球囊扩张覆膜支架作为复杂EVAR中的髂支:可行性与结局
本研究前瞻性纳入64例复杂EVAR患者共89条髂动脉,因髂着陆区直径<16 mm或闭塞/迂曲不适合标准自膨式髂支而使用球囊扩张覆膜支架(BECS)作为髂支。中位随访19.7个月,主要通畅率98.9%,免于再干预率93.8%,I b型内漏3例(3.4%),III型内漏1例(1.6%)。BECS是精选小或病变髂动脉患者中可行的髂支策略。
要点
- 髂着陆区直径<16 mm或闭塞/迂曲不适合标准髂支时,可考虑BECS作为髂支。
- 中位随访19.7个月,主要通畅率98.9%,免于再干预率93.8%。
- I b型内漏3例(3.4%),III型内漏1例(1.6%),远端密封良好。
核心结论
- BECS可作为精选小或病变髂动脉、股动脉入路不佳患者EVAR中可行的髂支替代策略。
- 本系列中BECS与高中期髂支通畅率、低再干预率及良好远端密封相关。
研究设计与样本
- 前瞻性、非随机、研究者发起的IDE研究,2017至2025年纳入64例复杂EVAR患者共89条髂动脉。
- 当髂着陆区直径<16 mm或闭塞/迂曲不适合标准自膨式EVAR髂支时使用BECS(VBX)作为髂支。
- 排除BECS与髂分支装置联用或用于加固/延长市售髂支的病例。
主要发现
- 患者平均年龄74.0岁(范围61-89岁),女性占54.7%。
- 平均髂着陆区直径10.6 mm(范围5.6-15.6 mm)。
- 15条髂动脉(16.9%)经腋动脉入路释放BECS,74条(83.1%)经股动脉入路。
- 57条髂动脉(64%)使用1枚BECS,29条(32.6%)使用2枚,3条(3.4%)使用3枚;最常用直径11 mm(62条,69.7%)。
- 中位随访19.7个月(范围0.8-60.2),主要通畅率98.9%,免于再干预率93.8%;I b型内漏3例(3.4%),III型内漏1例(1.6%)。
临床怎么用
- 对于髂着陆区直径<16 mm或闭塞/迂曲不适合标准自膨式髂支的复杂EVAR患者,可考虑使用BECS作为髂支。
- BECS平台低profile输送系统、精准释放、良好柔顺性和强径向支撑力使其尤其适合此应用。
局限性
- 需要更长随访和对比研究以进一步明确长期耐久性和最佳患者选择。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Calcified, stenotic, or occluded iliac and femoral arteries can pose a challenge for iliac limb delivery and deployment during endovascular aneurysm repair (EVAR). The purpose of this study is to evaluate the feasibility and outcomes of balloon-expandable covered stents (BECSs) used as iliac limbs during complex EVAR. METHODS: From 2017 to 2025, BECS (VBX; W.L. Gore & Associates, Flagstaff, AZ) were used as an iliac limb for 89 iliac arteries in 64 patients undergoing complex EVAR in a prospective, nonrandomized, investigator-sponsored Investigational Device Exemption study. A BECS was used as an iliac limb when the iliac landing zone diameter was <16 mm in patients with occlusive disease or tortuosity that was deemed unfavorable for treatment using a standard self-expanding EVAR limb. Cases in which a BECS was used with an iliac branch device or to reinforce or extend commercially available iliac limbs were excluded. RESULTS: The mean age of patients treated with a BECS as an iliac limb was 74.0 years (range: 61-89 years), and 54.7% were women. The mean iliac landing zone diameter was 10.6 mm (range: 5.6-15.6 mm). BECSs were delivered via axillary artery access for 15 iliac arteries (16.9%) and were delivered via femoral artery access for the remaining 74 iliac arteries (83.1%). A single BECS was used in 57 iliac arteries (64%), 2 BECSs were used in 29 iliac arteries (32.6%), and 3 BECSs were used in 3 iliac arteries (3.4%). The most commonly used BECS size was 11 mm in diameter (62 iliac arteries; 69.7%). BECSs were implanted into 71 native iliac arteries and into 18 iliac limbs from prior EVARs or prosthetic surgical grafts. The median follow-up was 19.7 months (range: 0.8-60.2). Primary patency was 98.9%, and freedom from reintervention was 93.8%. Three patients (3.4%) had a type 1b endoleak, and one patient (1.6%) had a type III endoleak on follow-up imaging. CONCLUSION: BECS can serve as a feasible and promising alternative iliac limb strategy during EVAR in selected patients with small or diseased iliac arteries or unfavorable femoral access. In this series, the use of BECS was associated with high, intermediate-term limb patency, low reintervention rates, and favorable distal sealing despite challenging iliac anatomy. The low-profile delivery system, precise deployment, favorable flexibility, and strong radial force of the VBX platform make it particularly well suited for this application. Longer follow-up and comparative studies are needed to further define long-term durability and optimal patient selection.