本研究评估新型机器人显微手术平台在3例患者上肢建立透析用动静脉瘘的可行性。传统血管准备后,由经验丰富的血管外科医生用该平台完成静脉至动脉端侧连续缝合吻合,吻合时间28至36分钟。全部3例吻合成功,随访超声显示瘘管重建成功且无并发症。

要点

  • 机器人显微手术平台可完成上肢透析用动静脉瘘的端侧连续缝合吻合。
  • 3例患者均成功建立动静脉瘘,吻合时间28至36分钟。
  • 随访超声显示全部瘘管重建成功且无并发症。

核心结论

  • 机器人显微手术平台已成功用于建立透析治疗的精准动静脉吻合。

研究设计与样本

  • 临床可行性研究,共纳入3例患者,在传统血管准备后使用机器人显微手术平台完成动静脉瘘吻合。

主要发现

  • 全部3例患者均使用该显微手术设备成功建立动静脉瘘。
  • 吻合时间在28至36分钟之间。
  • 1例在血流恢复后需追加一针缝合,同样使用该设备完成。
  • 随访超声检查显示全部患者瘘管重建成功且无任何并发症。

关键数值与推荐

  • 吻合方式为静脉至动脉端侧连续缝合,由经验丰富的血管外科医生操作。
  • 吻合完成后恢复血流并检测渗漏。

临床怎么用

  • 该平台可用于小血管的精准动静脉吻合,可能改善组织保护、精度和可视化。

局限性

  • 样本量仅3例,随访信息有限,未报告长期通畅率。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Modern robotic systems, particularly for microsurgical procedures, are offering increasingly new perspectives in the field of vascular surgery. A novel robot-assisted platform specially designed for microanastomoses is already used successfully in lymphatic vessel surgery and special flap plasty procedures. To expand these applications to the vascular field, this clinical feasibility study evaluated this system for creating arteriovenous fistulas in the upper extremity for dialysis. METHODS: Following conventional vessel preparation, the anastomoses for the arteriovenous fistulas were performed using the microsurgical robotic platform. An experienced vascular surgeon performed the anastomoses (vein to artery, end-to-side) with continuous sutures. Once the anastomoses had been completed and blood flow was restored, they were tested for leaks. RESULTS: In all 3 patients, an arteriovenous fistula was successfully created using the microsurgical device. Preparing and positioning the robotic system was straightforward and uncomplicated. Anastomosis times varied between 28 and 36 minutes. In one case, a second stitch was required after blood flow was restored, which was also performed using the device. Follow-up ultrasound examinations revealed successful fistula reconstruction in all patients without any complications. CONCLUSIONS: The microsurgical robotic platform has been successfully used to create precise arteriovenous anastomoses for dialysis therapy. This robotic system may offer a variety of advantages, especially for small vessels, including improved tissue preservation, precision, and visualization. Further development of these technologies could lead to new applications, particularly in vascular medicine.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Rusch M, Hoffmann G, Spille J, et al. A Technical and Safety Feasibility Study of Creating Arteriovenous Fistulas for Dialysis Using a Novel Robotic Microsurgical System. Ann Vasc Surg. 2026. PMID: 41796663. doi:10.1016/j.avsg.2026.02.043

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。