associated with lower EDF in most configurations. Pull-out force exceeded EDF in all tested configurations except the 1-activated-staple configuration. CONCLUSION: In this in vitro study, Aortoseal frame incorporation alone substantially increased EDF relative to the graft-only configuration at 10% and 20% oversizing. Ten percent endograft oversizing yielded EDF values comparable with 20% oversizing across most configurations. Circumferential staple distribution was the primary determinant of EDF in the Aortoseal Endostapling system, with EDF decreasing with increasing MUCA and increasing with staple number across all endograft oversizing conditions, consistent with circumferential load sharing as a potential underlying mechanism. These findings may serve as a mechanistic reference framework for preoperative planning in anatomically challenging cases and warrant further experimental and clinical evaluation.Clinical ImpactThe Aortoseal Endostapling system (Endoron Medical Holdings Ltd., Dublin, Ireland) represents a novel approach to proximal endograft fixation, enabling simultaneous circumferential single-stage activation of 6 nitinol endostaples via balloon activation, providing a uniform and reproducible fixation pattern intended to mimic the longitudinal and radial anchoring of a surgical anastomosis. To support clinical implementation, this study systematically evaluated endograft displacement force across a range of activated staple configurations, including suboptimal deployment scenarios. The findings provide physicians with a mechanistic reference framework for preoperative planning, supporting procedural decision-making in anatomically challenging cases.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Pierie SLR, Ligtenberg LW, David DB, et al. Effect of Endostaple Configuration and Oversizing on Endograft Displacement Force: An In Vitro Evaluation of the Aortoseal Endostapling System. J Endovasc Ther. 2026. PMID: 42961093. doi:10.1177/15266028261494945

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。