超越颈部:EVAR术后支架依赖的近端密封——AFX与Endurant的倾向匹配比较
本研究回顾性分析365例EVAR(Endurant 267例,AFX 98例),经倾向匹配75对后比较目标预期密封区(TASZ)与实际达到密封区(RASZ)。AFX的RASZ中位数34 mm长于Endurant的25 mm,且密封面积更大;Endurant常未覆盖全部预期颈部长度。两组Ia型内漏率均<5%且无差异。
要点
- AFX的RASZ中位数34 mm,Endurant为25 mm,AFX密封更充分。
- Endurant的RASZ较TASZ中位短8 mm,AFX中位超2 mm。
- 器械类型、颈斜度及过度尺寸是密封利用的独立预测因素。
核心结论
- 尽管Ia型内漏率无器械间差异,AFX更常延伸至传统颈部以外,Endurant更常未覆盖整个预期颈部长度。
研究设计与样本
- 回顾性单中心研究,纳入2011-2022年连续365例EVAR(Endurant 267例,AFX 98例),排除TASZ短于15 mm者后倾向匹配75对。
主要发现
- 匹配后AFX的RASZ中位数34 mm vs Endurant 25 mm(p<0.001),密封面积2355 vs 1862 mm2(p<0.001)。
- Endurant的RASZ较TASZ中位短8 mm(IQR -20.5至+1 mm),AFX中位超2 mm(IQR -4.5至+11 mm;p<0.001)。
- 回归分析确认器械类型、颈斜度和过度尺寸是密封利用的独立预测因素。
- Ia型内漏发生率<5%,无器械间差异。
关键数值与推荐
- TASZ定义为指南规定的目标预期密封区,RASZ为实际达到的密封区,颈/贴附比=RASZ/TASZ。
临床怎么用
- 早期术后CTA量化RASZ可提供基于器械特性的密封行为信息,补充说明书颈部指标。
局限性
- 回顾性单中心设计,样本来自单一中心,可能存在选择偏倚。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: To evaluate how the guideline-defined target anticipated sealing zone (TASZ) translates into the real achieved sealing zone (RASZ), and whether this relationship is stent-graft-dependent. METHODS: In a retrospective single-center study, 365 consecutive endovascular abdominal aortic aneurysm repairs (Endurant, n=267; AFX, n=98; 2011-2022) were reviewed. Preoperative and first postoperative computed tomography angiography (CTA) scans were evaluated with a previously reported centerline-driven approach. Target anticipated sealing zone and RASZ were systematically evaluated. After exclusion of patients with a TASZ shorter than 15 mm, propensity-score matching produced 75 device pairs. The primary study outcome was to evaluate the relationship between TASZ and RASZ and its dependence on stent-graft design. The neck/apposition ratio (RASZ/TASZ) was calculated to quantify the proportion of the available preoperative neck length effectively used for circumferential sealing. Linear regression assessed the independent effects of oversizing and device type on the neck-to-seal ratio. RESULTS: After matching, AFX achieved longer RASZ than Endurant (median 34 vs 25 mm; p<0.001) and larger sealing surface (2355 vs 1862 mm2, p<0.001). Compared to TASZ length, Endurant showed a median shortfall of -8 mm (IQR=-20.5 to +1 mm), whereas AFX exhibited a median surplus of +2 mm (IQR=-4.5 to +11 mm; p<0.001). Regression analysis confirmed that device type, neck slope, and oversizing are independent predictors of seal utilization. Type Ia endoleak incidence was <5% without inter-device difference. CONCLUSIONS: In this retrospective analysis, despite no difference in type Ia endoleak rate, achieved proximal sealing differed between devices, with AFX more often extending beyond the traditional neck and Endurant more often not covering the entire projected neck length.Clinical ImpactThis study highlights that guideline-defined neck length does not uniformly translate into achieved proximal sealing across different stent-graft technologies. Quantifying the achieved sealing zone on early postoperative CTA provides additional anatomical information beyond instructions for use-based neck metrics and may help to better characterize device-specific sealing behavior.