文献速递主动脉疾病

LumiGuide与标准FORS导丝在开窗分支腔内修复术中的结局与辐射暴露比较

Comparison of outcomes and radiation exposure using Fiber-Optic RealShape-enabled standard guidewire versus LumiGuide for fenestrated-branched endovascular aortic aneurysm repair.
PubMed 文献 · J Vasc Surg · 2026-06-22

本研究比较了标准FORS导丝与LumiGuide在开窗分支腔内修复术(F/B-EVAR)中的技术成功率和辐射暴露。纳入218例F/B-EVAR(129例标准FORS,89例LumiGuide),共931次目标血管插管。LumiGuide组FORS技术成功率更高(79.4% vs 56.2%,P<.001),手术时间、透视时间和对比剂剂量显著降低,但每目标血管导航时间和DAP无显著差异。

要点

  • LumiGuide提高FORS技术成功率至79.4%,标准FORS为56.2%。
  • LumiGuide显著缩短手术时间(257±91 vs 298±100分钟,P=.002)。
  • LumiGuide显著减少透视时间(58±38 vs 80±40分钟,P=.01)和对比剂剂量(130±40 vs 150±50 mL,P=.05)。

核心结论

  • LumiGuide与3D-Hub的应用提高了F/B-EVAR中FORS技术成功率并缩短手术时间。
  • LumiGuide可使用不同市售导管,提升便利性并降低辐射暴露,从而提高效率和安全性。

研究设计与样本

  • 回顾性比较研究,纳入218例F/B-EVAR(129例标准FORS,89例LumiGuide),共931次目标血管插管(590次标准FORS,341次LumiGuide)。
  • 技术成功定义为仅使用FORS成功完成目标血管插管。
  • 连续变量采用Welch t检验,比例采用z检验。

主要发现

  • FORS技术成功率:标准FORS组56.2%,LumiGuide组79.4%(P<.001)。
  • 每目标血管导航时间:LumiGuide组6.4±7.4分钟,标准FORS组6.7±9.6分钟(P=.6)。
  • 每目标血管DAP:LumiGuide组3.2±5.3 Gy·cm2,标准FORS组3.36±9.3 Gy·cm2(P=.79)。
  • 手术时间:LumiGuide组257±91分钟,标准FORS组298±100分钟(P=.002)。
  • 透视时间:LumiGuide组58±38分钟,标准FORS组80±40分钟(P=.01)。

临床怎么用

  • 在F/B-EVAR中可优先考虑使用LumiGuide与3D-Hub,以提高目标血管插管成功率并减少手术时间、透视时间和对比剂剂量。
  • LumiGuide允许使用不同市售导管,可能提升操作便利性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Target-vessel (TV) catheterization during fenestrated-branched endovascular aortic aneurysm repair (FBEVAR) relies primarily on fluoroscopy. Fiber-Optic RealShape (FORS) technology (Philips Medical Systems) uses light-refracted through optical fibers to generate real-time renderings of wires and catheters in three-dimensional (3D) space allowing their navigation without radiation. The LumiGuide is an enhanced longer version of the standard guidewire that is coupled with a 3D Hub connector to enhance visualization of commercially available catheters. This study aims to compare the outcomes and radiation exposure using the standard FORS-enabled guidewire vs the LumiGuide for F/B-EVAR. METHODS: Data collected from 218 F/B-EVARs (129 standard FORS, 89 LumiGuide) included 931 TV catheterizations (590 with standard FORS and 341 with LumiGuide). Technical success was defined as the successful catheterization of the TVs using only FORS. FORS and LumiGuide were compared in terms of technical success rates, procedure times, catheterization duration, fluoroscopy times, and cumulative dose-area products (DAP). Welch t-test was used for continuous variables, whereas a z-test was applied for proportions. RESULTS: Patient characteristics did not demonstrate significant differences between groups. FORS technical success was 56.2% with standard FORS vs 79.4% with LumiGuide (P < .001). Compared with standard FORS, the use of LumiGuide resulted in nonsignificant reduced navigation time per TV (6.4 ± 7.4 vs 6.7 ± 9.6 minutes; P = .6), DAP per TV (3.2 ± 5.3 vs 3.36 ± 9.3 Gy·cm2; P = .79), and significant reduction in procedure time (257 ± 91 vs 298 ± 100 minutes; P = .002), fluoroscopy time (58 ± 38 vs 80 ± 40 minutes; P = .01), and contrast dose (130 ± 40 vs 150 ± 50 mL; P = .05). Cumulative DAP remained similar between groups (59 ± 38 Gy·cm2 for LumiGuide vs 64 ± 43 Gy·cm2 with standard FORS; P = .5). CONCLUSIONS: The introduction of LumiGuide and the 3D-Hub for F/B-EVAR improved FORS technical success and reduced procedure times compared with earlier FORS procedures. The ability to use different commercially available catheters with LumiGuide results in enhanced convenience and lower radiation exposure, thereby improving efficiency and safety.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Huertas P, Porras-Colon J, Figueroa A, et al. Comparison of outcomes and radiation exposure using Fiber-Optic RealShape-enabled standard guidewire versus LumiGuide for fenestrated-branched endovascular aortic aneurysm repair. J Vasc Surg. 2026. PMID: 42331152. doi:10.1016/j.jvs.2026.06.067

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。