文献速递主动脉疾病

现货开窗支架腔内修复腹主动脉瘤的解剖可行性研究

Anatomic feasibility of off-the-shelf fenestrated endovascular aneurysm repair devices: An anatomic study.
PubMed 文献 · J Vasc Surg · 2026-06-02

基于8238例患者CT血管造影数据,评估标准化现货开窗支架移植物对近肾及肾旁腹主动脉瘤的解剖可行性。采用贪心集合覆盖算法寻找最少配置实现开窗目标血管对齐,结果显示双开窗平台(腹腔干和肠系膜上动脉)对多数患者解剖可行,而肾动脉变异性限制多血管标准化。

要点

  • 双开窗现货平台(腹腔干和肠系膜上动脉)对多数患者解剖可行,肾动脉变异性限制多血管标准化。
  • 8种配置覆盖81.9%的肠系膜上动脉、49.7%的左肾动脉和52.9%的右肾动脉解剖。
  • 肠系膜上动脉覆盖率达90%以上需16种设计,每侧肾动脉达80%覆盖需20种配置。

核心结论

  • 包含腹腔干和肠系膜上动脉的标准化双开窗现货平台对多数患者解剖可行,而肾动脉变异性限制多血管标准化。
  • 带肾分支的双开窗装置也可能规避这些限制。

主要发现

  • 解剖数据可用率:肠系膜上动脉57.1%,左肾动脉96.1%,右肾动脉96.8%。
  • 8种配置覆盖81.9%的肠系膜上动脉、49.7%的左肾动脉和52.9%的右肾动脉解剖。
  • 肠系膜上动脉覆盖率达90%以上需16种设计,每侧肾动脉达80%覆盖需20种配置。
  • 未观察到血管起源按主动脉直径(中位数28 mm,四分位距26-30.5 mm)聚集。
  • 前100种配置对三血管组合覆盖低于70%,对四血管组合覆盖低于25%。

临床怎么用

  • 双开窗现货平台可考虑用于腹腔干和肠系膜上动脉区域,但肾动脉受累时需个体化或带肾分支设计。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Fenestrated endovascular aneurysm repair enables treatment of complex aortic aneurysms involving visceral branches but is limited by patient-specific anatomical variability. Off-the-shelf (OTS) devices may reduce manufacturing delay; however, their feasibility depends on anatomic compatibility across populations. DESIGN: A large multicenter dataset was analyzed to evaluate the anatomic feasibility of standardized OTS fenestrated stent-graft configurations for juxtarenal and pararenal abdominal aortic aneurysms in this retrospective anatomic study. METHODS: Computed tomography angiography data from a sample of 8238 patients evaluated for fenestrated endovascular aneurysm repair between January 2018 and December 2023 were analyzed. Measurements of the celiac axis (CA), superior mesenteric artery (SMA), left renal artery (LRA), and right renal artery (RRA) included ostial distance and branching angle relative to the CA. A greedy set-cover algorithm identified the smallest number of OTS configurations achieving fenestration target vessel alignment (FTVA). Cumulative coverage and clustering by aortic diameter were assessed. Analyses were exploratory and descriptive. RESULTS: Anatomic data were available in 57.1% of cases for the SMA, 96.1% for the LRA, and 96.8% for the RRA. The CA served as the reference point for all analyses. Eight configurations covered 81.9% of SMA, 49.7% of LRA, and 52.9% of RRA anatomies. SMA coverage exceeded 90% with 16 designs, whereas 20 configurations were required for each renal artery to reach 80% coverage. No clustering of vessel origin by aortic diameter (median, 28 mm; interquartile range, 26-30.5 mm) was observed. Below 70% coverage for three-vessel and <25% for four-vessel combinations was achieved with the top 100 configurations. CONCLUSIONS: A standardized dual-fenestrated OTS platform incorporating the CA and SMA appears anatomically feasible for most patients, whereas renal artery variability limits multivessel standardization. Dual-fenestrated devices with renal branches could also circumvent these constraints.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Hofmann A, Leinweber M, Kliewer M, et al. Anatomic feasibility of off-the-shelf fenestrated endovascular aneurysm repair devices: An anatomic study. J Vasc Surg. 2026. PMID: 42235837. doi:10.1016/j.jvs.2026.05.038

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。