新型电磁线圈(EMAG)开窗支架覆膜系统用于快速插管以减少开窗EVAR中的造影剂使用和辐射暴露
本研究设计并制作了一种新型电磁(EMAG)系统,通过磁力吸引导丝或导管尖端快速选择性插管主动脉分支血管,以缩短开窗EVAR中的插管时间。对比研究表明,EMAG系统平均插管时间为5.73±2.67秒,显著短于标准导丝的34.37±39.91秒(P<.01)。该装置有潜力减少EVAR过程中的辐射暴露,提高患者安全性。
要点
- EMAG系统通过磁力吸引快速选择性插管分支血管,平均插管时间5.73±2.67秒。
- 标准导丝平均插管时间34.37±39.91秒,EMAG显著更短(P<.01)。
- 该装置有潜力减少开窗EVAR中的辐射暴露,尚需进一步临床研究。
核心结论
- EMAG系统可显著缩短开窗EVAR中分支血管插管时间,有潜力减少辐射暴露。
研究设计与样本
- 设计并制作了EMAG原型系统,并进行EMAG系统与标准导丝的对比研究。
- 研究评估了EMAG装置的吸引力、脱位力和温度升高特性。
主要发现
- EMAG系统平均插管时间为5.73±2.67秒,显著短于标准导丝的34.37±39.91秒(P<.01)。
- 研究表征了EMAG装置的吸引力、脱位力和温度升高。
临床怎么用
- EMAG系统有潜力改善开窗EVAR工作流程、提高操作一致性并减少透视依赖。
- 需进一步开展临床前和临床研究以推进该技术应用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Endovascular aortic repair (EVAR) has emerged as a popular option for minimally invasive treatment of abdominal and thoracic aortic aneurysms, offering advantages over traditional open surgical methods by reducing immediate postoperative morbidity and mortality. Although the endovascular repair option is attractive, it can potentially increase patient risks associated with x-ray exposure during the procedure, as significant levels of ionizing radiation are generated during preoperative diagnostic angiography, perioperative device placement, and lifelong follow-up imaging. For treatment of complex aortic aneurysms involving visceral vessels, radiation exposure during fenestrated stent-graft repair is typically double that of conventional unfenestrated stent-graft delivery and placement, which puts patients and clinicians at risk of reaching lifetime radiation limits and potentially developing cancer. To address this issue, a novel electromagnetic (EMAG) system has been developed and studied to potentially reduce radiation overexposure during the procedures by expediting a key step of the process, namely cannulation of fenestrations. The solenoid electromagnetic EMAG system embedded surrounding the ostia of the fenestrations attracts the magnetized tip of either a guidewire or catheter into a side branch, rapidly and selectively cannulating branch vessels in the aorta. The EMAG prototype was designed and fabricated with selective design requirements. A comparative study between the EMAG system and a standard guidewire demonstrated that the average time to cannulate with the EMAG system was 5.73 ± 2.67 seconds, which was significantly less than the average time taken to cannulate with the standard guidewire (34.37 ± 39.91 seconds, P < .01). In addition, the feasibility of the EMAG device was evaluated by characterizing its attraction force, dislodging force, and temperature rise. This device has the potential to reduce radiation exposure during EVAR procedures, offering a promising approach to improve patient safety in an increasingly common procedure.Clinical ImpactThe findings of this study demonstrate the feasibility of the EMAG system as a novel approach for assisting branch-vessel cannulation during fenestrated endovascular aortic repair. By advancing this technology through further preclinical and clinical studies, the EMAG system has the potential to improve workflow, enhance procedural consistency, reduce fluoroscopy dependence, and contribute to safer and more efficient complex endovascular interventions.
引用
- Tillman BW, Bryson GK, Go CC, et al. A Novel Electromagnetic Coil (EMAG) Fenestrated Stent-Graft for Rapid Cannulation to Reduce Use of Contrast Agent and Radiation Exposure During Fenestrated EVAR. J Endovasc Ther. 2026. PMID: 42557869. doi:10.1177/15266028261471720
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。